ASTM E1867-2011 Standard Test Method for Temperature Calibration of Dynamic Mechanical Analyzers《动态机械分析仪的温度校准的标准试验方法》.pdf
《ASTM E1867-2011 Standard Test Method for Temperature Calibration of Dynamic Mechanical Analyzers《动态机械分析仪的温度校准的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1867-2011 Standard Test Method for Temperature Calibration of Dynamic Mechanical Analyzers《动态机械分析仪的温度校准的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1867 11Standard Test Method forTemperature Calibration of Dynamic Mechanical Analyzers1This standard is issued under the fixed designation E1867; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisi
2、on. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the temperature calibrationof dynamic mechanical analyzers (DMA) from 150 C to500 C.1.2 The values sta
3、ted in SI units are to be regarded asstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regul
4、atory limitations prior to use. Specific precau-tionary statements are given in Note 7.2. Referenced Documents2.1 ASTM Standards:2E473 Terminology Relating to Thermal Analysis and Rhe-ologyE1142 Terminology Relating to Thermophysical PropertiesE2161 Terminology Relating to Performance Validation inT
5、hermal Analysis3. Terminology3.1 Definitions:3.1.1 The technical terms used in this test method aredefined in Terminologies E473, E1142, and E2161, includingdynamic mechanical analysis, frequency, stress, strain andstorage modulus.4. Summary of Test Method4.1 An equation is developed for the linear
6、correlation ofexperimentally observed program or sensor temperature andthe actual melting temperature for known melting referencematerials. This is accomplished by loading melting pointreference materials into a polymer tube, or wrapping them withpolymer tape and subjecting it to a mechanical oscill
7、ation ateither fixed or resonant frequency. The extrapolated onset ofmelting is identified by a rapid decrease in the ordinate signal(the apparent storage modulus, stress, inverse strain or probeposition). This onset is used for temperature calibration withtwo melting point reference materials.5. Si
8、gnificance and Use5.1 Dynamic mechanical analyzers monitor changes in theviscoelastic properties of a material as a function of tempera-ture and frequency, providing a means to quantify thesechanges. In most cases, the value to be assigned is thetemperature of the transition (or event) under study.
9、Therefore,the temperature axis (abscissa) of all DMA thermal curvesmust be accurately calibrated by adjusting the apparent tem-perature scale to match the actual temperature over thetemperature range of interest.6. Interferences6.1 An increase or decrease in heating rates or change inpurge gas type
10、or rate from those specified may alter results.6.2 Once the temperature calibration procedure has beenexecuted, the measuring temperature sensor position shall notbe changed, nor shall it be in contact with the specimen orspecimen holder in a way that would impede movement. If thetemperature sensor
11、position is changed or is replaced, then theentire calibration procedure shall be repeated.6.3 Once the temperature calibration has been executed, thegeometry deformation (bending study, versus tensile, and thelike) shall not be changed. If the specimen testing geometrydiffers significantly from tha
12、t of the calibrants, then thecalibration shall be repeated in the geometry matching that ofspecimen testing.6.4 This method does not apply to calibration for shear orcompressive geometries of deformation.7. Apparatus7.1 The function of the apparatus is to hold a specimen ofuniform dimension so that
13、the specimen acts as the elastic anddissipative element in a mechanically oscillated system. Dy-namic mechanic analyzers typically operate in one of severalmodes as outlined in Table 1.1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsib
14、ility of Subcommittee E37.10 on Funda-mental, Statistical and Mechanical Properties.Current edition approved Aug. 1, 2011. Published September 2011. Originallyapproved in 1997. Last previous edition approved in 2006 as E1867 06. DOI:10.1520/E1867-11.2For referenced ASTM standards, visit the ASTM web
15、site, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United Sta
16、tes.7.1.1 The apparatus shall consist of the following:7.1.1.1 ClampsAclamping arrangement that permits grip-ping of the specimen. This may be accomplished by clampingat both ends (most systems), one end (for example, torsionalpendulum) or neither end (for example, free bending betweenknife edges).7
17、.1.1.2 Device to Apply Oscillatory Stress or StrainAdevice for applying an oscillatory deformation (strain) oroscillatory stress to the specimen. The deformation may beapplied and then released, as in freely vibrating devices, orcontinually applied, as in forced vibration devices.7.1.1.3 DetectorA d
18、evice or devices for determining thedependent and independent experimental parameters, such asforce (stress), deformation (strain), frequency, and temperature.Temperature shall be measurable with an accuracy of 60.1 C,force to 61 % and frequency to 61%.7.1.1.4 Temperature Controller and OvenA device
19、 forcontrolling the specimen temperature, either by heating, cool-ing (in steps or ramps), or by maintaining a constant experi-mental environment. The temperature programmer shall besufficiently stable to permit measurement of specimen tempera-ture to 0.1 C.7.1.1.5 A Data Collection Device, to provi
20、de a means ofacquiring, storing, and displaying measured or calculatedsignals, or both. The minimum output signals required fordynamic mechanical analysis are storage modulus, loss modu-lus, tangent delta, temperature, and time.NOTE 1Some instruments, suitable for this test, may display onlylinear o
21、r logarithmic storage modulus while others may display linearand/or logarithmic storage modulus. Care must be taken to use the samemodulus scale when comparing unknown specimens, and in the compari-son of results from one instrument to another.7.2 High Temperature Polymer Tubing such as PTFE (Poly-t
22、etrafluoroethylene) or PEEK (Polyetheretherketone), of 3-mmoutside diameter and wall thickness of 0.5-mm (0.002 in.)3inner diameter may be used for low temperature standards (thatis, less than 160 C). The tubing may be sealed with suitablemelting temperature wax plugs, or similar sealant.NOTE 2PTFE
23、tubing is selected for its flexibility and inert nature forthe solvents in use at the temperatures of interest. Furthermore itstransitions should not produce any interference in the DMA signal withinthe range of the suggested calibrant materials. PEEK provides increasedstiffness for ease of loading.
24、 For other temperature ranges, a suitablereplacement for the high temperature polymer tubing may be used.7.3 Where the melting material is to be confined to a tube7.4 PTFE Tape, for wrapping metal point standards.7.5 Calibration MaterialsOne or more suitable materialspresented in Table 2.7.6 Caliper
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME18672011STANDARDTESTMETHODFORTEMPERATURECALIBRATIONOFDYNAMICMECHANICALANALYZERS 动态 机械 分析 温度 校准 标准

链接地址:http://www.mydoc123.com/p-529677.html