ASTM UOP603-2018 Trace CO and CO2 in Hydrogen and Light Gaseous Hydrocarbons by Gas Chromatography.pdf
《ASTM UOP603-2018 Trace CO and CO2 in Hydrogen and Light Gaseous Hydrocarbons by Gas Chromatography.pdf》由会员分享,可在线阅读,更多相关《ASTM UOP603-2018 Trace CO and CO2 in Hydrogen and Light Gaseous Hydrocarbons by Gas Chromatography.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1971, 1988, 2012, 2013, 2018 UOP LLC. All
3、 rights reserved. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.
4、832.9555 FAX, or 610.832.9585 PHONE. Trace CO and CO2 in Hydrogen and Light Gaseous Hydrocarbons by Gas Chromatography UOP Method 603-18 Scope This method is for determining low concentrations (0.2 to 500 mol-ppm) of carbon monoxide (CO) and carbon dioxide (CO2) in high purity hydrogen, methane and
5、other gas-phase samples. The lower limit of quantitation for CO and CO2 in samples containing high concentrations of hydrogen, nitrogen, and other gases, is 0.2 mol-ppm. For samples containing high methane concentrations, an analysis under a different valve configuration is necessary to resolve carb
6、on dioxide from interfering methane. References Analyzer manufacturers operating manual UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method A repeatable volume of sample is introduced through the gas sampling/backflush valve into a two-column system. After the eluti
7、on of CO2 from the first column into the second column, the valve is switched to back flush the components remaining in the first column to the vent. For samples containing high methane, the 4-port switching valve is used as a heart cut valve to vent the bulk of methane eluting from the first column
8、. CO, CO2 and methane are separated on the second column and converted in the methanizer into methane. A flame ionization detector is used to detect the methane produced. Concentrations of CO and CO2, are calculated using the external standard method of quantitation. Apparatus References to catalog
9、numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Analyzer, pre-configured, including a data system, AC Analytical Controls (PAC). A flow diagram of the AC analyzer is shown in Figure 1. Pre-configured analyzers are also available from Agilent Techn
10、ologies, PerkinElmer Separation Systems, and other suppliers. Confirm with the supplier that the analyzer is appropriate for the stream to be analyzed and the analysis required. Instrument gas supply regulators; not all instruments will require all regulators listed below. Regulator, air, two-stage,
11、 high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590 2 of 6 603-18 Regulator, helium, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Regulator, hydrogen, two-stage, high purity, delivery pressure rang
12、e 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-550 Regulator, nitrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Leak detector, gas, ThermoFisher Scientific GLD Gas Leak Detector, Cat. No. 66002-001 Reagents and Materials Refere
13、nces to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Columns, replacement, 2 m of 1/16-inch OD x 0.050-inch ID stainless steel tubing, micropacked “Hayesep Q”, two required. These are typical columns used for the analysis. Contact the a
14、nalyzer manufacturer for specific columns and part numbers. Instrument gases; not all instruments will require all gases listed below. Air, zero gas, total hydrocarbons less than 2.0 ppm as methane Helium, zero gas, total hydrocarbons less than 0.5 ppm as methane Hydrogen, zero gas, total hydrocarbo
15、ns less than 0.5 ppm as methane Nitrogen, zero gas, total hydrocarbons less than 0.5 ppm as methane Gas calibration blends; purchase only those required for the sample compositions to be analyzed Blend 1, 5 mol-ppm CO, 5 mol-ppm CO2 and 25 mol-ppm methane in hydrogen, Matheson Tri-Gas or local suppl
16、y Blend 2, 200 mol-ppm CO, 200 mol-ppm CO2 and 200 mol-ppm methane in hydrogen, Matheson Tri-Gas or local supply Blend 3, 5 mol-ppm CO, 5 mol-ppm CO2 in methane, Matheson Tri-Gas or local supply Gas purifiers, for carrier gases, to remove oxygen and moisture, VICI Metronics, Cat. Nos. P100-1 (helium
17、), P200-1 (hydrogen), P300-1 (nitrogen), and (optional) indicating oxygen traps, Restek, Cat. No. 22010, as needed for the specific instrument used Methanizer, replacement, nickel catalyst tube. Contact the analyzer manufacturer for specific configuration and part number. Procedure The analyst is ex
18、pected to be familiar with general laboratory practices, the technique of gas chromatography, and the equipment being used. Dispose of used reagents, materials, and samples in an environmentally safe manner according to local regulations. Preparation of Apparatus 1. Install the gas purifier in the s
19、upply line between the carrier gas source and the carrier gas inlets on the gas chromatograph. Column life is significantly reduced if gas purifiers are not used. Replace the gas purifiers at intervals determined by good laboratory practice. Indicating oxygen traps may be placed downstream of the ga
20、s purifiers. When the indicator shows one-half used, replace both the gas purifier and the 3 of 6 603-18 indicating trap. 2. Follow the instrument set-up procedure provided by the manufacturer but do not initially apply heat to the column oven; see Column Conditioning, below. CAUTION: Hydrogen leaka
21、ge into the confined volume of the column oven can cause a violent explosion. Therefore, it is mandatory to check for leaks each time a connection is made and periodically thereafter. Column Conditioning The columns must be conditioned to remove any traces of solvent or volatile components in the li
22、quid phase and allow the phase to distribute itself evenly over the surface of the solid support. For porous polymers, conditioning is necessary to remove residual solvents used during polymerization and any compounds adsorbed during storage and handling. This column conditioning is required when th
23、e instrument is first installed, and whenever columns are replaced. Pre-condition the chromatographic columns individually in the chromatograph column oven using the following procedures: 1. Connect the column to the inlet and flush the column with carrier gas to remove air. Do not connect the colum
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMUOP6032018TRACECOANDCO2INHYDROGENANDLIGHTGASEOUSHYDROCARBONSBYGASCHROMATOGRAPHYPDF

链接地址:http://www.mydoc123.com/p-1244049.html