ASTM UOP1009-2015 Total Mercury in Solids《固体中的总汞含量》.pdf
《ASTM UOP1009-2015 Total Mercury in Solids《固体中的总汞含量》.pdf》由会员分享,可在线阅读,更多相关《ASTM UOP1009-2015 Total Mercury in Solids《固体中的总汞含量》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2015 UOP LLC. All rights reserved. Nonconfidenti
3、al UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. To
4、tal Mercury in Solids UOP Method 1009-15 Scope This method is for determining total mercury in solid materials using a Nippon Instruments Corporation (NIC) MA-3000 Mercury Analyzer. The instrument uses a dual path detector to measure from the lower limit of 0.1 ng/g (mass-ppb) to an upper limit of 5
5、00 g/g (mass-ppm), when analyzing a 50 mg sample. This procedure can also be used for aqueous liquid samples and high boiling hydrocarbons (such as crude or shale oil) through the use of appropriate combustion programs. Volatile hydrocarbons (e.g. naphtha and BTX aromatics) should be analyzed by UOP
6、 Method 938. References UOP Method 938, “Total Mercury and Mercury Species in Liquid Hydrocarbons,” www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The NIC Mercury Analyzer/Model MA-3000, consisting of an autosampler, a mercury atomizer and integrat
7、ed mercury detector, is specified for this method. The sample is decomposed by heating in oxygen and the mercury vapor in the gaseous products is trapped by the mercury collector as gold amalgam. The temperature of the mercury collector is kept at 210 C to prevent the adsorption of non- mercury comb
8、ustion products on the collector. After collection, the mercury is liberated by heating the collector to 700+ C. The vaporized mercury is carried to a dual absorption cell (long and short path lengths) with a pure oxygen carrier gas and detected by the cold vapor atomic absorption technique. After i
9、ntroducing the sample into boats and loading the autosampler rack, all operations from the sample decomposition process to the mercury detection and calculation are carried out automatically by the instrument. The instrument collects data from both cells, and the selection of High or Low range can b
10、e changed by the operator after the sample analysis has completed. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used, unless stated otherwise. Balance, readability 0.01-mg, Mettler-Toledo XA105 or equivalent. If a diff
11、erent balance is used, it should be one that will directly communicate with the MA3Win software. 2 of 10 1009-15 Cylinder, graduated, 500-ml, Class B, VWR, Cat. No. 89000-258 Desiccator, VWR, Cat. No. 25031-048 (Corning 3120-250), optional Flasks, volumetric, glass, Class A, 100- (several required),
12、 500- and 1000-mL, Fisher Scientific, Cat. Nos. 10-210-8C, F and G, respectively Furnace, muffle, 120 V, 50/60 Hz, 1200 C maximum, Fisher Scientific, Cat. No. 10-554-28, optional Gloves, thermal protection, VWR, Cat. No. 32885-407, optional Mercury analyzer, NIC, Model MA-3000 (no substitute), with
13、the following accessories: Boats, ceramic, for MA-3000, NIC, Cat. No. S1106-6100-00 Sample heating tube MA-3, NIC, Cat. No. S1106-0620-00. Tubes with catalyst for Appendix K should be used with high halogen samples. Sample boat fixture (for measurement), NIC, Cat. No. N-1106-901-400 Mortar and Pestl
14、e, Agate, Fisher Scientific, Cat. No. 12-950B Pipet, digital, variable volume 20-200 L, Fisher Scientific, Cat. No. 21-377-289 (required for the analysis of heavy hydrocarbon and calibration samples) Pipets, volumetric, Class A, 1-, 2-, 5-, and 10-mL, Fisher Scientific, Cat. Nos. 13-660A, B, E, and
15、F, respectively Pipet filler, Fisher Scientific, Cat. No. 13-681-51 Refrigerator, explosion proof or flammable storage Regulator, 2-stage Stainless Steel, CGA 540, Matheson Cat. No. 3810-540 Sample containers, glass or Teflon bottles, acid cleaned, see Procedure Tongs, crucible, VWR, Cat. No. 82027-
16、372, for muffle furnace use Tweezers, flat, round point, Baxter Scientific, Cat. No. E1200-9 Wash bottle, Nalgene , LDPE, 500-mL, Fisher Scientific, Cat. No. 03-409-17F (used for nitric acid washing of all glassware) Wash bottle, Nalgene, Right to Know Safety, 500-mL, for acetone, Fisher Scientific,
17、 Cat. No. 03-409- 11A Wash bottle, Nalgene, Right to Know Safety, 500-mL, for water, Fisher Scientific, Cat. No. 03-409- 11E Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used, unless stated otherwise. Refe
18、rences to water mean deionized or distilled water. Acetone, Fisher Scientific, Cat. No. A18-4 Aluminum Foil, local supply, optional L-Cysteine, 99% purity, Pfaltz and Bauer, Cat. No. C33800 Mercuric chloride, mercury (II) chloride, 99.999% purity, Sigma-Aldrich, Cat. No. 20-377-7 (Note: Toxic materi
19、al, review SDS before handling and wear appropriate personal protective equipment) 3 of 10 1009-15 Nitric acid, concentrated, Trace Metal Grade, Fisher Scientific, Cat. No. A509-212 Nitric Acid, 1:1. Combine 200 mL of concentrated nitric acid with 200 mL of water in a plastic wash bottle, swirl to m
20、ix, and use freshly prepared while still warm. Nitrogen, or oil-free clean air, for blow drying glassware Oxygen, UHP, Airgas Cat. No. OX UHP200 Pipet tips, disposable, Fisher Scientific, Cat. No. 21-377-45 QC standard, a certified Hg standard should be obtained for use as a Quality Control (QC) sta
21、ndard. An example is NIST 1632d (Trace Elements in Coal) at 0.0928 mg/kg Hg. Transfer pipets, disposable, Fisher Scientific Cat. No. 13-711-9AM Water, deionized or distilled. Certified low Hg reagent water can be purchased from O2SI, Cat. No. 190007-16 Weighing paper, Fisher Scientific, Cat. No. 09-
22、898-12A Procedure The analyst is expected to be familiar with general laboratory practices and the equipment being used. Portions of these procedures are specific to the specified apparatus. Sampling Samples should be collected in a manner that assures representative sampling. Elemental mercury and
23、organic-mercury species are unstable and will react with the walls of the sample containers. Do not collect samples in metal containers. Mercury present can adsorb on the metal. Liquid samples must be collected in previously acid cleaned glass containers and then analyzed as soon as possible. Instru
24、ment Set Up 1. Set up the instrument according to the manufacturers instructions. 2. Turn on the power to the instrument and start the MA3Win control software on the PC. The instrument is normally turned off overnight and when not in use. Standard operating conditions are shown in Table 1. 3. Allow
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMUOP10092015TOTALMERCURYINSOLIDS 固体 中的 含量 PDF

链接地址:http://www.mydoc123.com/p-540818.html