ASTM E1866-1997(2013) Standard Guide for Establishing Spectrophotometer Performance Tests《分光光度计性能试验确定用标准指南》.pdf
《ASTM E1866-1997(2013) Standard Guide for Establishing Spectrophotometer Performance Tests《分光光度计性能试验确定用标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1866-1997(2013) Standard Guide for Establishing Spectrophotometer Performance Tests《分光光度计性能试验确定用标准指南》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1866 97 (Reapproved 2013)Standard Guide forEstablishing Spectrophotometer Performance Tests1This standard is issued under the fixed designation E1866; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last r
2、evision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers basic procedures that can be used todevelop spectrophotometer performance tests. The guide isintended to b
3、e applicable to spectrophotometers operating inthe ultraviolet, visible, near-infrared and mid-infrared regions.1.2 This guide is not intended as a replacement for specificpractices such as Practices E275, E925, E932, E958, E1421,orE1683 that exist for measuring performance of specific types ofspect
4、rophotometers. Instead, this guide is intended to provideguidelines in how similar practices should be developed whenspecific practices do not exist for a particular spectrophotom-eter type, or when specific practices are not applicable due tosampling or safety concerns. This guide can be used to de
5、velopperformance tests for on-line process spectrophotometers.1.3 This guide describes univariate level zero and level onetests, and multivariate level A and level B tests which can beimplemented to measure spectrophotometer performance.These tests are designed to be used as rapid, routine checks of
6、spectrophotometer performance. They are designed to uncovermalfunctions or other changes in instrument operation, but donot specifically diagnose or quantitatively assess the malfunc-tion or change. The tests are not intended for the comparison ofspectrophotometers of different manufacture.1.4 This
7、standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2
8、.1 ASTM Standards:2E131 Terminology Relating to Molecular SpectroscopyE275 Practice for Describing and Measuring Performance ofUltraviolet and Visible SpectrophotometersE925 Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidthdoes not Exceed 2 nm
9、E932 Practice for Describing and Measuring Performance ofDispersive Infrared SpectrometersE958 Practice for Measuring Practical Spectral Bandwidthof Ultraviolet-Visible SpectrophotometersE1421 Practice for Describing and Measuring Performanceof Fourier Transform Mid-Infrared (FT-MIR) Spectrom-eters:
10、 Level Zero and Level One TestsE1655 Practices for Infrared Multivariate QuantitativeAnalysisE1683 Practice for Testing the Performance of ScanningRaman Spectrometers3. Terminology3.1 DefinitionsFor terminology relating to molecularspectroscopic methods, refer to Terminology E131.3.2 Definitions of
11、Terms Specific to This Standard:3.2.1 action limit, nthe limiting value from an instrumentperformance test, beyond which the spectrophotometer isexpected to produce potentially invalid results.3.2.2 check sample, na single pure compound, or aknown, reproducible mixture of compounds whose spectrum is
12、constant over time such that it can be used in a performancetest.3.2.3 level A test, na pass/fail spectrophotometer perfor-mance test in which the spectrum of a check or test sample iscompared against historical spectra of the same sample via amultivariate analysis.3.2.4 level B test, na pass/fail s
13、pectrophotometer perfor-mance test in which the spectrum of a check or test sample isanalyzed using a multivariate model, and the results of theanalysis are compared to historical results for prior analyses ofthe same sample.3.2.5 level one (1) test, na simple series of measurementsdesigned to provi
14、de quantitative data on various aspects ofspectrophotometer performance and information on which tobase the diagnosis of problems.3.2.6 level zero (0) test, na routine check of spectropho-tometer performance, which can be done in a few minutes,designed to visually detect significant changes in instr
15、ument1This guide is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.03 on Infrared and Near Infrared Spectroscopy.Current edition approved Jan. 1, 2013. Published January 2013. Originallyapproved in 19
16、97. Last previous edition approved in 2007 as E1866 97 (2007).DOI: 10.1520/E1866-97R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summar
17、y page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1performance and provide a database to determine instrumentperformance over time.3.2.7 optical reference filter, nan optical filter or otherdevice which can be i
18、nserted into the optical path in thespectrophotometer or probe producing an absorption spectrumwhich is known to be constant over time such that it can beused in place of a check or test sample in a performance test.3.2.8 test sample, na process or product sample, or amixture of process or product s
19、amples which has a constantspectrum for a finite time period and which can be used in aperformance test. Test samples and their spectra are generallynot reproducible in the long term.4. Significance and Use4.1 If ASTM Committee E13 has not specified an appropri-ate test procedure for a specific type
20、 of spectrophotometer, orif the sample specified by a Committee E13 procedure isincompatible with the intended spectrophotometer operation,then this guide can be used to develop practical performancetests.4.1.1 For spectrophotometers which are equipped with per-manent or semi-permanent sampling acce
21、ssories, the testsample specified in a Committee E13 practice may not becompatible with the spectrophotometer configuration. Forexample, for FT-MIR instruments equipped with transmittanceor IRS flow cells, tests based on polystyrene films are imprac-tical. In such cases, these guidelines suggest mea
22、ns by whichthe recommended test procedures can be modified so as to beperformed on a compatible test material.4.1.2 For spectrophotometers used in processmeasurements, the choice of test materials may be limited dueto process contamination and safety considerations. Theseguidelines suggest means of
23、developing performance testsbased on materials which are compatible with the intended useof the spectrophotometer.4.2 Tests developed using these guidelines are intended toallow the user to compare the performance of a spectropho-tometer on any given day with prior performance. The tests areintended
24、 to uncover malfunctions or other changes in instru-ment operation, but they are not designed to diagnose orquantitatively assess the malfunction or change. The tests arenot intended for the comparison of spectrophotometers ofdifferent manufacture.5. Test Conditions5.1 When conducting the performanc
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME186619972013STANDARDGUIDEFORESTABLISHINGSPECTROPHOTOMETERPERFORMANCETESTS 分光 光度计 性能 试验 确定 标准 指南

链接地址:http://www.mydoc123.com/p-529675.html