ASTM E1866-1997(2007) Standard Guide for Establishing Spectrophotometer Performance Tests《建立光谱仪性能试验的标准指南》.pdf
《ASTM E1866-1997(2007) Standard Guide for Establishing Spectrophotometer Performance Tests《建立光谱仪性能试验的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1866-1997(2007) Standard Guide for Establishing Spectrophotometer Performance Tests《建立光谱仪性能试验的标准指南》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1866 97 (Reapproved 2007)Standard Guide forEstablishing Spectrophotometer Performance Tests1This standard is issued under the fixed designation E 1866; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers basic procedures that can be used todevelop spectrophotometer performance tests. The guide isintended t
3、o be applicable to spectrophotometers operating inthe ultraviolet, visible, near-infrared and mid-infrared regions.1.2 This guide is not intended as a replacement for specificpractices such as Practices E 275, E 925, E 932, E 958, E 1421,or E 1683 that exist for measuring performance of specifictype
4、s of spectrophotometers. Instead, this guide is intended toprovide guidelines in how similar practices should be devel-oped when specific practices do not exist for a particularspectrophotometer type, or when specific practices are notapplicable due to sampling or safety concerns. This guide canbe u
5、sed to develop performance tests for on-line processspectrophotometers.1.3 This guide describes univariate level zero and level onetests, and multivariate level A and level B tests which can beimplemented to measure spectrophotometer performance.These tests are designed to be used as rapid, routine
6、checks ofspectrophotometer performance. They are designed to uncovermalfunctions or other changes in instrument operation, but donot specifically diagnose or quantitatively assess the malfunc-tion or change. The tests are not intended for the comparison ofspectrophotometers of different manufacture.
7、1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced D
8、ocuments2.1 ASTM Standards:2E 131 Terminology Relating to Molecular SpectroscopyE 275 Practice for Describing and Measuring Performanceof Ultraviolet, Visible, and Near-Infrared Spectrophotom-etersE 387 Test Method for Estimating Stray Radiant PowerRatio of Dispersive Spectrophotometers by the Opaqu
9、eFilter MethodE 925 Practice for Monitoring the Calibration ofUltraviolet-Visible Spectrophotometers whose SpectralSlit Width does not Exceed 2 nmE 932 Practice for Describing and Measuring Performanceof Dispersive Infrared SpectrometersE 958 Practice for Measuring Practical Spectral Bandwidthof Ult
10、raviolet-Visible SpectrophotometersE 1421 Practice for Describing and Measuring Performanceof Fourier Transform Mid-Infrared (FT-MIR) Spectrom-eters: Level Zero and Level One TestsE 1655 Practices for Infrared Multivariate QuantitativeAnalysisE 1683 Practice for Testing the Performance of ScanningRa
11、man Spectrometers3. Terminology3.1 DefinitionsFor terminology relating to molecularspectroscopic methods, refer to Terminology E 131.3.2 Definitions of Terms Specific to This Standard:3.2.1 action limit, nthe limiting value from an instrumentperformance test, beyond which the spectrophotometer isexp
12、ected to produce potentially invalid results.3.2.2 check sample, na single pure compound, or aknown, reproducible mixture of compounds whose spectrum isconstant over time such that it can be used in a performancetest.3.2.3 level A test, na pass/fail spectrophotometer perfor-mance test in which the s
13、pectrum of a check or test sample iscompared against historical spectra of the same sample via amultivariate analysis.3.2.4 level B test, na pass/fail spectrophotometer perfor-mance test in which the spectrum of a check or test sample isanalyzed using a multivariate model, and the results of theanal
14、ysis are compared to historical results for prior analyses ofthe same sample.1This guide is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.03 on Infrared and Near Infrared Spectroscopy.Current edition
15、 approved Dec. 1, 2007. Published December 2007. Originallyapproved in 1997. Last previous edition approved in 2002 as E 1866 - 97 (2002).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume
16、information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.2.5 level one (1) test, na simple series of measurementsdesigned to provide quantitative data on various a
17、spects ofspectrophotometer performance and information on which tobase the diagnosis of problems.3.2.6 level zero (0) test, na routine check of spectropho-tometer performance, which can be done in a few minutes,designed to visually detect significant changes in instrumentperformance and provide a da
18、tabase to determine instrumentperformance over time.3.2.7 optical reference filter, nan optical filter or otherdevice which can be inserted into the optical path in thespectrophotometer or probe producing an absorption spectrumwhich is known to be constant over time such that it can beused in place
19、of a check or test sample in a performance test.3.2.8 test sample, na process or product sample, or amixture of process or product samples which has a constantspectrum for a finite time period and which can be used in aperformance test. Test samples and their spectra are generallynot reproducible in
20、 the long term.4. Significance and Use4.1 If ASTM Committee E13 has not specified an appropri-ate test procedure for a specific type of spectrophotometer, orif the sample specified by a Committee E13 procedure isincompatible with the intended spectrophotometer operation,then this guide can be used t
21、o develop practical performancetests.4.1.1 For spectrophotometers which are equipped with per-manent or semi-permanent sampling accessories, the testsample specified in a Committee E13 practice may not becompatible with the spectrophotometer configuration. For ex-ample, for FT-MIR instruments equipp
22、ed with transmittance orIRS flow cells, tests based on polystyrene films are impractical.In such cases, these guidelines suggest means by which therecommended test procedures can be modified so as to beperformed on a compatible test material.4.1.2 For spectrophotometers used in process measure-ments
23、, the choice of test materials may be limited due toprocess contamination and safety considerations. These guide-lines suggest means of developing performance tests based onmaterials which are compatible with the intended use of thespectrophotometer.4.2 Tests developed using these guidelines are int
24、ended toallow the user to compare the performance of a spectropho-tometer on any given day with prior performance. The tests areintended to uncover malfunctions or other changes in instru-ment operation, but they are not designed to diagnose orquantitatively assess the malfunction or change. The tes
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便