ASTM D3920-2002(2007)e1 Standard Test Method for Strontium in Water《水中锶含量的标准试验方法》.pdf
《ASTM D3920-2002(2007)e1 Standard Test Method for Strontium in Water《水中锶含量的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D3920-2002(2007)e1 Standard Test Method for Strontium in Water《水中锶含量的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 3920 02 (Reapproved 2007)e1Standard Test Method forStrontium in Water1This standard is issued under the fixed designation D 3920; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in
2、 parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.e1NOTEEditorial changes were made throughout in September 2007.1. Scope1.1 This test method covers the determination of dissolvedand total recoverable stro
3、ntium in water and wastewater byatomic absorption spectroscopy.1.2 The test method is applicable in the range from 0.1 to 1mg/L of strontium. The range may be extended by dilution ofthe original sample.1.3 Round-robin data were obtained in natural and reagentwater matrices. It is the users responsib
4、ility to ensure thevalidity of the test method for waters of untested matrices.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and dete
5、rmine the applica-bility of regulatory limitations prior to use. Specific precau-tionary statements are given in 8.4 and 8.9.2. Referenced Documents2.1 ASTM Standards:2D 1129 Terminology Relating to WaterD 1193 Specification for Reagent WaterD 2777 Practice for Determination of Precision and Bias of
6、Applicable Test Methods of Committee D19 on WaterD 3370 Practices for Sampling Water from Closed ConduitsD 4691 Practice for Measuring Elements in Water by FlameAtomic Absorption SpectrophotometryD 4841 Practice for Estimation of Holding Time for WaterSamples Containing Organic and Inorganic Constit
7、uentsD 5810 Guide for Spiking into Aqueous SamplesD 5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water Analysis3. Terminology3.1 For definitions of terms used in this test method, refer toTerminology D 1129.3.2 Definitions of Terms Specific to This Standard:3
8、.2.1 total recoverable strontium, nan arbitrary analyticalterm relating to the forms of strontium that are determinable bythe digestion procedure described in this test method.3.2.2 laboratory control sample, na solution with a cer-tified concentration of the strontium.4. Summary of Test Method4.1 S
9、trontium is determined by atomic absorption spectro-photometry. The sample is aspirated into an air-acetylene flamefollowing the addition of lanthanum chloride/potassium chlo-ride solution. Samples containing particulate matter that mayclog the aspirator capillary or burner, thus producing inaccurat
10、eresults, are filtered through a 0.45-m membrane filter prior totesting.5. Significance and Use5.1 Although most potable supplies contain little strontium,some well waters in the midwestern part of the United Stateshave levels as high as 39 mg/L.35.2 This test method affords a reliable means of accu
11、ratelydetermining strontium and correcting calcium results obtainedby the methods cited in 6.2.6. Interferences6.1 Chemical interference caused by silicon, aluminum, andphosphate is controlled by adding lanthanum chloride. Potas-sium chloride is added to suppress the ionization of strontium.NOTE 1A
12、nitrous oxide-acetylene flame has been used successfullyby some to remove chemical interferences.6.2 Strontium chemically resembles calcium and causes apositive error in gravimetric and titrimetric methods for cal-cium determination.1This test method is under the jurisdiction of ASTM Committee D19 o
13、n Waterand is the direct responsibility of Subcommittee D19.05 on Inorganic Constituentsin Water.Current edition approved Aug. 1, 2007. Published September 2007. Originallyapproved in 1980. Last previous edition approved in 2002 as D 3920 02.2For referenced ASTM standards, visit the ASTM website, ww
14、w.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Standard Method for the Examination of Water and Wastewater, 14th Ed.American Public Health Assn., Washington, DC 2000
15、5.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.7. Apparatus7.1 Atomic Absorption Spectrophotometer for use at 460.7nm. A general guide for the use of flame atomic absorptionapplications is given in Practice D 4691.NOTE 2The manufa
16、cturers instructions should be followed for set-ting instrumental parameters.7.2 Strontium Hollow-Cathode Lamp.8. Reagents and Materials8.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specificat
17、ions of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are available.4Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.8.2 Pu
18、rity of WaterUnless otherwise indicated, referenceto water shall be understood to mean reagent water conformingto Specification D 1193, Type I, II, and III water. Type I ispreferred and more commonly used. Other reagent water typesmay be used provided it is first ascertained that the water is ofsuff
19、iciently high purity to permit its use without adverselyaffecting the precision and bias of the test method. Type IIwater was specified at the time of round robin testing of thistest method.NOTE 3The user must ensure the type of reagent water chosen issufficiently free of interferences. The water sh
20、ould be analyzed using thetest method.8.3 Hydrochloric Acid (sp gr 1.19)Concentrated hydro-chloric acid (HCl).8.4 Lanthanum Chloride/Potassium Chloride SolutionDissolve 11.73 g of lanthanum oxide (La2O3) in a minimumamount of concentrated hydrochloric acid (approximately 50mL.). Add 1.91 g of potass
21、ium chloride (KCl). Allow solutionto cool to room temperature and dilute to 100 mL with water.(WarningAdd acid slowly and in small portion to controlthe reaction rate upon mixing.)8.5 Nitric Acid (sp gr 1.42)Concentrated HNO3.8.6 Strontium Solution, Stock (1.0 mL = 1.0 mg Sr)Dissolve 2.415 g of stro
22、ntium nitrate (Sr(NO3)2) in a 1-L flaskcontaining 10 mL of concentrated HCl and 700 mL of water.Dilute to 1000 mL with water.8.7 Strontium Solution, Standard (1 mL = 0.010 mg Sr)Dilute 5.0 mL of strontium solution standard to 500 mL withwater.8.8 OxidantAir that has been cleaned and dried through as
23、uitable filter to remove oil, water, and other foreign sub-stances, is the usual oxidant.8.9 FuelStandard, commercially available acetylene isthe usual fuel. Acetone, always present in acetylene cylinders,can be prevented from entering the burner system by replacingthe cylinder when the pressure rea
24、ches 490 kPa (70 psig).(WarningPurified grade acetylene containing a special pro-prietary solvent other than acetone should not be used withpoly (vinyl chloride) tubing as weakening of the walls mayresult and cause a potentially hazardous situation.)9. Sampling9.1 Collect the samples in accordance w
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD392020022007E1STANDARDTESTMETHODFORSTRONTIUMINWATER 水中 含量 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-515809.html