1、Designation: D1512 15aD1512 15bStandard Test Methods forCarbon BlackpH Value1This standard is issued under the fixed designation D1512; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parenthe
2、ses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope1.1 These test methods, Test Method ABoiling Slurry and Test Method BSo
3、nic Slurry, are used to indicate the pH of thecarbon black surface by measuring the pH of water in contact with the carbon black.NOTE 1The pH of the carbon black is often used in this industry to indicate the relative acidity or alkalinity of carbon black and will be used in theremainder of these te
4、st methods to describe this property.NOTE 2Test Method A and Test Method B do not always give the same results.1.2 The values stated in SI units are to be regarded as the standard. The values in parentheses are for information only.1.3 This standard does not purport to address all of the safety conc
5、erns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterD1799
6、 Practice for Carbon BlackSampling Packaged ShipmentsD1900 Practice for Carbon BlackSampling Bulk ShipmentsD4483 Practice for Evaluating Precision for Test Method Standards in the Rubber and Carbon Black Manufacturing IndustriesE70 Test Method for pH of Aqueous Solutions With the Glass Electrode3. S
7、ignificance and Use3.1 The pH level of a carbon black is known to affect the vulcanization of some rubber compounds.TEST METHOD ABOILING SLURRY4. Apparatus4.1 pH Meter, (digital is recommended) having an accuracy of 60.05 pH and equipped with a combination electrode and RNCconnector.4.2 Container, s
8、tainless steel or copper, 125 cm3 or larger.4.3 Hot Plate.4.4 High Speed Mill, Mixer or Mortar and Pestle.4.5 Beakers, glass, 100 cm3 graduated with watch glasses.4.6 Magnetic Stir Plate.4.7 Magnetic Stir Bars.1 These test methods are under the jurisdiction ofASTM Committee D24 on Carbon Black and a
9、re the direct responsibility of Subcommittee D24.31 on Non-Carbon BlackComponents of Carbon Black.Current edition approved Sept. 1, 2015Dec. 1, 2015. Published December 2015February 2016. Originally approved in 1975. Last previous edition approved in 2015 asD1512 15.D1512 15a. DOI: 10.1520/D1512-15A
10、.10.1520/D1512-15B.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and
11、is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only th
12、e current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Reagents5.1 Purity of ReagentsReagent grade chemicals shall be used in all tests. Unles
13、s otherwise indicated, it is intended that allreagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, wheresuch specifications are available.3 Other grades may be used, provided it is first ascertained that the reagent is of sufficiently
14、 highpurity to permit its use without lessening the accuracy of the determination.5.2 Purity of WaterUnless otherwise indicated, references to water shall be understood to mean reagent water as defined byType 1 in Specification D1193.5.3 Distilled Water, high purity.5.4 Buffer Solutions, pH of 4.00,
15、 7.00, and 10.00.5.5 Acetone, reagent grade.6. Sampling6.1 Samples shall be taken in accordance with Practices D1799 or D1900.7. Calibration7.1 Calibrate the pH meter using buffer solutions according to manufacturers instructions.8. Procedure8.1 Pulverize pelleted or lumpy carbon black to a fine pow
16、der, using either the high speed mixer or mortar and pestle.8.2 Weigh 5 g of carbon black into a 100 cmPrepare boiling distilled water in a stainless steel3 glass beaker.8.3 Add 50 cmWeigh carbon black3 of boiling,and boiling distilled water prepared in a stainless steel beaker and into a glassbeake
17、r or boiling flask and add 2 to 3 drops of acetone to facilitate wetting of the sample. The carbon black weight and volumeof water is maintained at a constant ratio of 1:10. See Table 1 for selection of carbon black weight, volume of water and beakeror boiling flask volume.NOTE 3A stainless steel be
18、aker is used to eliminate contamination during boiling.8.4 Cover the glass beaker with a watch glass and boil the mixture for 15 min, but do not allow all the liquid to evaporate.8.5 Let the mixture cool to room temperature in an atmosphere free from chemical fumes which might contaminate the sample
19、.8.6 Standardize the pH meter with the buffer solutions. Rinse the electrode with distilled water and wipe clean after each test.8.7 Place a magnetic stir bar into the glass beaker and place on magnetic stir plate (or similar mechanical stirring device) andadjust the stir speed to achieve a continuo
20、us uniform slurry. Carefully place the pH electrode into the slurry taking care not to allowthe electrode to contact the stir bar. Once a constant pH is obtained record to the nearest 0.05 unit.NOTE 4Refer to Test Method E70 for a definition of pH and a highly detailed procedure for making pH measur
21、ements.8.8 Rinse the electrode with distilled water and wipe clean. Keep the electrode soaking in distilled water when not in use.9. Report9.1 Report the following information:9.1.1 Proper identification of the sample,9.1.2 Result obtained, reported to the nearest 0.05 unit, and9.1.3 Test Method use
22、d, A or B.3 Reagent Chemicals, American Chemical Society Specifications, American Chemical Society, Washington, DC. For suggestions on the testing of reagents not listed bythe American Chemical Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United Stat
23、es Pharmacopeia and NationalFormulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD.TABLE 12 Precision Parameters for ASTM D1512, Carbon Black - pH Value, Method A, (Type 1 Precision)Units pH UnitsMaterial Period Number of Laboratories Mean Level Sr r SR RHS Tread Fall 2003 22 6.2 0.15
24、0.43 0.73 2.07SRB A6 (N134) Fall 2004 23 6.5 0.15 0.41 0.53 1.49LS Carcass Spring 2004 26 7.2 0.14 0.39 0.50 1.41SRB C6 (N326) Spring 2003 27 8.5 0.11 0.32 0.50 1.41N774 Fall 2002 19 8.6 0.16 0.45 0.62 1.75Average 7.4Pooled Values 0.14 0.40 0.58 1.65D1512 15b210. Precision and Bias10.1 Test Method A
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD15122015BSTANDARDTESTMETHODSFORCARBONBLACKX2014PHVALUE 碳黑 标准 试验 方法 PH PDF

链接地址:http://www.mydoc123.com/p-509950.html