ASTM C1689-2008 Standard Practice for Subsampling of Uranium Hexafluoride.pdf
《ASTM C1689-2008 Standard Practice for Subsampling of Uranium Hexafluoride.pdf》由会员分享,可在线阅读,更多相关《ASTM C1689-2008 Standard Practice for Subsampling of Uranium Hexafluoride.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1689 08Standard Practice forSubsampling of Uranium Hexafluoride1This standard is issued under the fixed designation C 1689; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in paren
2、theses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice is applicable to subsampling uraniumhexafluoride (UF6), using heat liquefaction techniques, frombulk containers, obtained in conformance
3、 with C 1052, intosmaller sample containers, which are required for laboratoryanalyses.1.2 It is assumed that the liquid UF6being sampled com-prises a single quality and quantity of material. This practicedoes not address any special additional arrangement that mightbe required for taking proportion
4、al or composite samples.1.3 The number of samples to be taken, their nominalsample weight, and their disposition shall be agreed uponbetween the parties.1.4 The scope of this practice does not include provisionsfor preventing criticality incidents.1.5 This standard does not purport to address all of
5、 thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C 787 Specification for
6、Uranium Hexafluoride for Enrich-mentC 859 Terminology Relating to Nuclear Materials3C 996 Specification for Uranium Hexafluoride Enriched toLess Than 5 %235UC 761 Test Methods for Chemical, Mass Spectrometric,Spectrochemical, Nuclear, and Radiochemical Analysis ofUranium HexafluorideC 1052 Practice
7、for Bulk Sampling of Liquid UraniumHexafluoride2.2 Other Documents:ANSI N14.1 Uranium Hexafluoride: Packaging for Trans-port4ISO/DIS 7195 Packaging of Uranium Hexafluoride (UF6)for Transport5USEC-651 The UF6Manual: Good Handling Practices forUranium Hexafluoride, latest revision63. Terminology3.1 Te
8、rms shall be defined in accordance with TerminologyC 859 except for the following:3.1.1 sample bottlethe vessel (typically a 1S or 2S bottle)into which the sample of UF6is withdrawn from the containerfor transfer to the laboratory, analysis or dispatch to thecustomer.3.1.2 subsample tubethe small ve
9、ssel (for example, a P10tube7) into which a subsample of UF6is withdrawn from thesample bottle for analysis of UF6quality or dispatch to thecustomer.3.1.3 subsample rigthe equipment to perform the transferof liquid UF6from the sample bottle into the subsample tube,typically a vacuum manifold equippe
10、d with heating and aliquid nitrogen trap.4. Summary of Practice4.1 Two methods of withdrawing a subsample of UF6aredescribed which differ based on safety requirements namely:(1) homogenizing of liquefied UF6by agitation before liquidtransfer and (2) homogenizing of liquefied UF6by convectionbefore l
11、iquid transfer. The first method involves homogeniza-tion of liquified UF6in sample bottle by vigorous shaking.Subsequently the sample bottle is inverted and connected to thetop of a heated vacuum-manifold system, and the subsample1This practice is under the jurisdiction of ASTM Committee C26 on Nuc
12、learFuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved Aug. 1, 2008. Published September 2008.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMSt
13、andards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.5Available from International Organization for Standardization (ISO),
14、1, ch. dela Voie-Creuse, Case postale 56, CH-1211, Geneva 20, Switzerland, http:/www.iso.ch.6Available from United States Enrichment Corp., 6903 Rockledge Dr., Bethesda,MD 20817, http:/.7Polychlorotrifluoroethylene P10 tubes are widely accepted by the industry forsubsample collection and subsequent
15、UF6quality analyses or dispatch to thecustomer. Other types of subsample tubes, for example P-20, P-80 or P100 , can beused for internal subsample collection and processing. Dispatch of these subsampletubes may be agreed upon by buyer and seller and subject to (local) transportregulations.1Copyright
16、 ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.tube is attached to the appropriate port of the system. Thesystem is evacuated and the liquid UF6allowed to flow bygravity into the subsample tube. In the second method thesample bottle containin
17、g solid UF6is connected to the top of amanifold system, and a subsample tube is attached to theappropriate port of the system. The whole system is enclosed insecondary containment that can be heated (hot-box). Afterevacuation the complete system is heated for specific period(typically 1.5 hr) to all
18、ow for complete homogenization ofthe liquid UF6by convection. Subsequently the liquid UF6isallowed to flow by gravity either directly or via graduatedvolume into the subsample tube.4.2 For both methods of sampling, the presence of residuesmay have significant implications for the quality of UF6. For
19、safety and quality reasons, sample bottles and an subsampletubes shall be clean, dry, and empty before filling.4.3 Various types of sample bottles and tubes are in use andare described in detail in the applicable national and interna-tional standards, for example, ANSI N14.1 and ISO/DIS 7195.For a g
20、iven type of sample bottle, the detailed configuration,for example valve orientation, terminal fittings and the like,may vary. Hence the type and configuration of bottles used forthe withdrawal of samples shall be agreed upon between theparties.5. Significance and Use5.1 Uranium hexafluoride is norm
21、ally produced and handledin large (typically 1- to 14-ton) quantities and must, thereforebe characterized by reference to representative samples. Thequantities involved, physical properties, chemical reactivity,and hazardous nature of UF6are such that for representativesampling, specially designated
22、 equipment must be used andoperated in accordance with the most carefully controlled andstringent procedures. This practice indicates appropriate prin-ciples, equipment and procedures currently in use for subsam-pling of liquid UF6. It is used by UF6converters, enrichers andfuel fabricators to revie
23、w the effectiveness of existing proce-dures or to design equipment and procedures for future use.Other subsampling procedures such as UF6vapor sampling arenot directly representative of the chemical quality of liquidUF6.5.2 It is emphasized that this test guide is not meant toaddress conventional or
24、 nuclear criticality safety issues.6. Apparatus6.1 Hot Water Bath.6.2 Subsample RigFor Procedure 1 see Fig. 1 and Proce-dure 2 see Fig. 2. Materials of construction in direct contactwith liquid UF6are made from nickel, high nickel alloys, ormaterials having comparable resistance to UF6corrosion.6.3
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC16892008STANDARDPRACTICEFORSUBSAMPLINGOFURANIUMHEXAFLUORIDEPDF

链接地址:http://www.mydoc123.com/p-465529.html