ASTM F448-2011 Test Method for Measuring Steady-State Primary Photocurrent《测量稳态初级光电流的标准试验方法》.pdf
《ASTM F448-2011 Test Method for Measuring Steady-State Primary Photocurrent《测量稳态初级光电流的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F448-2011 Test Method for Measuring Steady-State Primary Photocurrent《测量稳态初级光电流的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F448 11Standard Test Method forMeasuring Steady-State Primary Photocurrent1This standard is issued under the fixed designation F448; the number immediately following the designation indicates the year of originaladoption or, in the case of revision, the year of last revision.Anumber in
2、parentheses indicates the year of last reapproval.Asuperscriptepsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the measurement of steady-stateprimary photocurrent, Ipp, generated in semiconductor deviceswhen these devices are exposed
3、to ionizing radiation. Theseprocedures are intended for the measurement of photocurrentsgreater than 109As/Gy(Si or Ge), in cases for which therelaxation time of the device being measured is less than 25 %of the pulse width of the ionizing source. The validity of theseprocedures for ionizing dose ra
4、tes as great as 108Gy(Si or Ge)/shas been established. The procedures may be used for mea-surements at dose rates as great as 1010Gy(Si or Ge)/s;however, extra care must be taken. Above 108Gy/s, thepackage response may dominate the device response for anydevice. Additional precautions are also requi
5、red when measur-ing photocurrents of 109As/Gy(Si or Ge) or lower.1.2 Setup, calibration, and test circuit evaluation proceduresare also included in this test method.1.3 Because of the variability between device types and inthe requirements of different applications, the dose rate rangeover which any
6、 specific test is to be conducted is not given inthis test method but must be specified separately.1.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, if a
7、ny, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E668 Practice for Application of Thermoluminescen
8、ce-Dosimetry (TLD) Systems for DeterminingAbsorbed Dosein Radiation-Hardness Testing of Electronic DevicesF526 Test Method for Using Calorimeters for Total DoseMeasurements in Pulsed LinearAccelerator or Flash X-rayMachines3. Terminology3.1 Definitions:3.1.1 fall time, nthe time required for a signa
9、l pulse todrop from 90 to 10 % of its steady-state value.3.1.2 photocurrent relaxation time, nthe time required forthe radiation induced photocurrent to decrease to 1/e (0.368) ofits initial value. The relaxation time depends upon therecombination-controlled photocurrent decay in the media,which is
10、often a semiconductor. The relaxation time candepend upon the temperature and the strength of theirradiation/illumination.3.1.3 primary photocurrent, nthe flow of excess chargecarriers across a p-n junction due to ionizing radiation creatingelectron-hole pairs throughout the device. The charges asso
11、ci-ated with this current are only those produced in the junctiondepletion region and in the bulk semiconductor materialapproximately one diffusion length on either side of thedepletion region (or to the end of the semiconductor material,whichever is shorter).3.1.4 pulse width, nthe time a pulse-amp
12、litude remainsabove 50 % of its maximum value.3.1.5 rise time, nthe time required for a signal pulse torise from 10 to 90 % of its steady-state value.4. Summary of Test Method4.1 In this test method, the test device is irradiated in theprimary electron beam of a linear accelerator. Both the irradia-
13、tion pulse and junction current (Fig. 1) are displayed andrecorded. Placement of a thin, low atomic number (Z#13)scattering plate in the beam is recommended to improve beamuniformity; the consequences of the use of a scattering platerelating to interference from secondary electrons are described.The
14、 total dose is measured by an auxiliary dosimeter. Thesteady-state values of the dose rate and junction current and therelaxation time of the junction current are determined from thedata trace and total dose.4.2 In special cases, these parameters may be measured at asingle dose rate under one bias c
15、ondition if the test is designedto generate information for such a narrow application. The1This test method is under the jurisdiction of ASTM Committee F01 onElectronics and is the direct responsibility of Subcommittee F01.11 on Nuclear andSpace Radiation Effects.Current edition approved June 1, 201
16、1. Published July 2011. Originally approvedin 1975 as F448 75 T. Last previous edition approved in 2005 as F448 99(2005).DOI: 10.1520/F0448-11.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards vo
17、lume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.preferred approach, described in this test method, is to char-acterize the radiation response of a devi
18、ce in a way that isuseful to many different applications. For this purpose, theresponse to pulses at a number of different dose rates isrequired. Because of the bias dependence of the depletionvolume, it is possible that more than one bias level will berequired during the photocurrent measurements.5
19、. Significance and Use5.1 PN Junction DiodeThe steady-state photocurrent of asimple p-n junction diode is a directly measurable quantity thatcan be directly related to device response over a wide range ofionizing radiation. For more complex devices the junctionphotocurrent may not be directly relate
20、d to device response.5.2 Zener Diode In this device, the effect of the photo-current on the Zener voltage rather than the photocurrent itselfis usually most important. The device is most appropriatelytested while biased in the Zener region. In testing Zener diodesor precision voltage regulators, ext
21、ra precaution must be takento make certain the photocurrent generated in the device duringirradiations does not cause the voltage across the device tochange during the test.5.3 Bipolar TransistorAs device geometries dictate thatphotocurrent from the base-collector junction be much greaterthan curren
22、t from the base-emitter junction, measurements areusually made only on the collector-base junction with emitteropen; however, sometimes, to obtain data for computer-aidedcircuit analysis, the emitter-base junction photocurrent is alsomeasured.5.4 Junction Field-Effect DeviceA proper photocurrentmeas
23、urement requires that the source be shorted (dc) to thedrain during measurement of the gate-channel photocurrent. Intetrode-connected devices, the two gate-channel junctionsshould be monitored separately.5.5 Insulated Gate Field-Effect DeviceIn this type ofdevice, the true photocurrent is between th
24、e substrate and thechannel, source, and drain regions. A current which cangenerate voltage that will turn on the device may be measuredby the technique used here, but it is due to induced conduc-tivity in the gate insulator and thus is not a junction photocur-rent.6. Interferences6.1 Air Ionization
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF4482011TESTMETHODFORMEASURINGSTEADYSTATEPRIMARYPHOTOCURRENT 测量 稳态 初级 电流 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-536688.html