UOP 991-2017 Trace Chloride Fluoride and Bromide in Liquid Organics by Combustion Ion Chromatography (CIC).pdf
《UOP 991-2017 Trace Chloride Fluoride and Bromide in Liquid Organics by Combustion Ion Chromatography (CIC).pdf》由会员分享,可在线阅读,更多相关《UOP 991-2017 Trace Chloride Fluoride and Bromide in Liquid Organics by Combustion Ion Chromatography (CIC).pdf(14页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2011, 2013, 2017 UOP LLC. All rights reserved. N
3、onconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.95
4、85 PHONE. Trace Chloride, Fluoride, and Bromide in Liquid Organics by Combustion Ion Chromatography (CIC) UOP Method 991-17 Scope This method is for determining trace concentrations of chloride, fluoride, and bromide in liquid organics by Combustion Ion Chromatography (CIC). This method has a lower
5、limit of quantitation of 0.1 mg/kg (mass-ppm) for fluoride and chloride, and 0.2 mg/kg for bromide. References Instruction Manual for Automatic Quick Furnace, Model AQF-2100H, www.cosa- UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method A reproducible volume of sam
6、ple is injected into a quartz boat. The boat is introduced into a multi-position horizontal furnace under controlled temperature and moisture. The sample is then combusted in a pyrohydrolytic, oxygen enriched environment. The halogens in the sample are converted to hydrogen halides and halogen gas.
7、These gases are then absorbed into an aqueous solution of constant volume. An aliquot of the solution is injected into an ion chromatograph (IC) where chloride, fluoride, and bromide are separated by anion exchange and measured by a conductivity detector with ion suppression. External standards are
8、used for quantitation. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, semi-micro, capacity 220 g, readable to 0.00001 g at 81 g Chromatographic column, IonPac AS11-HC Analytical, 250-mm length by 4-mm ID,
9、Dionex, Thermo Fisher, Cat. No. 052960 Combustion system, equipped with AQF-2100H furnace (HF-210 horizontal furnace), GA-210 gas adsorption unit, ABC-210 automatic boat controller, ASC-250L liquid autosampler, and AQF Software, COSA Instrument Corp. Newer models from the same manufacturer are also
10、suitable. Deionized water system, Barnstead GenPure xCAD Plus UV TOC (bench version), Thermo Scientific, ThermoFisher Cat. No. 50136153 Dry block heater, digital, 2 block capacity, 120V, 50/60Hz, 210W, VWR, Cat. No. 12621-088, and heating blocks with 15- and 28-mm diameter vial wells, Cat. Nos. 1262
11、1-126 and -138, respectively 2 of 14 991-17 Flask, volumetric, 1-L, borosilicate glass, Class A, VWR, Cat No. 89000-412 Ion chromatograph, equipped with a pump, injection valve/autosampler, eluent generator, continuously regenerated trap column, suppressor, conductivity detector, computer, and Chrom
12、eleon software, Dionex, Thermo Fisher, Model ICS-2100 Muffle furnace, capable of operation at 800C, Barnstead/Thermolyne compact benchtop muffle furnace, Thermo Scientific, VWR Cat. No. 30631-230 Pipettor, Eppendorf Reference single-channel, variable volume, 100-1000 mL, VWR, Cat. No. 10032-682 Refr
13、igerator, laboratory, explosion proof or flammable storage, Fisher Scientific, Cat. No. 05EREETSA Regulator, argon, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590 Regulator, oxygen, two-stage, high purity, delivery pressure range 30-700 kPa (
14、4-100 psi), cleaned for oxygen service, Matheson Tri-Gas, Model 3810-540 Ultrasonic bath, Electron Microscopy Sciences Model 3510-MT, VWR, Cat. No. 100493-068 Vortex mixer, digital, 120V, 30W, VWR, Cat. No. 97043-564 Reagents and Materials References to catalog numbers and suppliers are included as
15、a convenience to the method user. Other suppliers may be used. References to water mean deionized and distilled water that is subsequently treated to produce ionically pure, 18.2 megaohm-cm, organic-free (99.9%, VWR, Cat. No. BJ010 Adsorption tube, 10-mL, COSA Instrument Corp., Cat. No. MC25000 (rep
16、lacement) Argon, 99.999% pure Ball joint with U-shaped tube, COSA Instrument Corp., Cat. No. MC28017 (replacement) 4-Bromoacetanilide, 98% pure, Sigma-Aldrich, Cat No. 161659 4-Fluorobenzoic acid, 99+% pure, Sigma-Aldrich, Cat. No. 418846 Methanol, B then draw 80 L of sample/standard into the syring
17、e. 4. Place syringe on balance and weigh syringe. Record the weight to nearest 0.00001 g. For a sample that must be heated, weigh the heated syringe with sample and immediately dispense the contents. 5. Enter the weight in the chromatography software under the weight column for the associated sample
18、/standard. 6. Dispense weighed liquid into the waste bottle of the autosampler. 7. Rinse syringe with 3 syringe volumes of toluene. a. Repeat Steps 2-6 for all samples and standards. b. Proceed to Step 8 when all standards and samples have been weighed. 8. Return the clean syringe to the autosampler
19、 and proceed to Step 3 of Liquid Injection for liquid samples or Step 5 of Heated Sample Injection for viscous or waxy samples to start the analysis. Analysis of Standard Calibration Solutions 1. Fill sample vials with all four Standard Calibration Solutions and load them onto the liquid autosampler
20、. 2. Inject 80 L of Standard Calibration Solution 1 into the quartz boat of the automatic boat controller (ABC) and start the combustion and IC sequences. Identify and integrate each peak for the calibration standards. Identification can be aided by comparing to the typical standard chromatogram (se
21、e Figure). Individual standards may need to be prepared to identify retention times. Cumulative multiple injections in the absorption solution increase the effective concentration of the analytes of interest. For low concentrations, from 0.1-10 mg/kg, it is required to run 3 cumulative injections to
22、 increase the response of fluoride, chloride, and bromide. For concentrations from 10-1000 mg/kg, a single injection is sufficient. 3. Calculate the response factor for each analyte using the chromatography software or Equation 3 in Calculations. 4. Repeat Steps 3 and 4 with Standard Calibration Sol
23、utions 2, 3, and 4. Analysis of Blank Blank runs may be needed to check for interference or carryover from a previous analysis. 1. Start the combustion and IC process with an empty sample boat under the same burn program as the standards and/or samples. 2. Identify and integrate peak areas of intere
24、st. 3. If background peaks are higher than the lowest standard, run the default boat prebake method 10 of 14 991-17 five times and re-analyze another empty sample boat blank injection. If background is still high, remove the old quartz boat. Replace with a new quartz boat and quartz wool. Run five b
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9912017TRACECHLORIDEFLUORIDEANDBROMIDEINLIQUIDORGANICSBYCOMBUSTIONIONCHROMATOGRAPHYCICPDF

链接地址:http://www.mydoc123.com/p-1074608.html