ASTM F448-2018 Standard Test Method for Measuring Steady-State Primary Photocurrent《测量稳态初级光电流的标准试验方法》.pdf
《ASTM F448-2018 Standard Test Method for Measuring Steady-State Primary Photocurrent《测量稳态初级光电流的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F448-2018 Standard Test Method for Measuring Steady-State Primary Photocurrent《测量稳态初级光电流的标准试验方法》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F448 11F448 18Standard Test Method forMeasuring Steady-State Primary Photocurrent1This standard is issued under the fixed designation F448; the number immediately following the designation indicates the year of originaladoption or, in the case of revision, the year of last revision.Anum
2、ber in parentheses indicates the year of last reapproval.Asuperscriptepsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope1.1 This test method covers the measurement of steady-stat
3、e primary photocurrent, Ipp, generated in semiconductor deviceswhen these devices are exposed to ionizing radiation. These procedures are intended for the measurement of photocurrents greaterthan 109As/Gy(Si or Ge), in cases for which the relaxation time of the device being measured is less than 25
4、% of the pulse widthof the ionizing source. The validity of these procedures for ionizing dose rates as great as 108Gy(Si or Ge)/s has been established.The procedures may be used for measurements at dose rates as great as 1010Gy(Si or Ge)/s; however, extra care must be taken.Above 108Gy/s, the packa
5、ge response may dominate the device response for any device. Additional precautions are also requiredwhen measuring photocurrents of 109 As/Gy(Si or Ge) or lower.1.2 Setup, calibration, and test circuit evaluation procedures are also included in this test method.1.3 Because of the variability betwee
6、n device types and in the requirements of different applications, the dose rate range overwhich any specific test is to be conducted is not given in this test method but must be specified separately.1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are i
7、ncluded in this standard.1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine theapplicability of
8、 regulatory limitations prior to use.1.6 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade
9、Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E668 Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining Absorbed Dose inRadiation-Hardness Testing of Electronic DevicesF526 Test Method for Using Calorimeters for
10、Total Dose Measurements in Pulsed Linear Accelerator or Flash X-ray Machines3. Terminology3.1 Definitions:3.1.1 fall time, nthe time required for a signal pulse to drop from 90 to 10 % of its steady-state value.3.1.2 photocurrent relaxation time, nthe time required for the radiation induced photocur
11、rent to decrease to 1/e (0.368) of itsinitial value. The relaxation time depends upon the recombination-controlled photocurrent decay in the media, which is often asemiconductor. The relaxation time can depend upon the temperature and the strength of the irradiation/illumination.3.1.3 primary photoc
12、urrent, nthe flow of excess charge carriers across a p-n junction due to ionizing radiation creatingelectron-hole pairs throughout the device. The charges associated with this current are only those produced in the junctiondepletion region and in the bulk semiconductor material approximately one dif
13、fusion length on either side of the depletion region(or to the end of the semiconductor material, whichever is shorter).1 This test method is under the jurisdiction of ASTM Committee F01 on Electronics and is the direct responsibility of Subcommittee F01.11 on Nuclear and SpaceRadiation Effects.Curr
14、ent edition approved June 1, 2011March 1, 2018. Published July 2011April 2018. Originally approved in 1975 as F448 75 T. Last previous edition approved in20052011 as F448 99F448 11.(2005). DOI: 10.1520/F0448-11.10.1520/F0448-18.2 For referencedASTM standards, visit theASTM website, www.astm.org, or
15、contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have
16、 been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official docu
17、ment.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.1.4 pulse width, nthe time a pulse-amplitude remains above 50 % of its maximum value.3.1.5 rise time, nthe time required for a signal pulse to rise from 10 to 90 % of its steady-s
18、tate value.4. Summary of Test Method4.1 In this test method, the test device is irradiated in the primary electron beam of a linear accelerator. Both the irradiation pulseand junction current (Fig. 1) are displayed and recorded. Placement of a thin, low atomic number (Z13) scattering plate in thebea
19、m is recommended to improve beam uniformity; the consequences of the use of a scattering plate relating to interference fromsecondary electrons are described. The total dose is measured by an auxiliary dosimeter. The steady-state values of the dose rateand junction current and the relaxation time of
20、 the junction current are determined from the data trace and total dose.4.2 In special cases, these parameters may be measured at a single dose rate under one bias condition if the test is designed togenerate information for such a narrow application. The preferred approach, described in this test m
21、ethod, is to characterize theradiation response of a device in a way that is useful to many different applications. For this purpose, the response to pulses at anumber of different dose rates is required. Because of the bias dependence of the depletion volume, it is possible that more thanone bias l
22、evel will be required during the photocurrent measurements.5. Significance and Use5.1 PN Junction DiodeThe steady-state photocurrent of a simple p-n junction diode is a directly measurable quantity that canbe directly related to device response over a wide range of ionizing radiation. For more compl
23、ex devices the junction photocurrentmay not be directly related to device response.5.2 Zener Diode In this device, the effect of the photocurrent on the Zener voltage rather than the photocurrent itself is usuallymost important. The device is most appropriately tested while biased in the Zener regio
24、n. In testing Zener diodes or precisionvoltage regulators, extra precaution must be taken to make certain the photocurrent generated in the device during irradiations doesnot cause the voltage across the device to change during the test.5.3 Bipolar TransistorAs device geometries dictate that photocu
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF4482018STANDARDTESTMETHODFORMEASURINGSTEADYSTATEPRIMARYPHOTOCURRENT 测量 稳态 初级 电流 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-536689.html