ASTM F2182-2011 Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging《测定在磁共振成像中靠近被植入物导致发热的无线电频率的标准试验方法》.pdf
《ASTM F2182-2011 Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging《测定在磁共振成像中靠近被植入物导致发热的无线电频率的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2182-2011 Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging《测定在磁共振成像中靠近被植入物导致发热的无线电频率的标准试验方法》.pdf(15页珍藏版)》请在麦多课文档分享上搜索。
1、Designation:F218209 Designation: F2182 11Standard Test Method forMeasurement of Radio Frequency Induced Heating On orNear Passive Implants During Magnetic ResonanceImaging1This standard is issued under the fixed designation F2182; the number immediately following the designation indicates the year o
2、foriginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers measurement of radio frequency (RF
3、) induced heating on or near a passive medical implant andits surroundings during magnetic resonance imaging (MRI).1.2 This test method is one of those required to determine if the presence of a passive implant may cause injury to the patientwith the implant during an MR procedure. Other safety issu
4、es that should be addressed include magnetically induced displacementforce and torque.1.3 The amount of RF-induced temperature rise for a given specific absorption rate (SAR) will depend on the RF frequency,which is dependent on the static magnetic field strength of the MR system. Because of possibl
5、e additional heating, particularlywhen implant dimensions approaches or exceeds one quarter of the wavelength of the RF field inside the phantom, conclusionsfrom measurements made at one static magnetic field strength do not apply to other field strengths and frequencies. While the focusin this test
6、 method is on 1.5 T or 3 Tesla cylindrical bore MR systems, the RF-induced temperature rise for an implant in openMR systems can be evaluated by suitable modification of the method described herein.1.4 This test method assumes that testing is done on devices that will be entirely inside the body. Fo
7、r other implantationconditions (for example, external fixation devices, percutaneous needles, catheters or tethered devices such as ablation probes),modifications of this test method are necessary.1.5 This test method applies to whole body magnetic resonance equipment, as defined in section 2.2.103
8、of the IEC Standard60601-2-33, Ed. 2.0, with a whole body RF transmit coil as defined in section 2.2.100. The RF coil is assumed to have quadratureexcitation.1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.7 This standar
9、d does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM S
10、tandards:2F2052 Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the MagneticResonance EnvironmentF2119 Test Method for Evaluation of MR Image Artifacts from Passive ImplantsF2213 Test Method for Measurement of Magnetically Induced Torque on Medical Device
11、s in the Magnetic ResonanceEnvironmentF2503 Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment2.2 IEC Standard:360601-2-33, Ed. 2.0 Medical Electrical EquipmentPart 2: Particular Requirements for the Safety of Magnetic ResonanceEquipment for Medical
12、 Diagnosis, 20021This test method is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved Nov. 15, 2009. Published January 2010. Originally approved in 2002.
13、 Last previous edition approved in 2002 as F218202a. DOI:10.1520/F2182-09.Current edition approved March 1, 2011. Published March 2011. Originally approved in 2002. Last previous edition approved in 2009 as F2182 09. DOI:10.1520/F2182-11.2For referenced ASTM standards, visit the ASTM website, www.as
14、tm.org, or contact ASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3Available from the International Electrotechnical Commission (IEC), 3 rue de Varembe, Case postale 131, CH-1211 Geneva 2
15、0, Switzerland.1This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users c
16、onsult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.2.3 NEMA Standard:4NEMA MS 82008
17、 Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems3. Terminology3.1 Definitions:3.1.1 gelled salinephantom medium consisting of sodium chloride and polyacrylic acid or sodium chloride and hydroxy-ethylcellulose in water as specified in this test method.3.1.2 imp
18、lant, nin medicine, an object, structure, or device intended to reside within the body for diagnostic, prosthetic, orother therapeutic purposes.3.1.3 isocentergeometric center of the gradient coil system, which generally is the geometric center of a scanner with acylindrical bore.3.1.33.1.4 local SA
19、Rspecific absorption rate (SAR) averaged over any 10 g of tissue of the patient body and over a specified time.60601-2-33, Ed. 2.03.1.43.1.5 magnetic resonance (MR) environmentvolume within the 0.50 mT (5 gauss (G) line of an MR system, which includesthe entire three dimensional volume of space surr
20、ounding the MR scanner. For cases where the 0.50 mT line is contained withinthe Faraday shielded volume, the entire room shall be considered the MR environment.3.1.53.1.6 magnetic resonance imaging (MRI)imaging technique that uses static and time varying magnetic fields to provideimages of tissue by
21、 the magnetic resonance of nuclei.3.1.63.1.7 magnetic resonance system (MR system)ensemble of MR equipment, accessories including means for display, control,energy supplies, and the MR environment.60601-2-33, Ed. 2.03.1.7medical implanta structure or device that is placed within the body of the pati
22、ent for medical diagnostic or therapeuticpurposes.3.1.8 MR Conditionalan item that has been demonstrated to pose no known hazards in a specified MR environment withspecified conditions of use. Field conditions that define the specified MR environment include field strength, spatial gradient, dB/dt(t
23、ime rate of change of the magnetic field), radio frequency (RF) fields, and specific absorption rate (SAR). Additional conditions,including specific configurations of the item, may be required.3.1.9 MR Safean item that poses no known hazards in all MR environments.NOTE 1MR Safe items include noncond
24、ucting, nonmagnetic items such as a plastic petri dish. An item may be determined to be MR Safe byproviding a scientifically based rationale rather than test data.3.1.10 MR test systemMR system or an apparatus that reproduces the RF field of this type of system.3.1.11 MR Unsafean item that is known
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF21822011STANDARDTESTMETHODFORMEASUREMENTOFRADIOFREQUENCYINDUCEDHEATINGNEARPASSIVEIMPLANTSDURINGMAGNETICRESONANCEIMAGING

链接地址:http://www.mydoc123.com/p-537984.html