ASTM F1840-2010e1 Standard Terminology for Surgical Suture Needles《外科手术缝合针的标准术语》.pdf
《ASTM F1840-2010e1 Standard Terminology for Surgical Suture Needles《外科手术缝合针的标准术语》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1840-2010e1 Standard Terminology for Surgical Suture Needles《外科手术缝合针的标准术语》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F1840 101Standard Terminology forSurgical Suture Needles1This standard is issued under the fixed designation F1840; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses ind
2、icates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTE“Typical Curvatures” figure was editorially included in May 2011.1. Scope1.1 This terminology covers general definitions for surgicalneedles.2. Terminology2.1 Definiti
3、ons of Terms Specific to the Instrument (See Fig.1):attachment area, nportion of the needle where the attach-ment of the suture takes place. For example, eyed, drilled,and channel.body, ncentral portion of the needle intended to be graspedby the needle holder.chord length, nthe straight line distanc
4、e between the twoends of a curved needle.curvature, nthe shape of the needle viewed in profile. Somecommon shapes include, but are not limited to: straight,12curve or “ski”,18 circle,14 circle,38 circle,12 circle,58circle, and compound curvature (see Fig. 2).cutting egde, ncutting edges are made of
5、various goemetricshapes, that is, triangular, diamond, and hexagonal. Thevarious edges may be sharpened by the manufacturer de-pending on the user performance.needle length, nthe distance measured along the needlecurvature from end to end.needle radius, nthe radius of the uniformly curved portionor
6、portions of the needle measured from the centerline of theneedle body.needle wire diameter, nthe gage or thickness of the needlewire, measured at a location between the needle body andthe attachment area, where either no or minimal work hastaken place.point, nportion of the needle intended to initia
7、te tissuepenetration.point configuration, nthe shape of the point. Some commonpoint configurations include, but are not limited to (see Fig.3): taper, trocar, blunt, spatulated, conventional cutting edge,reverse cutting edge, cutting taper, and side cutting needle.swage, nthe term used to describe a
8、ny attachment methodthat uses mechanical force to crimp the end of the needle andfirmly hold the suture in place2.2 Definitions of Terms Specific to Mechanical Propertiesof a Surgical Needle:finish, nfinal surface visual appearance classified as follows:bright or mirror finish, nhighly reflective su
9、rface.satin, matte, or black finish, nreduced reflective surface(as compared to bright or mirror finish) varying from a dullappearance to a blackened surface.maximum bend momentthe greatest moment applied to aneedle during a bend test.needle ducilitya measure of the amount of plastic bending aneedle
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF18402010E1STANDARDTERMINOLOGYFORSURGICALSUTURENEEDLES 外科手术 缝合 标准 术语 PDF

链接地址:http://www.mydoc123.com/p-535816.html