ASTM E1363-2008 Standard Test Method for Temperature Calibration of Thermomechanical Analyzers《热机械分析仪温度校准的标准试验方法》.pdf
《ASTM E1363-2008 Standard Test Method for Temperature Calibration of Thermomechanical Analyzers《热机械分析仪温度校准的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1363-2008 Standard Test Method for Temperature Calibration of Thermomechanical Analyzers《热机械分析仪温度校准的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1363 08Standard Test Method forTemperature Calibration of Thermomechanical Analyzers1This standard is issued under the fixed designation E 1363; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisi
2、on. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the temperature calibrationof thermomechanical analyzers from 50 to 1100 C. (SeeNote 1.)1.2 The values
3、 stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard is similar to ISO 113591 but addresses alarger temperature range and utilizes additional calibrationmaterials.1.4 This standard does not purport to address all of thesafety
4、 concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific precau-tionary statements are given in Section 7 and Note 9.2. Refe
5、renced Documents2.1 ASTM Standards:2E 473 Terminology Relating to Thermal Analysis and Rhe-ology2.2 Other Standards:113591 Thermomechanical Analysis (TMA)-Part 1: Gen-eral Principles33. Terminology3.1 Definitions:3.1.1 The terminology relating to thermal analysis appear-ing in E 473 shall be conside
6、red applicable to this document.4. Summary of Test Method4.1 An equation is developed for the linear correlation of theexperimentally observed program temperature and the actualmelting temperature for known melting standards. This isaccomplished through the use of a thermomechanical analyzerwith a p
7、enetration probe to obtain the onset temperatures fortwo melting point standards.An alternate, one-point method oftemperature calibration, is also given for use over very narrowtemperature ranges. (See Note 2.)NOTE 1This test method may be used for calibrating thermomechani-cal analyzers at temperat
8、ures outside this range of temperature. However,the accuracy of the calibration will be no better than that of thetemperature standards used.NOTE 2It is possible to develop a more elaborate method of tempera-ture calibration using multiple (more than two) fusion standards andquadratic regression ana
9、lysis. Since most modern instruments are capableof heating rates which are essentially linear in the region of use, theprocedure given here is limited to a two-point calibration.5. Significance and Use5.1 Thermomechanical analyzers are employed in theirvarious modes of operation (penetration, expans
10、ion, flexure,etc.) to characterize a wide range of materials. In most cases,the value to be assigned in thermomechanical measurements isthe temperature of the transition (or event) under study.Therefore, the temperature axis (abscissa) of all TMA thermalcurves must be accurately calibrated either by
11、 direct reading ofa temperature sensor or by adjusting the programmer tempera-ture to match the actual temperature over the temperature rangeof interest.6. Apparatus6.1 Thermomechanical Analyzer (TMA), The essential in-strumentation required to provide the minimum thermome-chanical analytical or the
12、rmodilatometric capability for thismethod includes:1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on ThermalTest Methods and Practices.Current edition approved Sept. 1, 2008. Published September 2008. Or
13、iginallyapproved in 1990. Last previous edition approved in 2003 as E 1363 03.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page on
14、the ASTM website.3Available from American National Standards Institute, 11 W. 42nd St., 13thFloor, New York, NY 10036.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6.1.1 A Rigid Specimen Holder or Platform, of inert, lowexpansivity
15、 material ( 1 m m-1K-1) to center the specimen inthe furnace and to fix the specimen to mechanical ground.6.1.2 A Rigid (expansion compression, flexure, tensile, etc)Probe, of inert, low expansivity material ( 1 m m-1K-1) thatcontacts with the specimen with an applied compressive ortensile force. Fo
16、r this test method the use of a penetration probeis recommended.6.1.3 A Sensing Elementlinear over a minimum range of2 mm to measure the displacement of the rigid probe to 6 50nm resulting from changes in the length/height of the speci-men.6.1.4 A Weight or Force Transducer, to generate a constantfo
17、rce of 50 6 5 mN (5.0 6 0.5 g) that is applied through therigid probe to the specimen.NOTE 3The recommendation of a 5.0 g load (or a force of 50 mN) isbased on the use of penetration probes commonly used in the commer-cially available thermomechanical analyzers. These probes have tipdiameters rangin
18、g from 0.89 to 2.0 mm and lead to pressures from 80 to16 kPa when using the recommended 5.0 g load. The use of probes whichdiffer greatly from this range of tip diameters may require differentloading (or force).6.1.5 A Furnace, capable of providing uniform controlledheating (cooling) at a rate of 10
19、 6 1 C min-1of a specimen toa constant temperature within the applicable temperature rangeof this methodNOTE 4The temperature range of operation of commercial thermo-mechanical analyzers vary by manufacturer and mode. The completerange of temperature of an instrument is sometimes achieved by the use
20、 oftwo different furnaces. In this case, temperature calibration must becarried out for each furnace.6.1.6 A Temperature Controller, capable of executing aspecific temperature program by operating the furnace betweenselected temperature limits at a rate of temperature change of10 6 1 C min-1.6.1.7 A
21、 Temperature Sensor, that may be positioned in closeproximity to the test specimen to provide an indication of thespecimen/furnace temperature to within 6 0.1 C min6.1.8 A means of sustaining an environment around thespecimen with an inert purge gas (e.g., nitrogen, helium, argon,etc.) at a purge ga
22、s flow rate of 20 to 50 mL min-1.6.1.9 A Data Collection Device, to provide a means ofacquiring, storing, and displaying measured or calculatedsignals, or both. The minimum output signals required forTMAare a change in linear dimension, temperature, and times.7. Hazards7.1 This test method may invol
23、ve the use of hazardousmaterials, operations, and equipment. It is the responsibility ofthe user of this test method to establish appropriate safetypractice and to determine the applicability of regulatorylimitations prior to use.NOTE 5Warning: Toxic or corrosive effluents, or both, may bereleased w
24、hen heating some materials and could be harmful to personneland the apparatus.7.2 Once this calibration procedure has been executed asdescribed in 10.1.2.1-10.1.2.7 of this test method, the measur-ing temperature sensor position should not be changed, norshould it be in contact with the sample or sa
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME13632008STANDARDTESTMETHODFORTEMPERATURECALIBRATIONOFTHERMOMECHANICALANALYZERS 机械 分析 温度 校准 标准 试验

链接地址:http://www.mydoc123.com/p-528549.html