ITU-T V 16-1988 MEDICAL ANALOGUE DATA TRANSMISSION MODEMS《医务上模拟数据传输用的调制解调器-电话网络上的数据交流 第XVII研究组 第7页》.pdf
《ITU-T V 16-1988 MEDICAL ANALOGUE DATA TRANSMISSION MODEMS《医务上模拟数据传输用的调制解调器-电话网络上的数据交流 第XVII研究组 第7页》.pdf》由会员分享,可在线阅读,更多相关《ITU-T V 16-1988 MEDICAL ANALOGUE DATA TRANSMISSION MODEMS《医务上模拟数据传输用的调制解调器-电话网络上的数据交流 第XVII研究组 第7页》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、INTERNATIONAL TELECOMMUNICATION UNION)45G134 6 TELECOMMUNICATIONSTANDARDIZATION SECTOROF ITU$!4!G0G0#/-5.)#!4)/.G0G0/6%2G0G04(%G0G04%,%0(/.%.%47/2+-%$)#!,G0G0!.!,/5%G0G0$!4!G0G042!.3-)33)/.-/$%-3)45G134G0G0RecommendationG0G06 (Extract from the “LUEG0“OOK)NOTES1 ITU-T Recommendation V.16 was publishe
2、d in Fascicle VIII.1 of the Blue Book. This file is an extract fromthe Blue Book. While the presentation and layout of the text might be slightly different from the Blue Book version, thecontents of the file are identical to the Blue Book version and copyright conditions remain unchanged (see below)
3、.2 In this Recommendation, the expression “Administration” is used for conciseness to indicate both atelecommunication administration and a recognized operating agency. ITU 1988, 1993All rights reserved. No part of this publication may be reproduced or utilized in any form or by any means, electroni
4、c ormechanical, including photocopying and microfilm, without permission in writing from the ITU.Fascicle VIII.1 - Rec. V.16 1Recommendation V.16Fascicle VIII.1 - Rec. V.16MEDICAL ANALOGUE DATA TRANSMISSION MODEMS(Geneva, 1976)The CCITT,considering that(a) computer-aided automatic ECG (electro-cardi
5、ogram) interpretation is being made available by specialdiagnostic centres to general practitioners and hospitals at remote places and suitable transmission equipment isnecessary for this reason;(b) such a service can be implemented to advantage in a special data collection system using simple remot
6、estations and a high-quality central unit;(c) for such applications particularly suitable and compatible transmission facilities are necessary whichmust not interfere with other telephone services;(d) analogue as well as digitalized transmission of the analogue data (e.g. ECG records) are in princip
7、lepossible;(e) in most cases, however, on-line transmission with analogue transmission methods can be implementedmore easily and economically;(f) in practice, analogue transmission generally promises a sufficient degree of quality;(g) in cases of emergency and monitoring of implanted pacemakers, ver
8、y simple, acoustically coupledequipment may be of great assistance to the persons concerned;unanimously declares the following view1 Analogue transmission of medical analogue data, e.g. ECGs, should be permitted in the public telephonenetwork. Reliable, sufficiently interference-free transmission ca
9、nnot be taken for granted on every connection or route.Therefore, it is necessary to test the connections under consideration before such a service is definitely introduced.2 This service requires two basically different transmission devices (modems):1) transmission equipment for simultaneous transm
10、ission of three ECG signals on a telephone channel froma remote station to the central station, preferably for direct galvanic coupling to the telephone channel;2) transmission equipment, preferably for emergency use and for monitoring of implanted pacemakers, tosimultaneously transmit only one ECG
11、signal from a remote station to a central station with acoustic orgalvanic coupling to a telephone channel.The ECG station usually consists of an ECG recorder, including separating amplifier, data input/output deviceand the modem specified in this Recommendation (see Figure 1/V.16).The central stati
12、on usually comprises the central modem specified herein and the interpretation system forECGs (e.g. a computer programmed for ECG interpretation).This Recommendation covers the modems, the desired transmission characteristics of the ECG transmissionchannel as well as the necessary interchange circui
13、ts and the method of transmitting the digital data associated with theECG (e.g. patients identification codes, control signals in both transmission directions and the interpretation record).2 Fascicle VIII.1 - Rec. V.163 Modems for simultaneous analogue transmission of three ECG records3.1 Basic cha
14、racteristics of the analogue channelsThe equipment specified below is mainly intended for operation with direct galvanic coupling to telephonelines.3.1.1 Baseband signalBaseband signal requirements at the modem input:- number of simultaneously transmitted ECGrecords 3- frequency response of the sepa
15、rating amplifier flat- signal-to-noise ratio with 10 Hz square wave 50 dBsignals 1 V (unweighted)- full scale limit (see Note 1) 2.5 V- linearity deviation of an ECG channel related tofull scale and the optimum straight line 1%- permissible group delay distortion of the input from 3 to 60 Hzsignal a
16、t the modulator input (including channel 2 ms (outside this range,filter in the baseband) see Figure 2/V.16)Fascicle VIII.1 - Rec. V.16 3- spectrum: if a.c. coupling is applied, a timeconstant of = 3.2 s, corresponding to a lowercut-off frequency of 0.5 Hz, should be used.- baseband pre-emphasis (se
17、e Note 2) (between rise of 6 dB/octave;separating amplifier and modem) cut-off frequency: 15 HzNote 1 - Existing instruments (ECG recorders, etc.) are designed for 2.5 V full scale. If, however, theInternational Electrotechnical Commission specifies 1 V or 1.25 V as the full scale limit, this value
18、should beadopted. The slope of the modulator characteristic (see 3.1.2 below) must then be adjusted accordingly.Note 2 - This value will require further study if, at a later date, amplitude compandors are used to improve thesignal-to-noise ratio.3.1.2 Requirements for the transmission equipment (mod
19、ems)The modem should be capable of transmitting baseband signals with a bandwidth of up to approximately 100Hz. The transmission equipment (modems) should not deteriorate the performance of the baseband signal as specifiedunder 3.1.1 above by more than 10%. The exact value of the admissible deterior
20、ation needs further study.Since the centre channel of the transmission equipment will in future be used for digital transmission of ECG-associated digital data and other biological data, it must be capable of transmitting d.c. components. The same shouldapply to the other channels.- line signals for
21、 transmitting the ECG: signals as specified under 3.1.1 above- modulation method: frequency modulation- subcarrier frequencies fnand associated maximumtransmission levels pn:f1= 950 Hz 6 Hz p1= 7 dB lower than the resultingf2= 1400 Hz 15 Hz (see Note 1) p2= 5 dB level p0as specified inf3= 2100 Hz 15
22、 Hz p3= 3 dB Recommendation V.2- resulting maximum level: p0as specified in Recommendation V.2- simultaneous transmission of all three subcarriersis mandatory, if subcarriers f1and/or f3are used.- maximum frequency deviation per channel in thecase of linear operation: f = 100 Hz4 Fascicle VIII.1 - R
23、ec. V.16- slope of the modulator characteristic (subcarrierdeviation sensitivity): 40 Hz/V (see Note 2)- a positive signal should cause a rise in thesubcarrier frequency- FM channel bandwidth (3 dB points): 350 Hz- resulting level accepted by receiver - 6 dBm to(upper threshold-level): -43 dBm- lowe
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ITUTV161988MEDICALANALOGUEDATATRANSMISSIONMODEMS 医务 模拟 数据传输 调制解调器 电话 网络 数据 交流 XVII 研究 PDF

链接地址:http://www.mydoc123.com/p-804228.html