ASTM UOP925-2016 Nickel Molybdenum Phosphorus Cobalt and Aluminum in Fresh Catalysts by ICP-OES《采用ICP-OES法测定新鲜催化剂中的镍 钼 磷 钴和铝》.pdf
《ASTM UOP925-2016 Nickel Molybdenum Phosphorus Cobalt and Aluminum in Fresh Catalysts by ICP-OES《采用ICP-OES法测定新鲜催化剂中的镍 钼 磷 钴和铝》.pdf》由会员分享,可在线阅读,更多相关《ASTM UOP925-2016 Nickel Molybdenum Phosphorus Cobalt and Aluminum in Fresh Catalysts by ICP-OES《采用ICP-OES法测定新鲜催化剂中的镍 钼 磷 钴和铝》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1992, 2016 UOP LLC. All rights reserved. Nonconfidential U
3、OP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. Nickel
4、, Molybdenum, Phosphorous, Cobalt, and Aluminum in Fresh Catalysts by ICP-OES UOP Method 925-16 Scope This method is for determining nickel, molybdenum, phosphorus and cobalt in fresh gamma-alumina supported catalysts; and nickel (Ni), molybdenum (Mo), phosphorus (P), cobalt (Co) and aluminum (Al) i
5、n fresh silica-alumina supported catalysts of known composition by Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES). The method is suitable for determining nickel over a range of 1 to 10 mass-%, molybdenum over a range of 5 to 20 mass-%, phosphorus (as phosphate only) over a range
6、of 0.5 to 25 mass-%, cobalt over a range of 1 to 5 mass-%, and aluminum over a range of 5 to 50 mass-%. Catalysts containing alpha- or theta-alumina phases cannot be analyzed by the method. Nickel bearing catalysts that contain tungsten are analyzed by UOP Method 924. References UOP Method 918, “Col
7、lection and Grinding of Adsorbents and Catalysts for Analysis,” www.astm.org UOP Method 924, “Nickel, Tungsten and Aluminium in Fresh Catalysts by ICP-AES,” www.astm.org UOP Method 954, “Loss on Ignition (LOI) for Fresh, Regenerated, Used, and Spent Catalysts, Catalyst Supports, and Adsorbents,” www
8、.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Perkin Elmer WinLab32 for ICP Instrument Control Software Manual Outline of Method Samples are ground using UOP Method 918, with alumina supported samples being dissolved using sulfuric acid, while silica-alumina supported
9、 samples are dissolved using sulfuric and hydrofluoric acids. The resulting sample solution is aspirated into an ICP-OES instrument and the element concentrations determined by comparison to a matrix-matched standard solution. Sample results are reported on a volatile-free basis, based on a 500 C LO
10、I (UOP Method 954, “Loss on Ignition (LOI) for Fresh, Regenerated, Used, and Spent Catalysts, Catalyst Supports, and Adsorbents”). Aluminum is determined on silica-alumina supported catalysts only. If desired, the results can be reported on an oxide basis also. 2 of 7 925-16 Apparatus References to
11、catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, readable to 0.0001 g Beaker, quartz, low form, 250-mL, Ace Glass, Inc., Cat. No. 5334-14 Beaker, PTFE, 250-mL, Fisher Scientific, Cat. No. 02-586-1H Crucible, wide form, 17 mL, porce
12、lain, Coors No. 60135, Fisher Scientific, Cat. No.FB-965-J Cylinders, graduated, polypropylene, 10- and 50-mL, Fisher Scientific, Cat. Nos. 08-572A and -572C, respectively Flasks, volumetric, Class A, 100-, 200-, 500- and 1000-mL, Fisher Scientific, Cat. Nos. 13-756-694, -695, -697 and -698, respect
13、ively, several of each required Funnel, disposable, 55-mm diameter, Fisher Scientific, Cat. No. 10-320A Hot plate, variable heat to at least 400 C, Fisher Scientific, Cat. No. 11-100-49H Inductively Coupled Plasma-Optical Emission Spectrometer, computer controlled, having a bandpass of 0.010 nm (FWH
14、M) or better, and covering the wavelength range specified in Table 2. The instrument must have the ability to use internal standards and perform background corrections, and must measure the internal standard response concurrently with each element response. A Perkin-Elmer 7300DV spectrometer with a
15、corrosion resistant sample introduction system, or its equivalent, is required. Pipets, total volume, Class A, 5-, 7-, 8-, 10-, 15-, 20-, 25-, 30- and 50-mL, Fisher Scientific, Cat. Nos. 13-660-5E, 13-646-42, -43, 13-660-5F, -5M, -5H, -5J, -5S, and -5K, several of each required. Alternatively, a dig
16、ital pipettor may be used. Pipets, total volume, 1-5 mL, 1-10mL, Eppendorf pipet Cat. No. 4920000105, 4920000113 Pipet filler, Fisher Scientific, Cat. No. 13-681-102A Regulator, argon, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Regulator,
17、 nitrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Watch glass, 100-mm, glass, for 250-mL beaker, Fisher Scientific, Cat. No. 02-612D Watch glass, 100-mm, PTFE, Cat. No. 02-617-1E Reagents and Materials References to catalog numbers an
18、d suppliers are included as a convenience to the method user. Other suppliers may be used. References to water mean deionized or distilled water. Aluminum solution, 1000-g/mL, Spex Certiprep, Inc, Cat. No. PLAL2-2X Argon, 99.995% minimum purity, ICP feed gas Cobalt solution, 1000-g/mL, Spex Certipre
19、p, Inc, Cat. No. PLCO2-2X Gloves, neoprene, Fisher Scientific, Cat. Nos. 11-391-129A through -129E for sizes 7 through 10, respectively 3 of 7 925-16 Hydrofluoric acid, concentrated, trace metals grade, VWR, Cat. No. EM-HX0628-7 Molybdenum solution, 1000-g/mL, Spex Certiprep, Inc, Cat. No. PLMO9-2X
20、Nickel solution, 1000-g/mL, Spex Certiprep, Inc, Cat. No. PLNI2-2X Nitric acid, concentrated, trace metals grade, VWR, Cat. No. EM-NX0408-2 Nitrogen, 99.99% minimum purity, water/oil free, ICP optics purge gas Phosphorus solution, 1000-g/mL, Spex Certiprep, Inc, Cat. No. PLP9-2X Scandium metal solut
21、ion, aqueous, 10,000-g/mL, SPEX CertiPrep, Inc., Cat. No. PLSC2-3X, for use as an internal standard Scandium internal standard, aqueous, 100-g/mL. To prepare 100 mL, pipet 1 mL of the 10,000- g/mL scandium metal solution into a 100-mL volumetric flask. Fill to the mark with deionized water. Cap and
22、invert several times to mix. Store in a 125-mL polypropylene bottle. The solution should remain stable for two months. Sulfuric acid, concentrated, trace metals grade, VWR, Cat. No. EM-SX1248-6 Yttrium solution, 1000-g/mL, Spex Certiprep, Inc, Cat. No. PLY2-2X. Not required if the spectrometer speci
23、fied in Apparatus is used. Procedure Caution: All subsequent steps involving additions of acids and heating of samples and digests must be performed in a properly operating fume hood, using appropriate precautions. Gloves, safety glasses, goggles, and/or face shield, and other appropriate personal p
24、rotective equipment must be worn. Consult safety data sheet (SDS) and local regulations for each material used. The analyst is expected to be familiar with general laboratory practices, the technique of ICP-OES, and the equipment being used. Dispose of used supplies and samples in an environmentally
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMUOP9252016NICKELMOLYBDENUMPHOSPHORUSCOBALTANDALUMINUMINFRESHCATALYSTSBYICPOES 采用 ICPOES 测定 新鲜 催化剂

链接地址:http://www.mydoc123.com/p-540831.html