《ASTM F2901-2013 Standard Guide for Selecting Tests to Evaluate Potential Neurotoxicity of Medical Devices《评估医疗设备潜在神经毒性的选择试验的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2901-2013 Standard Guide for Selecting Tests to Evaluate Potential Neurotoxicity of Medical Devices《评估医疗设备潜在神经毒性的选择试验的标准指南》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2901 13Standard Guide forSelecting Tests to Evaluate Potential Neurotoxicity ofMedical Devices1This standard is issued under the fixed designation F2901; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of las
2、t revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 Medical devices may cause adverse effects on thestructure and/or function of the nervous system. In this guide,these adve
3、rse effects are defined as neurotoxicity. This guideprovides background information and recommendations onmethods for neurotoxicity testing. This guide should be usedwith Practice F748, and may be helpful where neurotoxicitytesting is needed to evaluate medical devices that contactnervous system tis
4、sue or cerebral spinal fluid (CSF).1.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limit
5、ations prior to use.2. Referenced Documents2.1 ASTM Standards:2F748 Practice for Selecting Generic Biological Test Methodsfor Materials and DevicesF1904 Practice for Testing the Biological Responses toParticles in vivo2.2 Other Referenced Documents:ISO/AAMI/ANSI 10993-3 :2003 Biological Evaluation o
6、fMedical DevicesPart 3: Tests for Genotoxicity,Carcinogenicity, and Reproductive Toxicity3ISO/AAMI/ANSI 10993-5 :2009 Biological Evaluation ofMedical DevicesPart 5: Tests for In Vitro Cytotoxicity3ISO 1099311 : 2006 Biological Evaluation of MedicalDevicesPart 11: Tests for Systemic ToxicityISO/AAMI/
7、ANSI 10993-18 Biological Evaluation of Medi-cal DevicesPart 18: Chemical Characterization of Ma-terials3ANSI/AAMI ST72 :2010 Bacterial EndotoxinsTestMethodologies, Routine Monitoring, and Alternatives toBatch Testing3USP Rabbit Pyrogen Test4USP Transfusion and Infusion Assemblies and Simi-lar Medica
8、l Devices43. Summary of Guide3.1 This is an informative guide and should be used withPractice F748.3.2 The duration of contact between the tissue and medicaldevice should be considered when determining the appropriatepanel of testing. This guide may not address neurosurgicalinstruments or medical de
9、vices that have transient incidentalcontact with the nervous system due to the limited tissuecontact duration.3.3 The evaluation of neurotoxicity should be considered inconjunction with material characterization and other informa-tion such as non-clinical tests, clinical studies, post-marketexperien
10、ce, and intended use.4. Significance and Use4.1 The objective of this guide is to recommend a panel ofbiological tests that can be used in addition to the testingrecommended in Practice F748. This guide is designed todetect neurotoxicity caused by medical devices that contactnervous tissue.4.2 The t
11、esting recommendations should be considered fornew materials, established materials with different manufactur-ing methods that could affect nervous tissue response, ormaterials used in new nervous tissue applications.4.3 Chemical characterization can be used to evaluate simi-larity for materials wit
12、h a history of clinical use in a similarnervous tissue application.1This guide is under the jurisdiction of ASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.16 on Biocompatibility Test Methods.Current edition approved Feb. 1, 2013. P
13、ublished February 2013. Originallyapproved in 2012. Last previous edition approved in 2012 as F2901 12. DOI:10.1520/F290113.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, r
14、efer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.4Available from U.S. Pharmacopeia (USP), 12601 Twinbrook Pkwy., Rockville,MD 20852-1790, http:/www.usp.org.Cop
15、yright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Tests for Neurotoxicity5.1 Testing should be performed on the final sterilizeddevice, representative samples from the final sterilized device,or materials processed in the same manner as
16、 the final sterilizeddevice. Testing of individual materials may be useful forresearch and development, but the definitive neurotoxicityevaluation should include all materials in the final version ofthe device. The test article should be exposed to all phases ofmanufacturing including processing, cl
17、eaning, sterilization,and packaging.5.1.1 A complete description of all device materials andreagents used during manufacturing and processing should beprovided with information on the source, purity, and toxicityprofile. Chemical characterization studies can provide addi-tional information on the de
18、vice safety profile. See ISO/AAMI/ANSI 10993-18 for information on chemical characterizationof materials.5.2 The following tests should be considered to assessneurotoxicity of medical devices within the scope of this guide.5.2.1 CytotoxicityCytotoxicity assays are sensitivescreening tools that gener
19、ally serve as a starting point forevaluating medical device biocompatibility. See X1.4 forinformation on neuro-cytotoxicity testing.5.2.2 GenotoxicityNervous tissue contains proliferatingcell populations, and can respond to device implantation witha proliferative response. Nervous tissue is also kno
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF29012013STANDARDGUIDEFORSELECTINGTESTSTOEVALUATEPOTENTIALNEUROTOXICITYOFMEDICALDEVICES 评估 医疗 设备

链接地址:http://www.mydoc123.com/p-539693.html