ASTM E1866-1997(2002) Standard Guide for Establishing Spectrophotometer Performance Tests《建立光谱仪性能试验的标准指南》.pdf
《ASTM E1866-1997(2002) Standard Guide for Establishing Spectrophotometer Performance Tests《建立光谱仪性能试验的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1866-1997(2002) Standard Guide for Establishing Spectrophotometer Performance Tests《建立光谱仪性能试验的标准指南》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1866 97 (Reapproved 2002)Standard Guide forEstablishing Spectrophotometer Performance Tests1This standard is issued under the fixed designation E 1866; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide describes basic procedures that can be usedto develop spectrophotometer performance tests. The guide isintende
3、d to be applicable to spectrophotometers operating inthe ultraviolet, visible, near-infrared and mid-infrared regions.1.2 This guide is not intended as a replacement for specificpractices such as Practices E 275, E 925, E 932, E 958, E 1421,or E 1683 that exist for measuring performance of specifict
4、ypes of spectrophotometers. Instead, this guide is intended toprovide guidelines in how similar practices should be devel-oped when specific practices do not exist for a particularspectrophotometer type, or when specific practices are notapplicable due to sampling or safety concerns. This guide canb
5、e used to develop performance tests for on-line processspectrophotometers.1.3 This guide describes univariate level zero and level onetests, and multivariate level A and level B tests which can beimplemented to measure spectrophotometer performance.These tests are designed to be used as rapid, routi
6、ne checks ofspectrophotometer performance. They are designed to uncovermalfunctions or other changes in instrument operation, but donot specifically diagnose or quantitatively assess the malfunc-tion or change. The tests are not intended for the comparison ofspectrophotometers of different manufactu
7、re.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Reference
8、d Documents2.1 ASTM Standards:E 131 Terminology Relating to Molecular Spectroscopy2E 275 Practice for Describing and Measuring Performanceof Ultraviolet, Visible, and Near-Infrared Spectrophotom-eters2E 387 Test Method for Estimating Stray Radiant PowerRatio of Spectrophotometers by the Opaque Filte
9、r Method2E 925 Practice for the Periodic Calibration of Narrow Band-Pass Spectrophotometers2E 932 Practice for Describing and Measuring Performanceof Dispersive Infrared Spectrometers2E 958 Practice for Measuring Practical Spectral Bandwidthof Ultraviolet-Visible Spectrophotometers2E 1421 Practice f
10、or Describing and Measuring Performanceof Fourier Transform Infrared (FT-IR) Spectrometers:Level Zero and Level One Tests2E 1655 Practice for Infrared, Multivariate, QuantitativeAnalysis2E 1683 Practice for Testing the Performance of ScanningRaman Spectrometers23. Terminology3.1 DefinitionsFor termi
11、nology relating to molecularspectroscopic methods, refer to Terminology E 131.3.2 Definitions of Terms Specific to This Standard:3.2.1 action limit, nthe limiting value from an instrumentperformance test, beyond which the spectrophotometer isexpected to produce potentially invalid results.3.2.2 chec
12、k sample, na single pure compound, or aknown, reproducible mixture of compounds whose spectrum isconstant over time such that it can be used in a performancetest.3.2.3 level A test, na pass/fail spectrophotometer perfor-mance test in which the spectrum of a check or test sample iscompared against hi
13、storical spectra of the same sample via amultivariate analysis.3.2.4 level B test, na pass/fail spectrophotometer perfor-mance test in which the spectrum of a check or test sample isanalyzed using a multivariate model, and the results of theanalysis are compared to historical results for prior analy
14、ses ofthe same sample.3.2.5 level one (1) test, na simple series of measurementsdesigned to provide quantitative data on various aspects ofspectrophotometer performance and information on which tobase the diagnosis of problems.3.2.6 level zero (0) test, na routine check of spectropho-tometer perform
15、ance, which can be done in a few minutes,designed to visually detect significant changes in instrument1This guide is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.03 on Infrared and Near Infrared Spe
16、ctroscopy.Current edition approved Sept. 10, 2002. Published September 2002.2Annual Book of ASTM Standards, Vol 03.06.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.performance and provide a database to determine instrumentperforman
17、ce over time.3.2.7 optical reference filter, nan optical filter or otherdevice which can be inserted into the optical path in thespectrophotometer or probe producing an absorption spectrumwhich is known to be constant over time such that it can beused in place of a check or test sample in a performa
18、nce test.3.2.8 test sample, na process or product sample, or amixture of process or product samples which has a constantspectrum for a finite time period and which can be used in aperformance test. Test samples and their spectra are generallynot reproducible in the long term.4. Significance and Use4
19、.1 If ASTM Committee E-13 has not specified an appro-priate test procedure for a specific type of spectrophotometer,or if the sample specified by a Committee E-13 procedure isincompatible with the intended spectrophotometer operation,then this guide can be used to develop practical performancetests.
20、4.1.1 For spectrophotometers which are equipped with per-manent or semi-permanent sampling accessories, the testsample specified in a Committee E-13 practice may not becompatible with the spectrophotometer configuration. For ex-ample, for FT-MIR instruments equipped with transmittance orIRS flow cel
21、ls, tests based on polystyrene films are impractical.In such cases, these guidelines suggest means by which therecommended test procedures can be modified so as to beperformed on a compatible test material.4.1.2 For spectrophotometers used in process measure-ments, the choice of test materials may b
22、e limited due toprocess contamination and safety considerations. These guide-lines suggest means of developing performance tests based onmaterials which are compatible with the intended use of thespectrophotometer.4.2 Tests developed using these guidelines are intended toallow the user to compare th
23、e performance of a spectropho-tometer on any given day with prior performance. The tests areintended to uncover malfunctions or other changes in instru-ment operation, but they are not designed to diagnose orquantitatively assess the malfunction or change. The tests arenot intended for the compariso
24、n of spectrophotometers ofdifferent manufacture.5. Test Conditions5.1 When conducting the performance tests, the spectropho-tometer should be operated under the same conditions as willbe in effect during its intended use. Sufficient warm-up timeshould be allowed before the commencement of any measur
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME186619972002STANDARDGUIDEFORESTABLISHINGSPECTROPHOTOMETERPERFORMANCETESTS 建立 光谱仪 性能 试验 标准 指南 PDF

链接地址:http://www.mydoc123.com/p-529673.html