ASTM E1363-2016 Standard Test Method for Temperature Calibration of Thermomechanical Analyzers《热机械分析仪温度校准的标准试验方法》.pdf
《ASTM E1363-2016 Standard Test Method for Temperature Calibration of Thermomechanical Analyzers《热机械分析仪温度校准的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1363-2016 Standard Test Method for Temperature Calibration of Thermomechanical Analyzers《热机械分析仪温度校准的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1363 13E1363 16Standard Test Method forTemperature Calibration of Thermomechanical Analyzers1This standard is issued under the fixed designation E1363; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope Scope*1.1 This test method describes the temperature calibration of thermomechanical analyzers from 50from 50 to1100C.1500C. (
3、See Note 1.)1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.3 This standard is similar to ISO 113591 but addresses a larger temperature range and utilizes additional calibrationmaterials.1.4 This standard does not purpor
4、t to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use. Specific precautionary statements are given in Se
5、ction 7 and Note 1011.2. Referenced Documents2.1 ASTM Standards:2E473 Terminology Relating to Thermal Analysis and Rheology2.2 Other Standards:3ISO 113591 Thermomechanical Analysis (TMA)-Part 1: General Principles3. Terminology3.1 Definitions:3.1.1 The terminology relating to thermal analysis appear
6、ing in Terminology E473 shall be considered applicable to thisdocument.4. Summary of Test Method4.1 An equation is developed for the linear correlation of the experimentally observed program temperature and the actualmelting temperature for known melting standards. This is accomplished through the u
7、se of a thermomechanical analyzer with apenetration probe to obtain the onset temperatures for two melting point standards. An alternate, one-point method of temperaturecalibration,calibration is also given for use over very narrow temperature ranges. (See Note 2.)NOTE 1This test method may be used
8、for calibrating thermomechanical analyzers at temperatures outside this range of temperature. However, theaccuracy of the calibration will be no better than that of the temperature standards used.NOTE 2It is possible to develop a more elaborate method of temperature calibration using multiple (more
9、than two) fusion standards and quadraticregression analysis. Since most modern instruments are capable of heating rates which are essentially linear in the region of use, the procedure given hereis limited to a two-point calibration.5. Significance and Use5.1 Thermomechanical analyzers are employed
10、in their various modes of operation (penetration, expansion, flexure, etc.) tocharacterize a wide range of materials. In most cases, the value to be assigned in thermomechanical measurements is thetemperature of the transition (or event) under study. Therefore, the temperature axis (abscissa) of all
11、 TMA thermal curves must1 This test method is under the jurisdiction ofASTM Committee E37 on Thermal Measurements and is the direct responsibility of Subcommittee E37.10 on Fundamental,Statistical and Mechanical Properties.Current edition approved April 1, 2013Dec. 1, 2016. Published May 2013January
12、 2017. Originally approved in 1990. Last previous edition approved in 20082013 asE1363 08.E1363 13. DOI: 10.1520/E1363-13.10.1520/E1363-16.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume
13、information, refer to the standards Document Summary page on the ASTM website.3 Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http:/www.ansi.org.This document is not an ASTM standard and is intended only to provide the user of an ASTM sta
14、ndard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by AS
15、TM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1be accurately calibrated either by direct reading of a temperature sensor or
16、 by adjusting the programmer temperature to match theactual temperature over the temperature range of interest.6. Apparatus6.1 Thermomechanical Analyzer (TMA), The essential instrumentation required to provide the minimum thermomechanicalanalytical or thermodilatometric capability for this method in
17、cludes:6.1.1 A Rigid Specimen Holder or Platform, of inert, low expansivity material (1 m m-1 K-1) to center the specimen in thefurnace and to fix the specimen to mechanical ground.6.1.2 A Rigid (expansion compression, flexure, tensile, etc.) Probe, of inert, low expansivity material (1 m m-1 K-1) t
18、hatcontacts with the specimen with an applied compressive or tensile force. For this test method, the use of a penetration probe isrecommended.6.1.3 A Sensing Element, linear over a minimum range of 2 mm 2 mm to measure the displacement of the rigid probe to 650nm resulting from changes in the lengt
19、h/height of the specimen.6.1.4 A Weight or Force Transducer, to generate a constant force of 50 6 5 mN (5.0 6 0.5 g) that is applied through the rigidprobe to the specimen.NOTE 3The recommendation of a 5.0 g load (or a force of 50 mN) is based on the use of penetration probes commonly used in the co
20、mmerciallyavailable thermomechanical analyzers. These probes have tip diameters ranging from 0.89 to 2.0 mm and lead to pressures from 80 to 16 kPa when usingthe recommended 5.0 g load. The use of probes which differ greatly from this range of tip diameters may require different loading (or force).6
21、.1.5 A Furnace, capable of providing uniform controlled heating (cooling) at a rate of 1 1C min-1 to 10 6 1C min-1 of aspecimen to a constant temperature within the applicable temperature range of this method.NOTE 4The temperature range of operation of commercial thermomechanical analyzers vary by m
22、anufacturer and mode. The complete range oftemperature of an instrument is sometimes achieved by the use of two different furnaces. In this case, temperature calibration must be carried out for eachfurnace.6.1.6 A Temperature Controller, capable of executing a specific temperature program by operati
23、ng the furnace between selectedtemperature limits at a rate of temperature change of 10 6 1C min-1.6.1.7 A Temperature Sensor, that may be positioned in close proximity to the test specimen to provide an indication of thespecimen/furnace temperature to within 60.1C min-1.6.1.8 Ameans of sustaining a
24、n environment around the specimen with an inert purge gas (for example, nitrogen, helium, argon,etc.) at a purge gas flow rate of 20 20 mL min-1 to 50 mLmin-1.6.1.9 A Data Collection Device, to provide a means of acquiring, storing, and displaying measured or calculated signals, or both.The minimum
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME13632016STANDARDTESTMETHODFORTEMPERATURECALIBRATIONOFTHERMOMECHANICALANALYZERS 机械 分析 温度 校准 标准 试验

链接地址:http://www.mydoc123.com/p-528551.html