《ASTM D4629-2008 Standard Test Method for Trace Nitrogen in Liquid Petroleum Hydrocarbons by Syringe Inlet Oxidative Combustion and Chemiluminescence Detection.pdf》由会员分享,可在线阅读,更多相关《ASTM D4629-2008 Standard Test Method for Trace Nitrogen in Liquid Petroleum Hydrocarbons by Syringe Inlet Oxidative Combustion and Chemiluminescence Detection.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 4629 08Designation: 379/88An American National StandardStandard Test Method forTrace Nitrogen in Liquid Petroleum Hydrocarbons bySyringe/Inlet Oxidative Combustion and ChemiluminescenceDetection1This standard is issued under the fixed designation D 4629; the number immediately followi
2、ng the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved
3、 for use by agencies of the Department of Defense.1. Scope*1.1 This test method covers the determination of the tracetotal nitrogen naturally found in liquid hydrocarbons boiling inthe range from approximately 50 to 400C, with viscositiesbetween approximately 0.2 and 10 cSt (mm2/s) at roomtemperatur
4、e. This test method is applicable to naphthas, distil-lates, and oils containing 0.3 to 100 mg/kg total nitrogen. Forliquid hydrocarbons containing more than 100 mg/kg totalnitrogen, Test Method D 5762 can be more appropriate. Thistest method has been successfully applied, during interlabora-tory st
5、udies, to sample types outside the range of the scope bydilution of the sample in an appropriate solvent to bring thetotal nitrogen concentration and viscosity to within the rangecovered by the test method. However, it is the responsibility ofthe analyst to verify the solubility of the sample in the
6、 solventand that direct introduction of the diluted sample by syringeinto the furnace does not cause low results due to pyrolysis ofthe sample or solvent in the syringe needle.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.
7、1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. See 6.2, 6.4,
8、6.5,6.9, and Section 7.2. Referenced Documents2.1 ASTM Standards:2D 1298 Test Method for Density, Relative Density (SpecificGravity), or API Gravity of Crude Petroleum and LiquidPetroleum Products by Hydrometer MethodD 4052 Test Method for Density and Relative Density ofLiquids by Digital Density Me
9、terD 5762 Test Method for Nitrogen in Petroleum and Petro-leum Products by Boat-Inlet ChemiluminescenceD 6299 Practice for Applying Statistical Quality Assuranceand Control Charting Techniques to Evaluate AnalyticalMeasurement System Performance3. Summary of Test Method3.1 The sample of liquid petro
10、leum hydrocarbon is intro-duced either by syringe or boat inlet system, into a stream ofinert gas (helium or argon). The sample is vaporized andcarried to a high temperature zone where oxygen is introducedand organically bound nitrogen is converted to nitric oxide(NO). The NO contacts ozone, and is
11、converted to excitednitrogen oxide (NO2). The light emitted as the excited NO2decays is detected by a photomultiplier tube and the resultingsignal is a measure of the nitrogen contained in the sample.4. Significance and Use4.1 Some process catalysts used in petroleum and chemicalrefining may be pois
12、oned when even trace amounts of nitrog-enous materials are contained in the feedstocks. This test1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of SubcommitteeD02.03 on Elemental Analysis.Current edition approved
13、 July 1, 2008. Published August 2008. Originallyapproved in 1986. Last previous edition approved in 2007 as D 462902(2007).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, re
14、fer to the standards Document Summary page onthe ASTM website.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.method can be used to determine bound nitrogen in process
15、feeds and may also be used to control nitrogen compounds infinished products.5. Apparatus5.1 Furnace, electric, held at a temperature sufficient tovolatilize and pyrolyze all of the sample and oxidize theorganically bound nitrogen to NO. Furnace temperature(s)shall be as recommended by the manufactu
16、rer (typicallyaround 1000C).5.2 Combustion Tube, fabricated to meet the instrumentmanufacturers specifications.5.3 Drier TubeThe reaction products include water vaporthat must be eliminated prior to measurement by the detector.This can be accomplished with a magnesium perchlorateMg(ClO4)2scrubber or
17、 a membrane drying tube (permeationdrier), or by whatever other means the instrument manufacturerspecifies as appropriate for the instrument being used.5.4 Chemiluminescent Detector, capable of measuring lightemitted from the reaction between NO and ozone.5.5 Totalizer, having variable attenuation,
18、and capable ofmeasuring, amplifying, and integrating the current from thechemiluminescent detector. A built in microprocessor or at-tached computer system may perform these functions.5.6 Micro-litre Syringe, of 5, 10, 25, 50, or 250 L capacitycapable of accurately delivering micro-litre quantities i
19、s re-quired. The needle should be long enough to reach the hottestportion of the inlet section of the furnace when injecting thesample. The syringe may be part of an automatic sampling andinjection device used with the instrument.5.7 Strip Chart Recorder (Optional).5.8 Sample Inlet SystemOne of the
20、following must beused:5.8.1 Manually Operated Syringe.5.8.2 Syringe, with a constant rate injector system, capableof delivering a sample at a precisely controlled rate.5.8.3 Boat Inlet System, to facilitate analysis of samples thatwould react with the syringe or syringe needle. The pyrolysistube for
21、 boat inlet use may require specific construction topermit insertion of a boat fully into the inlet section of thefurnace. The boat inlet system external to the furnace may becooled to a temperature below room temperature to aid indissipating the heat from the boat when it is removed from thefurnace
22、. Cooling the boat inlet system may also reduce thechances of the sample combusting in the boat before introduc-tion into the furnace and may be necessary when runningvolatile samples such as naphtha using a boat inlet system.5.9 Quartz Insert Tube (Optional), may be packed withcupric oxide (CuO) or
23、 other oxidation catalyst as recom-mended by the instrument manufacturer, to aid in completingoxidation. This is inserted into the exit end of the pyrolysistube.5.10 Vacuum System (Optional), The chemiluminescencedetector may be equipped with a vacuum system to maintainthe reaction cell at reduced p
24、ressure (typically 20 to 25 mmHg). This can improve the signal to noise ratio of the detector.5.11 Analytical Balance (Optional), with a precision of60.01 mg.6. Reagents6.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagent
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD46292008STANDARDTESTMETHODFORTRACENITROGENINLIQUIDPETROLEUMHYDROCARBONSBYSYRINGEINLETOXIDATIVECOMBUSTIONANDCHEMILUMINESCENCEDETECTIONPDF

链接地址:http://www.mydoc123.com/p-517538.html