ASTM UOP1014-2017 Determination of Trace Propadiene and Methyl Acetylene Impurities in Polymer Grade Propylene by Gas Chromatography《采用气相色谱法测定聚合级丙烯中的痕量丙二烯和甲基乙炔杂质》.pdf
《ASTM UOP1014-2017 Determination of Trace Propadiene and Methyl Acetylene Impurities in Polymer Grade Propylene by Gas Chromatography《采用气相色谱法测定聚合级丙烯中的痕量丙二烯和甲基乙炔杂质》.pdf》由会员分享,可在线阅读,更多相关《ASTM UOP1014-2017 Determination of Trace Propadiene and Methyl Acetylene Impurities in Polymer Grade Propylene by Gas Chromatography《采用气相色谱法测定聚合级丙烯中的痕量丙二烯和甲基乙炔杂质》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2017 UOP LLC. All rights reserved. . Nonconfiden
3、tial UOP Methods are available from ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585
4、 PHONE. Determination of Trace Propadiene and Methyl Acetylene Impurities in Polymer Grade Propylene by Gas Chromatography UOP Method 1014-17 Scope This method is for determining trace methyl acetylene and propadiene (MAPD) in high purity polymer grade propylene liquid or gas using a preconfigured,
5、commercially available gas chromatograph. Other impurities of C3-C5 if present can also be determined. The range of quantitation for each component is 0.2 to 2000 mass- (or mol-) ppm. This method can be used to determine trace impurities in ethylene with appropriate gas injection valve and customize
6、d run program. For a full characterization of the C5 minus impurities to a lower limit of quantitation of 2 ppm refer to UOP 899. References UOP Method 899, “Trace Hydrocarbons in Hydrogen, Hydrocarbon Gases, or LPG by GC,” www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm
7、.org Wasson ECE Instrumentation, Application 263-SP instrument manual Wasson ECE Instrumentation, Vaporizer instrument manual Outline of Method The method requires the use of a dedicated gas chromatographic system that is configured for automated analysis which is capable, via valving and a vaporize
8、r, of determining impurities in polymer- grade propylene gas or liquid. A reproducible sample volume is injected into a GC configured with a vaporizer (external or on- board), gas sampling valve, split/splitless inlet, flame ionization detector (FID), and control and quantitation software. Through t
9、he use of multiple valves, a portion of the sample is injected onto the first column where molecules are separated by boiling point. Next, a portion of the sample, mostly propylene and higher boiling analytes, is cut onto a second column where molecules are separated based on polarity and sent to th
10、e FID for detection. The concentrations of methyl acetylene and propadiene (MAPD) in propylene sample matrices are determined by the external standard method of quantitation, wherein peak areas of the sample impurities are compared to peak areas of a calibration blend. 2 of 11 1014-17 Significance a
11、nd Use High purity propylene and ethylene are commonly used as the feedstock for production of polypropylene, and the quality of this monomer is critical to successful polymerization. The presence of trace amounts of certain hydrocarbon impurities can have adverse effects on the product quality. Thi
12、s test method is suitable for setting specifications, for use as an internal quality control tool, and for use in development or research work. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Analyzer, Wasson-ECE In
13、strumentation Application 263-SP “Analysis of Impurities in Polymer-grade Propylene” Gas chromatographs are available pre-configured, including a data system, from several suppliers. Confirm with the supplier that the analyzer is appropriate for the stream to be analyzed and the analysis required (s
14、ee Figure 1 for Analyzer configuration) which includes: Chromatographic column 1, 30 m of 0.53-mm ID fused silica capillary, internally coated to a film thickness of 0.5-m (bonded) with poly-dimethyl siloxane, boiling point column. Wasson ECE, Part No. 2337 Code KC79 Chromatographic column 2, 50 m o
15、f 0.53-mm ID PLOT fused silica capillary, internally coated to a Film thickness of 9-m with alumina (GS-Alumina), Wasson ECE, Part No. 2491 Code KC136 Gas chromatograph, capable of multiple temperature ramping, built for capillary column chromatography utilizing a split injection system with electro
16、nic pressure control (EPC), having a glass injection port insert and flame ionization detector. Three channels of additional electronic pressure control is required. Agilent Technologies, Model 7890. Vaporizer, model A1034-003, Wasson-ECE Instrumentation, Figure 2 shows Vaporizer flow path. Clamp, f
17、or ring stand and sample cylinders, Fisher Scientific Cat. No. 02-217-000 Clamp holder, for ring stand and clamp, Fisher Scientific Cat. No. 02-217-005 Tubing, stainless steel, 1.6-mm (1/16-inch) OD, 0.76-mm (0.030-inch) ID, Fisher Scientific, Cat. No. AT3003 Tubing, translucent, FEP Teflon, 3.2-mm
18、(1/8-inch) OD, 1.6-mm (1/16-inch) ID, 3450 kPa (500 psi), Thomas Scientific, Cat. No. 9567K20 Integrator, or data system, electronic, for obtaining peak areas. This device must integrate areas at a sufficiently fast rate so that narrow peaks typically resulting from use of a capillary column can be
19、accurately measured. ChemStation, Agilent Technologies. Leak detector, gas, Fisher Scientific, Cat. No. 0.-251-702 Regulator, air, two-stage, high purity, delivery pressure range 30 to 700 kPa (4 to 100 psi), Matheson Tri-Gas, Model 3122-590 Regulator, helium, two-stage, high purity, delivery pressu
20、re range 30 to 700 kPa (4 to 100 psi), Matheson Tri-Gas, Model 3122-580 Regulator, hydrogen, two-stage, high purity, delivery pressure range 30 to 700 kPa (4 to 100 psi), Matheson Tri-Gas, Model 3122-350 3 of 11 1014-17 Regulator, nitrogen, two-stage, high purity, delivery pressure range 30 to 700 k
21、Pa (4 to 100 psi), Matheson Tri-Gas, Model 3122-580 Ring stand, with rectangular base, 140- x 229-mm with 610-mm rod, Fisher Scientific, Cat. No. 14- 679Q Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used.
22、 Calibration blend, LPG, quantitative, primary Propylene standard, in a piston cylinder, pressurized with nitrogen to 800 psig, containing approximately 3 ppm of each of the following trace components: Isobutane, n-Butane, Propadiene, trans-2-Butene, 1-Butene, Isobutene, cis-2-Butene, Isopentane, Me
23、thyl Acetylene, n-Pentane, 1,3-Butadiene in Propylene. Local supply. Air, zero gas, total hydrocarbons less than 2.0 ppm as methane Helium, chromatographic grade, typically 99.995% purity Nitrogen, zero gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane Hydrogen, zero-gas, 9
24、9.95% minimum purity, total hydrocarbons less than 0.5 ppm as methane Gas purifier, helium, used to remove oxygen and moisture from carrier gas, VICI Metronics, Cat. No. P100-1 Procedure The analyst is expected to be familiar with general laboratory practices, the technique of gas chromatography, an
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMUOP10142017DETERMINATIONOFTRACEPROPADIENEANDMETHYLACETYLENEIMPURITIESINPOLYMERGRADEPROPYLENEBYGASCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-540820.html