《ASTM F2754 F2754M-2009(2013) Standard Test Method for Measurement of Camber Cast Helix and Direction of Helix of Coiled Wire《测量盘条拱形 浇铸 螺旋及螺旋方向的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2754 F2754M-2009(2013) Standard Test Method for Measurement of Camber Cast Helix and Direction of Helix of Coiled Wire《测量盘条拱形 浇铸 螺旋及螺旋方向的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2754/F2754M 09 (Reapproved 2013)Standard Test Method forMeasurement of Camber, Cast, Helix and Direction of Helixof Coiled Wire1This standard is issued under the fixed designation F2754/F2754M; the number immediately following the designation indicates the yearof original adoption or,
2、in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval.A superscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the various standard methodsthat are used for measuring c
3、amber, cast, helix, and helixdirection. The wire may be coiled with or without a spool.1.2 This test method applies to round wire that has adiameter between 0.0127 to 4.78 mm (0.0005 to 0.188 in.). Italso applies to flat or shaped wire.1.3 This test method does not apply to superelastic nitinolwire.
4、 It does apply to the as-drawn condition of nitinol wire.1.4 This test method does not apply to the measurement ofthe straightness of straightened to length wire and tubing.1.5 The values stated in either SI units or inch-pound unitsare to be regarded separately as standard. The values stated ineach
5、 system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-conformancewith the standard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is th
6、eresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Terminology2.1 Fig. 1(a) through Fig. 1(e) illustrate the physical mean-ing of cast, helix, direction of helix, and camber.2.2
7、 Definitions:2.2.1 castthe maximum diameter of coiled wire when onecomplete circumference rests completely on a flat surface suchas a table, workbench, or floor. Fig. 1(a).2.2.2 helixthere are two common methods for measuringhelixfree end lift and hanging helix. These definitions aredefined by Fig.
8、1(b) and Fig. 1(c), respectively.2.2.2.1 helix (free end lift)the maximum lift of the freeend of the wire when laid on a flat surface, Fig. 1(b).2.2.2.2 helix (hanging helix)the maximum distance be-tween two adjacent coils of wire, Fig. 1(c).Ahanging helix canalso be measured by suspending the coils
9、.2.2.3 helix directioncan be left or right-handed depend-ing upon how the wire was coiled, Fig. 1(d)2.2.3.1 left-handed helixthe wire is coiled in a counter-clockwise direction, Fig. 2(a).2.2.3.2 right-handed helixthe wire is coiled in a clockwisedirection, Fig. 2(b).2.2.4 camberthe deflection in th
10、e width direction of a flator shaped wire, Fig. 1(e).3. Summary of Test Method3.1 The maximum diameter of at least one complete cir-cumference is measured using a linear scale while it is restingcompletely flat on a flat surface such as a table, workbench, orfloor (cast measurement). The maximum lif
11、t of the free end ofthe wire when laid on a flat surface is the free end lift helix andis measured using a linear scale. A hanging helix can bemeasured using a linear scale while the wire is being suspended(hanging helix free end lift). Camber is the offset in the widthdimension of a flat or shaped
12、wire and can also be measuredusing a linear scale while the wire is resting on a flat surface.Alternatively, a coordinate measurement machine or opticalcomparator may be used. Helix direction is the direction whichthe wire has been coiled.4. Significance and Use4.1 The process of coiling wire causes
13、 the wire to take on acurvature from the process of being mechanically deformedinto a coiled geometry. The curvature in the wire is permanentunless the wire is straightened. It will affect how the coiledwire will react when it is subjected to additional wire formingoperations. In addition, residual
14、stresses induced from thecoiling operation can cause elastic recoil or spring back insubsequent wire forming operations unless the material isstraightened and stress relieved prior to forming. These re-sidual stresses can create wide variations in the dimensions ofcomponents and or parts that have b
15、een built using the coiledwire (cast).1This test method is under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved Oct. 1, 2013. Published October 2013. Original
16、lyapproved in 2009. Last previous edition approved in 2009 as F2754/F2754M 09.DOI: 10.1520/F2754_F2754M-09R13.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.2 The direction that the wire has been coiled affects howthe wire will be
17、taken off of the coil for subsequent wireforming operations (helix direction).4.3 Lift or spacing between adjacent coils also affects howthe wire will be taken off of the coil and can also affect thedimensions of components and or parts that have been builtusing the coiled wire due to residual stres
18、ses (helix).5. Apparatus5.1 A linear scale is required for measurement of the cast ofcoiled wire. An alternate approach is to use a profilometer orcoordinate measurement machine.5.2 A template is useful for measuring the cast of coiledwire that is greater than 0.6 m (24 in), Fig. 3. An alternativeap
19、proach is to use a coordinate measurement machine oroptical comparato5.3 Special equipment is not required for measurement ofcamber, helix, or the direction of helix.6. Test Specimen6.1 Test Specimen for Measuring Cast:6.1.1 It is preferred to fixture the spool or coil in a suitablemanner so it does
20、 not move as the wire is removed. It isimportant to keep a firm tension on the wire to prevent tanglingwhich could alter the measurement.6.1.2 Scrap the first 1.5 to3m(5to10ft)ofwire in orderto eliminate any possible end effects and wire damage frombeing tied off. Ensure that the wire is easily comi
21、ng off of thespool or coil without any crossing over of the wire.6.1.3 Allow the uncoiled wire to naturally form at least onecomplete circle that rests completely flat on a level surfacesuch as a smooth table, bench, or the floor.NOTE 1(a) Definition of cast. (b) Definition of helix as measured by l
22、ift method. (c) Definition of helix as measured by hanging method. (d) Definitionof helix direction. (e) Definition of camber.FIG. 1 Standard Definitions for Cast, Helix, and Camber of Spooled WireNOTE 1(a) Definition of left-handed helix direction. (b) Definition of right-handed helix direction.FIG
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF2754F2754M20092013STANDARDTESTMETHODFORMEASUREMENTOFCAMBERCASTHELIXANDDIRECTIONOFHELIXOFCOILEDWIRE

链接地址:http://www.mydoc123.com/p-539396.html