ASTM E1504-2006 Standard Practice for Reporting Mass Spectral Data in Secondary Ion Mass Spectrometry (SIMS)《次级离子质谱法(SIMS)中质谱数据报告的标准实施规程》.pdf
《ASTM E1504-2006 Standard Practice for Reporting Mass Spectral Data in Secondary Ion Mass Spectrometry (SIMS)《次级离子质谱法(SIMS)中质谱数据报告的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1504-2006 Standard Practice for Reporting Mass Spectral Data in Secondary Ion Mass Spectrometry (SIMS)《次级离子质谱法(SIMS)中质谱数据报告的标准实施规程》.pdf(2页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1504 06Standard Practice forReporting Mass Spectral Data in Secondary Ion MassSpectrometry (SIMS)1This standard is issued under the fixed designation E 1504; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year o
2、f last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides the minimum information neces-sary to describe the instrumental, experimental, and datareduc
3、tion procedures used in acquiring and reporting second-ary ion mass spectrometry (SIMS) mass spectral data.1.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety a
4、nd health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 673 Terminology Relating to Surface Analysis3. Terminology3.1 DefinitionsFor definitions of terms used in this prac-tice, refer to Terminology E 673.4. Summary of
5、Practice4.1 Experimental conditions and reporting procedures thataffect SIMS mass spectral data are presented in order tostandardize the reporting of such data to facilitate comparisonswith other laboratories and analytical techniques.5. Significance and Use5.1 This practice is intended for use in r
6、eporting theexperimental and data reduction procedures described in otherpublications.6. Information to be Reported6.1 Instrumentation:6.1.1 If a standard commercial SIMS instrument is used,specify the manufacturer, model number, and type of analyzerused. Specify the manufacturer and model number of
7、 anyaccessory or auxiliary equipment that would affect the datacontained within the mass spectrum (for example, additionalvacuum pumping attachments, primary ion mass filter, primaryion sources, electron flood guns, etc.). If any nonstandardmodification has been made to the instrumentation, describe
8、the modification in detail.6.1.2 If a noncommercial SIMS system is used, specify thecomponents composing the system (for example, ion gun,pumping system, vacuum chamber, and mass filter). Specifythe manufacturer and model number if the components are ofcommercial origin. If the components are home-b
9、uilt, specifythem in such detail that their potential effect on the obtainedmass spectrum may be deduced by an individual experiencedin SIMS and vacuum technology.6.2 SpecimenDescribe the specimen in as much detail aspossible. Such factors would include, but are not limited to,sample preparation and
10、 handling, sample history, bulk andtrace composition, physical dimensions, sample homogeneity,crystallinity, and any preanalysis cleaning procedure used.Describe in detail the method of sample mounting. Describeany conductive coating or grids placed on the sample forcharge compensation. If a subsrat
11、e is used, include substratecomposition, purity, and any methods of cleaning.6.3 Experimental Conditions:6.3.1 Primary Ion Source and Ion Optical ColumnIf acommercial ion source is being used, then the manufacturerand model number should be specified. If the ion source is acustom design, the it shou
12、ld be described in detail andappropriate literature references given, if applicable. Thespecies extracted from the ion source must be specified. If thePrimary Ion Column provides mass filtering, then the selectedmass-filer species must be specified. If there is no mass-filtering, then the purity of
13、the material used for ion productionmust be specified. State the ion energy and the impact energyof the primary beam. State the angle of incidence of theprimary ion beam with respect to the surface normal of thesample as well as the ion current (and the method by which thisis measured). State whethe
14、r the ion beam is rastered and the1This practice is under the jurisdiction of ASTM Committee E42 on SurfaceAnalysis and is the direct responsibility of Subcommittee E42.06 on SIMS.Current edition approved Nov. 1, 2006. Published November 2006. Originallyapproved in 1992. Last previous edition approv
15、ed in 2001 as E 1504 92 (2001).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International,
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME15042006STANDARDPRACTICEFORREPORTINGMASSSPECTRALDATAINSECONDARYIONMASSSPECTROMETRYSIMS 次级 离子 质谱法

链接地址:http://www.mydoc123.com/p-528826.html