ASTM E1106-2012 Standard Test Method for Primary Calibration of Acoustic Emission Sensors《声发射传感器一级校准的标准实施规程》.pdf
《ASTM E1106-2012 Standard Test Method for Primary Calibration of Acoustic Emission Sensors《声发射传感器一级校准的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1106-2012 Standard Test Method for Primary Calibration of Acoustic Emission Sensors《声发射传感器一级校准的标准实施规程》.pdf(13页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1106 12Standard Test Method forPrimary Calibration of Acoustic Emission Sensors1This standard is issued under the fixed designation E1106; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A n
2、umber in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the requirements for the abso-lute calibration of acoustic emission (AE) sensors. The cali-bration yields the f
3、requency response of a transducer to waves,at a surface, of the type normally encountered in acousticemission work. The transducer voltage response is determinedat discrete frequency intervals of approximately 10 kHz up to1 MHz. The input is a given well-established dynamic dis-placement normal to t
4、he mounting surface. The units of thecalibration are output voltage per unit mechanical input (dis-placement, velocity, or acceleration).1.2 UnitsThe values stated in SI units are to be regardedas standard. No other units of measurement are included in thisstandard.1.3 This standard does not purport
5、 to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E114 P
6、ractice for Ultrasonic Pulse-Echo Straight-BeamContact TestingE494 Practice for Measuring Ultrasonic Velocity in Mate-rialsE650 Guide for Mounting Piezoelectric Acoustic EmissionSensorsE1316 Terminology for Nondestructive Examinations3. Terminology3.1 Refer to Terminology E1316 for terminology used
7、inthis test method.4. Significance and Use4.1 Transfer StandardsOne purpose of this test method isfor the direct calibration of displacement transducers for use assecondary standards for the calibration ofAE sensors for use innondestructive evaluation. For this purpose, the transfer stan-dard should
8、 be high fidelity and very well behaved andunderstood. If this can be established, the stated accuracyshould apply over the full frequency range up to 1 MHz.NOTE 1The stated accuracy applies only if the transfer standardreturns to quiescence, following the transient input, before any wavereflected f
9、rom the boundary of the calibration block returns to the transferstandard (;100 s). For low frequencies with periods on the order of thetime window, this condition is problematical to prove.4.2 Applications SensorsThis test method may also beused for the calibration ofAE sensors for use in nondestru
10、ctiveevaluation. Some of these sensors are less well behaved thandevices suitable for a transfer standard. The stated accuracy forsuch devices applies in the range of 100 kHz to 1 MHz andwith less accuracy below 100 kHz.5. General Requirements5.1 A primary difficulty in any calibration of a mechanic
11、al/electrical transduction device is the determination of themechanical-motion input to the device. To address this diffi-culty, this calibration procedure uses (i) a standard transducerwhose absolute sensitivity is known from its design andphysical characteristics; and also (ii) a source that produ
12、cesmotion that approximates a waveform calculable from theory.The use of two independent sources of information confers adegree of redundancy that is employed to confirm the validityof the measurements and quantify the experimental errors.Briefly stated, the sensitivity of the transfer standard (or
13、othersensor under test) is determined by comparison with thestandard transducer, while knowledge of a part of the theoreti-cal waveform is used as a check.1This test method is under the jurisdiction of ASTM Committee E07 onNondestructive Testing and is the direct responsibility of Subcommittee E07.0
14、4 onAcoustic Emission Method.Current edition approved June 15, 2012. Published September 2012. Originallyapproved in 1986. Last previous edition approved in 2007 as E1106 - 07. DOI:10.1520/E1106-12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service
15、at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-29
16、59, United States.5.2 Test Block and Mechanical InputThe mechanicalinput to the sensors is obtained by pressing a glass capillarydown onto the surface of a large test block until it breaks. Thereasons for selecting this approach are: (a) capillary breaks arelocalized and short in duration, like natu
17、ral acoustic emissionevents; and (b) use of a large block simplifies wave propaga-tion and makes sensor output less dependent on arbitraryfeatures of block geometry.5.2.1 Prior to the fracture of the glass capillary, the force itexerts on the surface is distributed over an area on the order of2mm3 0
18、.3 mm.3When the glass capillary breaks, the forceit was applying to the surface is abruptly relieved, within a timeon the order of 0.2 to 0.3 s. Within the limitations arising fromthese finite dimensions, the breaking of the capillary approxi-mates a step force function at a point on the surface of
19、theblock. Theoretical solutions for the idealized response of ahalf-space to a normal point-force step function in time appliedto the surface are available.4,5The outputs of flat-responsetransducers have been found to be a good match (except for theinfinite amplitude part) to the theoretical wavefor
20、ms, support-ing the use of this theory as a check on the primary calibrationof sensors. An example with a flat response transducer isshown in Figure 9. The vertical component of the theoreticalwaveform comprises three parts: (a) a low-amplitude responsebeginning at time d/cL, where d is the distance
21、 from the sourceand cLis the longitudinal wave velocity; (b) a short impulsiveresponse between times d/cSand d/cR, where cSis the shearwave velocity and cRthe Rayleigh wave velocity; (c) a stepfunction beginning at d/cR. It is the last of these that is salientfor checking the sensor calibration. The
22、 theoretical height6(shelf value see Figure 9 for determination of the shelf value,relative to zero displacement) of this displacement step u3is:u35 F0A/4pdA21!where F0is the applied force (which is measured), is theshear modulus (calculated by use of the shear wave velocity) ofthe test block, A =(c
23、L/cS)2and d is the distance from the sourceto the transducer.5.3 Absolute Displacement MeasurementAn absolutemeasurement of the dynamic normal surface displacement ofthe block is required for this calibration test method. Thetransducer used for this measurement is a standard transduceragainst which
24、the device under test is compared. The standardtransducer should meet or exceed the performance of thecapacitive transducer described by Breckenridge and Greens-pan.7The important characteristics of the standard transducerinclude high fidelity, high sensitivity, and operating character-istics amenab
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME11062012STANDARDTESTMETHODFORPRIMARYCALIBRATIONOFACOUSTICEMISSIONSENSORS 声发 传感器 一级 校准 标准 实施 规程 PDF

链接地址:http://www.mydoc123.com/p-527917.html