ASTM E388-2004(2009) 2500 Standard Test Method for Wavelength Accuracy and Spectral Bandwidth of Fluorescence Spectrometers 《荧光分光计的光波波长精密度和光谱带宽的测试方法》.pdf
《ASTM E388-2004(2009) 2500 Standard Test Method for Wavelength Accuracy and Spectral Bandwidth of Fluorescence Spectrometers 《荧光分光计的光波波长精密度和光谱带宽的测试方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E388-2004(2009) 2500 Standard Test Method for Wavelength Accuracy and Spectral Bandwidth of Fluorescence Spectrometers 《荧光分光计的光波波长精密度和光谱带宽的测试方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E388 04 (Reapproved 2009)Standard Test Method forWavelength Accuracy and Spectral Bandwidth ofFluorescence Spectrometers1,2This standard is issued under the fixed designation E388; the number immediately following the designation indicates the year oforiginal adoption or, in the case of
2、 revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the testing of the spectralbandwidth and wavelength accuracy of fluores
3、cence spectrom-eters that use a monochromator for emission wavelengthselection and photomultiplier tube detection. This test methodcan be applied to instruments that use multi-element detectors,such as diode arrays, but results must be interpreted carefully.This test method uses atomic lines between
4、 250 nm and 1000nm.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standar
5、d to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Summary of Test Method2.1 The difference between the apparent wavelength and theknown wavelength for a series of atomic emission lines is usedas a test for wavelength ac
6、curacy. The apparent width of someof these lines is used as a test for spectral bandwidth.3. Apparatus3.1 Fluorescence Spectrometer to be tested.3.2 Atomic Discharge Lamps, Low-pressure, sufficientlysmall to be placed in the sample cell holder of the instrument.4. Reagent4.1 Scattering SuspensionDis
7、solve1gofglycogen perlitre of water, or use a dilute microsphere suspension contain-ing 1 mL of a commercially available, concentrated micro-sphere suspension.5. Procedure5.1 The emission lines given for mercury (Hg), neon (Ne),argon (Ar), krypton (Kr), and xenon (Xe) in Table 1 aretypically observa
8、ble using standard commercial fluorometers,although some of them may be too weak to detect on someinstruments.5.1.1 Most fluorescence instruments will not be able toresolve very closely spaced lines such as those for Hg at 312.57nm, 313.15 nm, and 313.18 nm, due to the relatively lowresolution monoc
9、hromators used in fluorescence equipmentcompared to those used in absorbance spectrometers. Evenlower resolution fluorometers may not resolve lines separatedby less than several nanometres such as those for Hg at 404.66and 407.78, or at 576.96 and 579.07 nm.5.1.2 In instruments using blazed grating
10、monochromators,additional weaker lines are found due to second order diffrac-tion of atomic lines. For instance, lines appear for Hg at 507.30and 593.46 nm, arising from the 253.65 and 296.73 nm lines,respectively.5.2 Calibration and Adjustment of Emission Monochroma-tor:5.2.1 With an atomic arc sou
11、rce properly aligned (see 5.3)inthe sample cell compartment, adjust the position of thewavelength dial to give maximum signal for each of the atomiclines and record the wavelength reading. The difference be-tween the observed value and the corresponding value in Table1 represents the correction that
12、 must be subtracted algebra-ically from the wavelength reading of the instrument. Thecorrections may be recorded or the monochromator adjusted togive the proper values. Since there may be some backlash inthe wavelength drive of scanning instruments, always approachthe peak position from the same dir
13、ection, if applicable.1This test method is under the jurisdiction of ASTM Committee E13 onMolecular Spectroscopy and Separation Science and is the direct responsibility ofSubcommittee E13.01 on Ultra-Violet, Visible, and Luminescence Spectroscopy.Current edition approved Oct. 1, 2009. Published Octo
14、ber 2009. Originallyapproved in 1969. Last previous edition approved in 2004 as E388 04. DOI:10.1520/E0388-04R09.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the
15、 standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.2.2 When calibrating scanning-type instruments, approachthe peak position in the same direction that the motor scans, ifyour instru
16、ment does not correct for backlash. Check theposition against that recorded while scanning and, if necessary,correct as in 5.2.1.5.3 In cases where the monochromator is designed so that alateral displacement of the calibration source from a positiondirectly in front of the entrance slit appears as a
17、 wavelengthshift, proceed as follows:5.3.1 Instead of placing the atomic lamp in front of theentrance slit of the monochromator, fill a sample cell with adilute scattering suspension, as described in 4.1.5.3.2 Place the cell in the sample position in the instrument.5.3.3 Illuminate the cell transver
18、sely with the atomic lamp,either from the side or from above.5.3.4 Adjust the wavelength to give the maximum signal foreach of the atomic lines given in Table 1; record the wave-length reading and proceed as in 5.2.5.4 Adjustment of Excitation Monochromator:5.4.1 After the emission monochromator has
19、 been cali-brated, adjust the excitation monochromator to match, asfollows:5.4.2 Place a sample cell containing either the dilute scat-tering suspension described in 4.1 in the sample cell compart-ment.5.4.3 With a continuous source in the normal source posi-tion of the instrument, illuminate the su
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME388200420092500STANDARDTESTMETHODFORWAVELENGTHACCURACYANDSPECTRALBANDWIDTHOFFLUORESCENCESPECTROMETERS

链接地址:http://www.mydoc123.com/p-527526.html