ASTM D6689-2001(2011) 5000 Standard Guide for Optimizing Controlling and Reporting Test Method Uncertainties from Multiple Workstations in the Same Laboratory Organization《来自同一实验室机.pdf
《ASTM D6689-2001(2011) 5000 Standard Guide for Optimizing Controlling and Reporting Test Method Uncertainties from Multiple Workstations in the Same Laboratory Organization《来自同一实验室机.pdf》由会员分享,可在线阅读,更多相关《ASTM D6689-2001(2011) 5000 Standard Guide for Optimizing Controlling and Reporting Test Method Uncertainties from Multiple Workstations in the Same Laboratory Organization《来自同一实验室机.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D6689 01 (Reapproved 2011)Standard Guide forOptimizing, Controlling and Reporting Test MethodUncertainties from Multiple Workstations in the SameLaboratory Organization1This standard is issued under the fixed designation D6689; the number immediately following the designation indicates
2、the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide describes a protocol for optimizing,
3、control-ling, and reporting test method uncertainties from multipleworkstations in the same laboratory organization. It does notapply when different test methods, dissimilar instruments, ordifferent parts of the same laboratory organization functionindependently to validate or verify the accuracy of
4、 a specificanalytical measurement.1.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory requir
5、ements prior to use.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD6091 Practice for 99 %/95 % Interlaboratory DetectionEstimate (IDE) for Analytical Methods with NegligibleCalibration ErrorD6512 Practice for Interlaboratory Quantitation EstimateE135 Terminology Relat
6、ing to Analytical Chemistry forMetals, Ores, and Related MaterialsE415 Test Method for Atomic Emission Vacuum Spectro-metric Analysis of Carbon and Low-Alloy SteelE1763 Guide for Interpretation and Use of Results fromInterlaboratory Testing of Chemical Analysis MethodsSTP 15D ASTM Manual on Presenta
7、tion of Data andControl Chart Analysis, Prepared by Committee E11 onStatistical Methods2.2 Other Documents:ISO 17025 (previously ISO Guide 25) General Require-ments for the Competence of Calibration and TestingLaboratories33. Terminology3.1 DefinitionsFor definitions of terms used in this Guide,refe
8、r to Terminology E135 and D1129.3.2 Defintions of Terms Specific to This Standard:3.2.1 laboratory organizationa business entity that pro-vides similar types of measurements from more than oneworkstation located in one or more laboratories, all of whichoperate under the same quality system.NOTE 1Key
9、 aspects of a quality system are covered in ISO 17025 andinclude documenting procedures, application of statistical control tomeasurement processes and participation in proficiency testing.3.2.2 maximum deviationthe maximum error associatedwith a report value, at a specified confidence level, for a
10、givenconcentration of a given element, determined by a specificmethod, throughout a laboratory organization.3.2.3 measurement quality objectivesa model used by thelaboratory organization to specify the maximum error associ-ated with a report value, at a specified confidence level.3.2.4 workstationa
11、combination of people and equipmentthat executes a specific test method using a single specifiedmeasuring device to quantify one or more parameters, witheach report value having an established estimated uncertaintythat complies with the measurement quality objectives of thelaboratory organization.4.
12、 Significance and Use4.1 Many analytical laboratories comply with acceptedquality system requirements such as NELAC chapter 5 (seeNote 2) and ISO 17025. When using standard test methods,their test results on the same sample should agree with thosefrom other similar laboratories within the reproducib
13、ilityestimates (R2) published in the standard. Reproducibilityestimates are generated during the standardization process aspart of the interlaboratory studies (ILS). Many laboratoriesparticipate in proficiency tests to confirm that they perform1This guide is under the jurisdiction of ASTM Committee
14、D19 on Water and isthe direct responsibility of Subcommittee D19.02 on Quality Systems, Specification,and Statistics.Current edition approved May 1, 2011. Published June 2011. Originallyapproved in 2001. Last previous edition approved in 2006 as D6689 01(2006).DOI: 10.1520/D6689-01R11.2For reference
15、d ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St
16、.,4th Floor, New York, NY 10036, http:/www.ansi.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.consistently over time. In both ILS and proficiency testingprotocols, it is generally assumed that only one workstation isused to gen
17、erate the data (see 6.5.1).NOTE 2NELAC chapter 5 allows the use of a Work Cell wheremultiple instruments/operators are treated as one unit: the performance ofthe Work Cell is tracked rather than each workstation independently. Thisguide is intended to go beyond the Work Cell to achieve the benefits
18、ofmonitoring workstations independently.4.2 Many laboratories have workloads and/or logisticalrequirements that dictate the use of multiple workstations.Some have multiple stations in the same area (central labora-tory format). Others stations are scattered throughout a facility(at-line laboratory f
19、ormat). Often, analysis reports do notidentify the workstation used for the testing, even if worksta-tions differ in their testing uncertainties. Problems can arise ifclients mistakenly attribute variation in report values to processrather then workstation variability. These problems can beminimized
20、 if the laboratory organization sets, complies with,and reports a unified set of measurement quality objectivesthroughout.4.3 This guide can be used to harmonize calibration andcontrol protocols for all workstations, thereby providing thesame level of measurement traceability and control. It stream-
21、lines documentation and training requirements, thereby facili-tating flexibility in personnel assignments. Finally, it offers anopportunity to claim traceability of proficiency test measure-ments to all included workstations, regardless on which work-station the proficiency test sample was tested. T
22、he potentialbenefits of utilizing this protocol increase with the number ofworkstations included in the laboratory organization.4.4 This guide can be used to identify and quantify benefitsderived from corrective actions relating to under-performingworkstations. It also provides means to track improv
23、ed perfor-mance after improvements have been made.4.5 It is a prerequisite that all users of this guide complywith ISO 17025, especially including the use of documentedprocedures, the application of statistical control of measure-ment processes, and participation in proficiency testing.4.6 The gener
24、al principles of this protocol can be adapted toother types of measurements, such as mechanical testing andon-line process control measurements such as temperature andthickness gauging. In these areas, users will likely need toestablish their own models for defining measurement qualityobjectives. Pr
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD6689200120115000STANDARDGUIDEFOROPTIMIZINGCONTROLLINGANDREPORTINGTESTMETHODUNCERTAINTIESFROMMULTIPLEWORKSTATIONSINTHESAMELABORATORYORGANIZATION

链接地址:http://www.mydoc123.com/p-523250.html