ASTM F2743-2011 Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems《涂料药物洗脱血管支架系统检查和急性颗粒表征的标准指南》.pdf
《ASTM F2743-2011 Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems《涂料药物洗脱血管支架系统检查和急性颗粒表征的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2743-2011 Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems《涂料药物洗脱血管支架系统检查和急性颗粒表征的标准指南》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2743 11Standard Guide forCoating Inspection and Acute Particulate Characterization ofCoated Drug-Eluting Vascular Stent Systems1This standard is issued under the fixed designation F2743; the number immediately following the designation indicates the year oforiginal adoption or, in the
2、case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide describes recommended in vitro test proce-dures for coating inspection and ac
3、ute particulate characteriza-tion of coated drug-eluting vascular (balloon-expandable andself-expanding) stent systems.1.2 Recommended practices for coating inspection andacute particulate characterization include baseline (deploy-ment) testing and simulated use testing. This guide describesthe capt
4、ure and analysis of particulates. This guide describesthe inspection of the coated stent surface. This guide wasdeveloped for characterization and not intended for productionrelease testing of coated drug-eluting vascular stent systemsalthough some sections may be appropriate.1.3 Chronic particulate
5、 characterization and coating inspec-tion are not included herein.1.4 Coating systems specifically designed to degrade orotherwise intentionally separate themselves from the perma-nent stent structure may not be fully addressed herein.1.5 The values stated in SI units are to be regarded asstandard.
6、No other units of measurement are included in thisstandard.1.6 The values stated in inch-pound units are to be regardedas standard. The values given in parentheses are mathematicalconversions to SI units that are provided for information onlyand are not considered standard.2. Referenced Documents2.1
7、 Other Standards:USP Particulate Matter in Injections2FDA Guidance for Industry and FDA Staff Non-ClinicalEngineering Tests and Recommended Labeling for Intra-vascular Stents and Associated Delivery Systems, April18, 20103AAMI TIR42:2010 Evaluation of Particulates Associatedwith Vascular Medical Dev
8、ices43. Terminology3.1 Definitions:3.1.1 mock vesselphysical simulation of the vasculature atthe intended clinical deployment site.3.1.2 stent systema system comprised of a vascular stentand its delivery system.3.1.3 trackingnavigation of a guide wire, guide catheter,and/or stent system through eith
9、er actual or simulated vascularanatomy.3.1.4 tracking fixturea model that simulates or replicatesthe geometry of a representative vasculature through which thestent system will be passed.3.2 Definitions of Terms Specific to This Standard:3.2.1 acutea test timeframe intended to include stentdelivery
10、and deployment beginning with the initial insertion ofstent system until full removal of the delivery system and itsaccessory devices.3.2.2 baselinecoating inspection and acute particulatecharacterization after stent expansion to the desired diameter inan unconstrained environment and without tracki
11、ng.3.2.3 chronica test timeframe intended to mimic theimplantation time after full removal of the delivery system andits accessory devices.3.2.4 constrained environmenta deployment site in whichthe stent is deployed into a mock vessel.3.2.5 simulated usecoating inspection and acute particu-late char
12、acterization after tracking in simulated anatomy andaqueous environment. It may also include deployment in bentconfiguration, deployment in overlapped configuration, post-dilatation, or other scenarios that can reasonably be expected inclinical use.3.2.6 unconstrained environmenta deployment site in
13、which the stent is not constrained by a mock vessel. Compareto “Constrained Environment”.4. Summary of Practice4.1 Test Sequence and SamplesBaseline and SimulatedUse Testing are conducted as two separate tests. Coating1This guide is under the jurisdiction of ASTM Committee F04 on Medical andSurgical
14、 Materials and Devices and is the direct responsibility of SubcommitteeF04.30 on Cardiovascular Standards.Current edition approved Oct. 1, 2011. Published November 2011. DOI:10.1520/F2743-11.2Available from U.S. Pharmacopeia (USP), 12601 Twinbrook Pkwy., Rockville,MD 20852-1790, http:/www.usp.org.3A
15、vailable from Food and Drug Administration (FDA), 10903 New HampshireAve., Silver Spring, MD 20993-0002, http:/www.fda.gov.4Available from Association for the Advancement of Medical Instrumentation(AAMI), 4301 North Fairfax Dr., Suite 301, Arlington, VA 22203-1633.1Copyright ASTM International, 100
16、Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.inspection and acute particulate characterization may be per-formed as two separate tests with independent samples.4.2 Baseline TestingA single stent is deployed to nominalor maximum labeled diameter. The stent is expand
17、ed in anunconstrained environment so as to characterize the stent only.Baseline testing includes coating inspection and acute particu-late characterization of the stent. Baseline coating inspectionmay be conducted after deployment in air or in an aqueousunconstrained environment. Baseline acute part
18、iculate charac-terization should be conducted in an aqueous unconstrainedenvironment. The surfaces of the stent coating are inspected fordefects or other adverse attributes caused by this procedure.Cumulative particulates released are captured or continuouslymonitored, counted and classified accordi
19、ng to size ranges.4.2.1 Particles released may be captured in a receptacle andsampled for count/size using light obscuration or filtration/microscopy, or4.2.2 Particles released may be acquired and continuouslycounted in an apparatus (for example, tube) for facilitatingflow.4.3 Simulated Use Testing
20、The stent system is tracked inan aqueous environment, through an appropriately clean, invitro model simulating the vascular anatomy to be navigated toaccess the targeted clinical deployment site. Accessory devices(for example, guidewires, guide catheters, and so forth) areutilized as indicated in th
21、e IFU. The stent is deployed eithersingly or overlapped with another stent and bent configurationto represent worst-case clinical condition, as appropriate. Aconstrained environment should be used as the deploymentlocation. Stents should be expanded in accordance with theIFU, including expansion to
22、post-dilatation limits, as appropri-ate. Cumulative particulates released from the stent(s), stentcoating(s), stent system(s) and accessory devices (if used)during the procedure are captured or continuously monitored,counted and classified according to size ranges. Particulatecharacterization may be
23、 necessary to aid in classifying poten-tial particulate sources, and the test developer should under-stand the constituents of the coated stent system. The surfacesof the stent coating(s) are inspected for defects or other adverseattributes caused by this procedure. Analysis of particulatesand surfa
24、ce inspection may be accomplished using the sametest articles subjected to tracking and deployment, if appropri-ate.4.3.1 Particles released may be captured in a collectionbeaker and sampled for count/size using light obscuration orfiltration/microscopy. The need for post-dilatation, overlappingor t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF27432011STANDARDGUIDEFORCOATINGINSPECTIONANDACUTEPARTICULATECHARACTERIZATIONOFCOATEDDRUGELUTINGVASCULARSTENTSYSTEMS

链接地址:http://www.mydoc123.com/p-539373.html