ASTM E2069-2006(2012) Standard Test Method for Temperature Calibration on Cooling of Differential Scanning Calorimeters《差示扫描热量计制冷却温度校准用标准试验方法》.pdf
《ASTM E2069-2006(2012) Standard Test Method for Temperature Calibration on Cooling of Differential Scanning Calorimeters《差示扫描热量计制冷却温度校准用标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2069-2006(2012) Standard Test Method for Temperature Calibration on Cooling of Differential Scanning Calorimeters《差示扫描热量计制冷却温度校准用标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2069 06 (Reapproved 2012)Standard Test Method forTemperature Calibration on Cooling of Differential ScanningCalorimeters1This standard is issued under the fixed designation E2069; the number immediately following the designation indicates the year oforiginal adoption or, in the case of
2、 revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the temperature calibration ofdifferential scanning calorimeters on coo
3、ling using the differ-ence between transition temperatures upon heating and coolingin the temperature range of 50 to 185C. An offset in theindicated temperature between heating and coolingexperiments, within this temperature range, may be used toprovide temperature calibration on cooling at other te
4、mperatureranges.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard t
5、o establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific precau-tionary statements are given in Section 6.2. Referenced Documents2.1 ASTM Standards:2D3418 Test Method for Transition Temperatures and En-thalpies of Fusion a
6、nd Crystallization of Polymers byDifferential Scanning CalorimetryE473 Terminology Relating to Thermal Analysis and Rhe-ologyE794 Test Method for MeltingAnd Crystallization Tempera-tures By Thermal AnalysisE928 Test Method for Purity by Differential Scanning Calo-rimetryE967 Test Method for Temperat
7、ure Calibration of Differen-tial Scanning Calorimeters and Differential Thermal Ana-lyzersE1970 Practice for Statistical Treatment of ThermoanalyticalData3. Terminology3.1 Specific technical terms used in this test method aredefined in Terminology E473.4. Summary of Test Method4.1 The temperature se
8、nsor of the DSC, used to determinethe temperature of a transition, is located close to but externalto the test specimen. This arrangement causes the indicatedtemperature to lead or lag the actual specimen temperature onheating/cooling causing the reported temperature to be higheron heating and lower
9、 on cooling than the actual specimentransition temperature. A DSC apparatus temperature, that hasbeen calibrated for heating experiments, needs to be re-calibrated for cooling experiments. Such a calibration oncooling is performed using a liquid crystal material, thetransition(s) for which are not s
10、ubject to super-heating orsuper-cooling.4.2 The transition temperature of a rapid, non-superheatingand non-supercooling transition is determined upon heatingand upon cooling. The difference between these two indicatedtemperatures provides an offset temperature value betweenheating and cooling experi
11、ments at the indicated rate. Thisoffset temperature value, when used with a precise temperaturecalibration upon heating, may serve as an instrument calibra-tion function upon cooling.5. Significance and Use5.1 This test method is useful in calibrating the temperaturesignal of a differential scanning
12、 calorimeter for cooling experi-ments such as the determination of crystallization temperaturesin Test Method D3418 and Test Method E794.5.2 This test method may be used for research,development, analytical, specification acceptance, quality as-surance and control purposes.6. Precautions6.1 Toxic or
13、 corrosive effluents, or both, may be releasedwhen heating the material of this test method and may beharmful to personnel and to the apparatus.1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on Calo-rime
14、try and Mass Loss.Current edition approved Sept. 1, 2012. Published September 2012. Originallyapproved in 2000. Last previous edition approved in 2006 as E2069 06. DOI:10.1520/E2069-06R12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at service
15、astm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States17. Apparatus7.1 Differential Scanning Calorimeter (DSC)The
16、essentialinstrumentation required providing the minimum differentialscanning calorimeter capability for this test method includes:7.1.1 A DSC Test Chamber, composed of:7.1.1.1 A Furnace(s), to provide uniform controlled heatingand cooling of a specimen and reference material to a constanttemperature
17、 or at a constant rate within the applicable tempera-ture range of this method.7.1.1.2 A Temperature Sensor, that indicates specimen orfurnace temperature to 60.01 C.7.1.1.3 A Differential Sensor, to detect a heat flow difference(DSC) between the specimen and reference with a range of atleast 100 mW
18、 readable to 61 W (DSC).7.1.1.4 A means of sustaining a purge gas rate of 10 to 1006 5 mL/min in the test chamber.NOTE 1Typically inert purge gases that inhibit specimen oxidation are99+ % pure nitrogen, argon or helium. Subambient operation requires drypurge gases. Dry gases are recommended for all
19、 experiments unless theeffect of moisture is part of the study.7.1.2 A Temperature Controller, capable of executing aspecific temperature program by operating the furnace orfurnaces between selected temperature limits at a rate oftemperature change of 10 C/min constant to within 60.1C/min or at an i
20、sothermal temperature constant to 60.1C.7.1.3 A Recording Device, capable of recording and display-ing fractions of the heat flow signal (DSC curve), including thesignal noise, on the Y-axis versus fractions of temperaturesignal, including the signal noise, on the X-axis.7.1.4 Containers, (pans, cru
21、cibles, vials, lids, closures,seals, etc.) that are inert to the specimen and referencematerials and that are of suitable structural shape and integrityto contain the specimen and reference in accordance with therequirements of this test method.NOTE 2DSC containers are commonly composed of aluminum
22、orother inert material of high thermal conductivity. Aluminum has beentested and found compatible with the materials used in this standard.7.1.5 Cooling Capability, at constant cooling rates of up to10C/min in the temperature range of 185 to 50C, to hastencool down from elevated temperatures, or to
23、sustain anisothermal subambient temperature, or both.7.2 A Balance, to weigh specimen and/or containers to 610g with a capacity of 100 mg or greater.8. Calibration Materials8.1 For the temperature range covered by manyapplications, the liquid crystal transitions of 99.8 to 99.9 %pure materials in Ta
24、ble 1 may be used for calibration. Thecalibrating liquid crystal materials3are known as M-24, BP-53and BCH-52.NOTE 3The purity of these liquid crystal materials may be determinedby Test Method E928 using the first liquid crystal transition prior to use(see Table 2).3The sole source of supply of thes
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME206920062012STANDARDTESTMETHODFORTEMPERATURECALIBRATIONONCOOLINGOFDIFFERENTIALSCANNINGCALORIMETERS

链接地址:http://www.mydoc123.com/p-530200.html