ASTM D5827-1995(2002) Standard Test Method for Analysis of Engine Coolant for Chloride and Other Anions by Ion Chromatography《用离子色谱法分析氯化物和其它阴离子的发动机冷却剂的标准试验方法》.pdf
《ASTM D5827-1995(2002) Standard Test Method for Analysis of Engine Coolant for Chloride and Other Anions by Ion Chromatography《用离子色谱法分析氯化物和其它阴离子的发动机冷却剂的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5827-1995(2002) Standard Test Method for Analysis of Engine Coolant for Chloride and Other Anions by Ion Chromatography《用离子色谱法分析氯化物和其它阴离子的发动机冷却剂的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 5827 95 (Reapproved 2002)Standard Test Method forAnalysis of Engine Coolant for Chloride and Other Anionsby Ion Chromatography1This standard is issued under the fixed designation D 5827; the number immediately following the designation indicates the year oforiginal adoption or, in the
2、 case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the chemical analysis of enginecoolant for chloride ion by hig
3、h-performance ion chromatog-raphy (HPIC). Several other common anions found in enginecoolant can be determined in one chromatographic analysis bythis test method.1.2 This test method is applicable to both new and usedengine coolant.1.3 Coelution of other ions may cause interferences for anyof the li
4、sted anions. In the case of unfamiliar formulations,identification verification should be performed by either orboth fortification and dilution of the sample matrix with theanions of interest.1.4 Analysis can be performed directly by this test methodwithout pretreatment, other than dilution, as requ
5、ired by thelinear ranges of the equipment. Table 1 indicates severalapplicable anions and approximate detection limits.1.5 The values stated in SI units are to be regarded as thestandard. The values given in parentheses are for informationonly.1.6 This standard does not purport to address all of the
6、safety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to its use.2. Referenced Documents2.1 ASTM Standards:D 1193 Specification for
7、Reagent Water2D 1176 Test Method for Sampling and Preparing AqueousSolutions of Engine Coolants or Antirusts for TestingPurposes33. Summary of Test Method3.1 A small volume of working sample is prepared bydilution of the sample with the method eluant. This dilutedsample is filtered and pumped throug
8、h two ion exchangecolumns and a suppressor and into a conductivity detector. Ionsare separated based on their affinity for exchange sites of theresin with respect to the resins affinity for the eluant. Thesuppressor increases the sensitivity of the method by bothincreasing the conductivity of the an
9、alytes and decreasing theconductivity of the eluant. The suppressor converts the eluantand the analytes to the corresponding hydrogen form acids.Anions are quantitated by integration of their response com-pared with an external calibration curve and are reported asmilligrams per litre (mg/L).4. Sign
10、ificance and Use4.1 This test method provides for the qualitative and quan-titative determination of common anions in engine coolant inthe milligrams per litre to low percent range and requires onlya few millilitres or microlitres of sample per test, with resultsavailable in less than 30 min. Accept
11、able levels of chloride andother anions vary with manufacturers blending specificationsand applicable ASTM minimum or maximum specifications.5. Interferences5.1 Interferences can be caused by substances with similarretention times, especially if they are in high concentrationcompared to those of the
12、 analyte of interest. Sample dilutionwill be used to minimize or solve most interference problems.5.2 A water dip (solvent system peak) can cause interfer-ence with some integrators. This is eliminated by dilution withthe eluant if the sample dilution factor is 49 + 1 (v/v) or greater.Below this dil
13、ution, it is best to add a spike of eluantconcentrate to the sample such that the sample is not diluted1This test method is under the jurisdiction of ASTM Committee D15 on EngineCoolants and is the direct responsibility of Subcommittee D15.04 on ChemicalProperties.Current edition approved Sept. 10,
14、1995. Published November 1995.2Annual Book of ASTM Standards, Vol 11.01.3Annual Book of ASTM Standards, Vol 15.05.TABLE 1 Analytes and Minimum Detection LimitsAnalyte Detection Limit, mg/LAChloride (Cl) 2.0Nitrite (NO2) 5.0Bromide (Br) 4.0Nitrate (NO3) 7.1o-Phosphate (HPO4)220.0Sulfate (SO4)28.0Oxal
15、ate (C2O4)212.0ADetermined using 100-L sample volume. Sample diluted 99 + 1 (v/v) withchromatographic eluant 30-S/cm full scale, suppressed conductivity detection.Dionex AS4ASC column with AG4ASC guard columns. Other systems will requireMDL determinations using chosen dilution factors, eluants, colu
16、mns, and detector.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.significantly and the resulting test solution matches the eluantused in the system. One method is the addition of 100 L of100X eluant concentrate to 10.0 mL of sample
17、or standard.5.3 Method interferences can be caused by the contamina-tion of glassware, eluant, reagents, etc. Great care must betaken to ensure that contamination, especially by chloride, iskept at the lowest possible levels.5.4 Pre-rinsing of the sample preparation containers withdeionized water is
18、 mandatory.5.5 The use of latex gloves is highly recommended toprevent contamination.6. Apparatus6.1 Analytical Balance, capable of weighing accurately to0.0001 g.6.2 Ion ChromatographAnalytical system with all re-quired accessories including syringes, columns, suppressor,gasses, and detector. Colum
19、n life and performance are en-hanced by the use of a two-eluant channel gradient pump, ifavailable.6.3 Guard Column, for protection of the analytical columnfrom strongly retained constituents. Better separations areobtained with additional plates.6.4 Anion Separator Column, capable of producing anal
20、yteseparation equivalent to or better than that shown in Fig. 1.46.5 Anion Suppressor DeviceMicro membrane suppressoror equivalent.5A cation exchange column in the hydrogen formhas been used successfully, but it will periodically need to beregenerated as required, being indicated by a high backgroun
21、dconductivity and low analyte response.6.6 Conductivity Detector, low volume (2 L) and flow,temperature compensated, capable of at least 0 to 1000 S/cmon a linear scale.6.7 Integrator or Chromatography Data System Software,capable of obtaining approximately the same detection limitsas are listed in
22、Table 1.6.8 Drying Oven, controlled at 105, 150, and 600 6 5C.6.9 Desiccator.7. Reagents7.1 Purity of ReagentsReagent grade or higher puritychemicals shall be used for the preparation of all samples,standards, eluants, and regenerator solutions. Unless otherwiseindicated, it is intended that all rea
23、gents conform to thespecifications of the Committee on Analytical Reagents of theAmerican Chemical Society, where such specification areavailable.6Other grades may be used, provided it is firstascertained that the reagent is of sufficiently high purity topermit its use without lessening the accuracy
24、 of the determi-nation.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean reagent water as definedby Type II of Specification D 1193. It is recommended that allwater be filtered through a 0.2-m filter. For eluant preparation,degas the water by sparging with
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD582719952002STANDARDTESTMETHODFORANALYSISOFENGINECOOLANTFORCHLORIDEANDOTHERANIONSBYIONCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-520807.html