《ASTM D5723-1995(2015) 6112 Standard Practice for Determination of Chromium Treatment Weight on Metal Substrates by X-Ray Fluorescence《采用X射线荧光法测定金属基体上铬处理重量的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5723-1995(2015) 6112 Standard Practice for Determination of Chromium Treatment Weight on Metal Substrates by X-Ray Fluorescence《采用X射线荧光法测定金属基体上铬处理重量的标准实施规程》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5723 95 (Reapproved 2015)Standard Practice forDetermination of Chromium Treatment Weight on MetalSubstrates by X-Ray Fluorescence1This standard is issued under the fixed designation D5723; the number immediately following the designation indicates the year oforiginal adoption or, in th
2、e case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the use of X-ray fluorescence(XRF) techniques for determination o
3、f the coating weight ofchromium treatments on metal substrates. These techniques areapplicable for determination of the coating weight as chro-mium or total coating weight of a chromium-containingtreatment, or both, on a variety of metal substrates.1.2 This standard does not purport to address all o
4、f thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Summary of Practice2.1 ExcitationThe measurement of chromium tr
5、eatmentcoating weights by XRF methods is based on the combinedinteraction of the chromium coating and substrate with anintense beam of primary radiation. Since each element fluo-resces at an energy characteristic of the particular element, thisinteraction results in the generation of X rays of defin
6、ed energy.The primary radiation may be generated by an X-ray tube orderive from a radioisotope.2.2 DetectionThe secondary beam (fluorescent X rays ofthe elements and scattered radiation) is read by a detector thatcan discriminate between the energy levels of fluorescingradiations in the secondary be
7、am. The detection system in-cludes the radiation detector with electronics for pulse ampli-fication and pulse counting.2.3 Basic Principle:2.3.1 A relationship exists between the treatment coatingweight and secondary radiation intensity. This relationship isusually linear within the desired coating
8、weights of thechromium treatments on metal substrates. The measurementsare based on primary standards of known coating weights andinstrument calibration that correlates the secondary radiationintensity with the coating weight quantitatively.2.3.2 The coating weight is determined by measurement ofthe
9、 fluorescent X rays of the coating. The detection system isset to count the number of X rays in an energy region that ischaracteristic of X rays from the element of interest. Theelement of interest in this practice is chromium.2.3.3 If a linear relationship exists, the coating weight andnumber of co
10、unts of X rays of a chromium treatment on aparticular substrate can be expressed by a conversion factorthat represents the number of counts for a particular coatingweight unit/unit area. This is usually expressed in mg/ft2ormg/m2of chromium or total coating weight.2.3.4 The exact relationship betwee
11、n the measured numberof counts and corresponding coating weight must be estab-lished for each individual combination of substrate andchromium-containing treatment. Usually determined by thetreatment supplier, this relationship is established by usingprimary standards having known amounts of the same
12、 treat-ment applied to the same substrate composition as the speci-mens to be measured.2.3.5 Some X-ray apparatus have a data handling systemwhereby a coating weight versus X-ray counts curve may beestablished within the system for the direct readout of coatingweight. If such apparatus does not perm
13、it the entry of aconversion factor as described in 2.3.3, it is calibrated using abare, untreated specimen and a minimum of three specimenswith known coating weights of the treatment and substratecombination of interest. The coating weight to be measuredmust be within the range of these known coatin
14、g weights. Morethan three known specimens must be used if the relationship ofX-ray counts to coating weight is not linear over the range tobe measured. The treatment supplier should be consulted forrecommendations for establishing the curve in the instrumentfor the particular treatment and substrate
15、 combination ofinterest.3. Significance and Use3.1 The procedure described in this practice is designed toprovide a method by which the coating weight of chromiumtreatments on metal substrates may be determined.1This practice is under the jurisdiction of ASTM Committee D01 on Paint andRelated Coatin
16、gs, Materials, and Applications and is the direct responsibility ofSubcommittee D01.53 on Coil Coated Metal.Current edition approved July 1, 2015. Published July 2015. Originally approvedin 1995. Last previous edition approved in 2010 as D5723 95 (2010). DOI:10.1520/D5723-95R15.Copyright ASTM Intern
17、ational, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2 This procedure is applicable for determination of thetotal coating weight and the chromium coating weight of achromium-containing treatment.4. Apparatus and Materials4.1 Measuring Instrument,2which is ca
18、pable of determiningthe coating weights of chromium-containing treatments onmetal substrates by X-ray fluorescence is required. The treat-ment supplier should be consulted for the suitability of theinstrumentation to be used.4.2 Calibration Standard,3necessary to calibrate the instru-ment. The count
19、 value of this standard must be specified by thetreatment supplier.4.3 Treated Coupon, on which the coating weight is to bedetermined must be cut to the required size for the instrumentfrom the treated substrate.4.4 Blank (Bare and Untreated) Coupon should be a sampleof the same metal substrate on w
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD5723199520156112STANDARDPRACTICEFORDETERMINATIONOFCHROMIUMTREATMENTWEIGHTONMETALSUBSTRATESBYXRAYFLUORESCENCE

链接地址:http://www.mydoc123.com/p-520521.html