ASTM D1823-1995(2001) Standard Test Method for Apparent Viscosity of Plastisols and Organosols at High Shear Rates by Extrusion Viscometer《用挤压粘度计测定在高速剪切时塑料溶胶和有机溶胶的表观粘度的标准试验方法》.pdf
《ASTM D1823-1995(2001) Standard Test Method for Apparent Viscosity of Plastisols and Organosols at High Shear Rates by Extrusion Viscometer《用挤压粘度计测定在高速剪切时塑料溶胶和有机溶胶的表观粘度的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D1823-1995(2001) Standard Test Method for Apparent Viscosity of Plastisols and Organosols at High Shear Rates by Extrusion Viscometer《用挤压粘度计测定在高速剪切时塑料溶胶和有机溶胶的表观粘度的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 1823 95 (Reapproved 2001)Standard Test Method forApparent Viscosity of Plastisols and Organosols at HighShear Rates by Extrusion Viscometer1This standard is issued under the fixed designation D 1823; the number immediately following the designation indicates the year oforiginal adopti
2、on or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the measurement of plastisoland organosol viscosi
3、ty at high shear rates by means of anextrusion viscometer.1.2 Apparent viscosity at low shear rates is covered in TestMethod D 1824.1.3 The values stated in SI units are to be regarded as thestandard. The values in parentheses are given for informationonly.1.4 This standard does not purport to addre
4、ss the safetyconcerns associated with its use. It is the responsibility of theuser of this standard to establish appropriate safety and healthpractices and determine the applicability of regulatory limita-tions prior to use.NOTE 1This standard resembles ISO 4575-1985 in title only. Thecontent is sig
5、nificantly different.2. Referenced Documents2.1 ASTM Standards:D 1475 Test Method for Density of Liquid Coatings, Inks,and Related Products2D 1755 Specification for Poly(Vinyl Chloride) Resins3D 1824 Test Method for Apparent Viscosity of Plastisolsand Organosols at Low Shear Rate by Brookfield Visco
6、m-eter3E1 Specification for ASTM Thermometers4E 691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method52.2 ISO Standard:ISO 4575-1985: Poly Vinyl Chloride PastesDetermination of Apparent Viscosity Using the SeversRheometer63. Summary of Test Method3.1 The sam
7、ple is conditioned to the proper temperature,placed into an extrusion chamber, and extruded under standardconditions. The viscosity is calculated from the extrusionpressure and the rate of flow through the orifice.4. Significance and Use4.1 The suitability of a dispersion resin for any givenapplicat
8、ion is dependent upon its viscosity characteristics.4.2 The extrusion viscosity defines the flow behavior of aplastisol or organosol under high shear. This viscosity relates tothe conditions encountered in mixing, pumping, knife coating,roller coating, and spraying processes.5. Apparatus5.1 Extrusio
9、n Rheometer.75.2 Orifice, 3.17 6 0.13 mm (0.125 6 0.005 in.) insidediameter and 50 6 1.0 mm (1.97 6 0.04 in.) long.5.3 Sample Containers, Tin Cans, or Glass Jars, 1-pt(500-mL) capacity.5.4 Paper Cups, 8-oz (250-mL) capacity.5.5 Nitrogen Cylinder, equipped with pressure regulator andgage.1This test m
10、ethod is under the jurisdiction ofASTM Committee D20 on Plasticsand is the direct responsibility of Subcommittee D20.15 on Thermoplastic Materials(Section D20.15.08).Current edition approved March 15, 1995. Published May 1995. Originallypublished as D 1823 61 T. Last previous edition D 1823 90.This
11、edition contains changes in Sections 1, 2, and 11 to include an ISOequivalency statement, to reference a low shear rate test method, and to includeKeywords.2Annual Book of ASTM Standards, Vol 06.01.3Annual Book of ASTM Standards, Vol 08.01.4Annual Book of ASTM Standards, Vol 14.03.5Annual Book of AS
12、TM Standards, Vol 14.02.6Available from American National Standards Institute, 11 W. 42nd St., 13thFloor, New York, NY 10036.7Burrell Severs, ModelA-120, or equivalent, has been found satisfactory for thispurpose. The Burrell Severs, ModelA-120 is available from the Burrell Corp., 2223Fifth Ave., Pi
13、ttsburgh, PA 15219.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.6 ThermometerASTM Solvents Distillation Thermom-eter having a range from 2 to +52C (28 to 126F) andconforming to the requirements for Thermometer 37C asprescribed i
14、n Specification E1.5.7 Timer.6. Conditioning6.1 Maintain the plastisol or organosol samples at 236 1C(73 6 2 F) and 50 6 5 % relative humidity at all times aftermixing and throughout the period of viscosity determinations.7. Procedure7.1 Set Up RheometerAttach the pressure regulator to thenitrogen t
15、ank. Connect the nitrogen supply to the rheometer bymeans of the copper tubing. Do not use oxygen or liquidpressure sources (Note 2). Set the three-way quick-acting valveto the IN position. Regulate the tank pressure to give 1.04 MPa(150 psi) pressure to the instrument. Do not use input or linepress
16、ure over 1.38 MPa (200 psi). Insert the medium-sizeorifice (approximately 3.2 mm (18 in.) inside diameter) in theorifice retaining cap, with the orifice and the barrel, then screwthe cap solidly in place. Mount the barrel in the instrument.NOTE 2Air may be used instead of nitrogen.7.2 Weigh four emp
17、ty paper cups for each sample to betested. Record tare weight of each cup to the nearest 0.1 g. Fillthe barrel with the sample to be tested to within 13 mm (12 in.)from the top of the barrel. Measure the sample temperature.7.3 Insert the top air cap and gasket into the air cap ring,screw it in place
18、 on top of the barrel, and connect the air supplyquick-connector.7.4 Set the rheometer regulator gage (on the right side of theinstrument) to 0.069 MPa (10 psi) pressure. Open the three-way quick-acting valve to the OUT position and allow the mixto extrude into an unweighed paper cup for 10 s. Adjus
19、t thegage pressure back to 0.069 MPa (10 psi).7.5 Quickly place a preweighed and labeled paper cup underthe nozzle and at the same time start the timer. Collect theextrudate until approximately 50 g of sample have entered thecup. Simultaneously remove the cup and stop the timer, againplacing the unw
20、eighed cup under the nozzle. (Use a maximumflow time of 200 s for extremely viscous samples.)7.6 Push the three-way quick-acting valve to the IN positionto turn off the nitrogen supply to the chamber.7.7 Record the extrudate weight to the nearest 0.1 g andefflux time to the nearest second.7.8 Increa
21、se the gage pressure to 0.28 MPa (40 psi) andrepeat 7.4-7.7. Increase to 0.48 MPa (70 psi) and repeat. Makea fourth determination at 0.69 MPa (100 psi). Report the exactefflux time for each determination. After all tests have beencompleted, turn off the nitrogen supply at the tank. Release thepressu
22、re in the instrument by pulling the three-way valve to theOUT position.7.9 Clean the orifice between runs using pipe cleaners thathave been wetted with mineral spirits. Take care that the innersurface of the orifice does not become scratched. After rinsingwith mineral spirits, dry the orifice in air
23、.8. Calculation8.1 Calculate the shear stress, shear rate, and viscosity asfollows:Shear stress, MPa or psi!5PR/2L (1)where:P = pressure in rheometer, MPa (or psi),R = radius of orifice, cm (in.), andL = length of orifice, cm (in.)Shear rate, s215 4W/3.1416R3DT (2)where:W = weight of material efflux
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD182319952001STANDARDTESTMETHODFORAPPARENTVISCOSITYOFPLASTISOLSANDORGANOSOLSATHIGHSHEARRATESBYEXTRUSIONVISCOMETER

链接地址:http://www.mydoc123.com/p-510517.html