BS 4272-2-1996 Anaesthetic and analgesic machines - Specification for intermittent (demand) flow analgesic machines for use with 50 50% (V V) nitrous oxide and oxygen《麻醉机和止痛机 使用50 .pdf
《BS 4272-2-1996 Anaesthetic and analgesic machines - Specification for intermittent (demand) flow analgesic machines for use with 50 50% (V V) nitrous oxide and oxygen《麻醉机和止痛机 使用50 .pdf》由会员分享,可在线阅读,更多相关《BS 4272-2-1996 Anaesthetic and analgesic machines - Specification for intermittent (demand) flow analgesic machines for use with 50 50% (V V) nitrous oxide and oxygen《麻醉机和止痛机 使用50 .pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、BRITISH STANDARD BS 4272-2:1996 Incorporating Amendment No. 1 Anaesthetic and analgesic machines Part 2: Specification for intermittent (demand) flow analgesic machines for use with 50/50 % (V/V) nitrous oxide and oxygen UDC 615.471:616-089.5 BS 4272-2:1996 This British Standard, having been prepare
2、d under the direction of the Health and Environment Sector Board, was published under the authority of the Standards Board and comes into effect on 15 February 1996 BSI 14 January 2003 First published January 1968 Second edition January 1996 The following BSI references relate to the work on this st
3、andard: Committee reference CH/121/1 Draft for comment 92/59246 DC ISBN 0 580 24776 7 Committees responsible for this British Standard The preparation of this British Standard was entrusted to Technical Committee CH/121/1, Anaesthetic machines, breathing systems and anaesthetic gas scavenging system
4、s, upon which the following bodies were represented: Association of Anaesthetists of Great Britain and Ireland Association of British Health-care Industries British Anaesthetic and Respiratory Equipment Manufacturers Association Department of Health Medical Sterile Products Association The following
5、 body was also, represented in the drafting of the standard, through subcommittees and panels: Guild of Hospital Pharmacists Amendments issued since publication Amd. No. Date Comments 13963 14 January 2003 See national forewordBS 4272-2:1996 BSI 14 January 2003 i Contents Page Committees responsible
6、 Inside front cover Foreword ii 1S c o p e 1 2 Normative references 1 3 Definitions 1 4M a t e r i a l s 2 5 Resistance to mechanical hazards 2 6 Legibility of marking on controls 2 7 Performance after cleaning and sterilization or disinfection 2 8 Connections for gas supply 2 9 Patient connection p
7、ort 3 10 Exhaust port 3 11 Resistance to environmental conditions 4 12 Performance 4 13 Additional requirements for portable analgesic machines 4 14 Marking 5 15 Instructions for use 5 Annex A (normative) Test methods 6 Figure 1 Symbol for “use no oil” 2 Figure A.1 Apparatus for measuring inspirator
8、y flow and differential pressure 8 Figure A.2 Apparatus for measuring expiratory resistance 9 List of references 11BS 4272-2:1996 ii BSI 14 January 2003 Foreword This Part of BS 4272 has been prepared by Technical Committee CH/121/1. It supersedes BS 4272-2:1968, which is withdrawn. The start and fi
9、nish of text introduced or altered by Amendment No. 1 is indicated in the text by tags . Since BS 4272-2:1968 was published, the use of analgesic machines of this type has been extended to emergency situations and to medical fields other than the original application in midwifery. The major differen
10、ces between this standard and BS 4272-2:1968 are as follows. a) Requirements related to the safety of high-pressure components have been added. b) Requirements related to mechanical construction have been added, with particular relevance to the use of the equipment in emergency situations. c) The re
11、quirements for the gas supply connection, pertaining to compatibility with anaesthetic and respiratory equipment specified in other British Standards, have been clarified. d) The test methods have been modified so that they are consistent with those in other British Standards for anaesthetic and res
12、piratory equipment. e) The requirements for marking and instructions have been aligned with the Essential Requirements of the EEC Directive for Medical Devices. f) The pressure/flow characteristics of the demand valve have been reconsidered. Whilst there have been no reports of hazards to the patien
13、t resulting from the previous peak flow of 300 l/min, evidence was presented to the committee responsible for this standard that rapid response to inspiratory effort may be as significant as the peak flow capability in satisfying the respiratory needs of the patient, and a lower peak flow may theref
14、ore be acceptable. Analgesic machines based on this premise have been in use without report of hazards. The other Parts of BS 4272 are as follows: BS 4272, Anaesthetic and analgesic machines Part 1:1968 Anaesthetic machines of the on-demand type supplied with nitrous oxide and oxygen from separate c
15、ontainers; Part 3:1989 Specification for continuous flow anaesthetic machines. Part 1 and Part 3 of BS 4272 were declared obsolescent following review in 1991 and 2001 respectively. This publication does not purport to include all the necessary provisions of a contract. Users are responsible for its
16、 correct application. Compliance with a British Standard does not of itself confer immunity from legal obligations. Summary of pages This document comprises a front cover, an inside front cover, pages i and ii, pages 1 to 11 and a back cover. The BSI copyright notice displayed in this document indic
17、ates when the document was last issued.BS 4272-2:1996 BSI 14 January 2003 1 1 Scope This Part of BS 4272 specifies general design and performance requirements for analgesic machines of the intermittent flow (demand flow) type for use with 50/50 % (V/V) mixture of oxygen and nitrous oxide contained i
18、n one cylinder, (hereinafter referred to as “analgesic machines”). This standard applies to analgesic machines intended for: a) connection directly to a high-pressure (approximately 17 MPa) cylinder; and b) connection to a low-pressure (280 kPa to 600 kPa) supply from a medical gas pipeline system,
19、or a similar source, filled from cylinder(s). Additional requirements for portable machines are included. This standard applies to analgesic machines which are intended to be used under supervision for the self-administration of an analgesic mixture in response to the patients inspiratory effort, e.
20、g. in obstetric practice or in emergency situations. It does not apply to analgesic machines with means to initiate gas flow other than by the patients inspiratory effort. This standard does not give requirements for the composition of the gas mixture, or for the markings on the gas cylinder. NOTE T
21、he gas is pre-mixed and specified in pharmacopoeia monographs. The markings on the gas cylinder are specified in BS EN 850 . 2 Normative references This Part of BS 4272 incorporates, by dated or undated reference, provisions from other publications. These normative references are made at the appropr
22、iate places in the text and the cited publications are listed on page 11. For dated references, only the edition cited applies; any subsequent amendments to or revisions of the cited publication apply to this Part of BS 4272 only when incorporated in the reference by amendment or revision. For undat
23、ed references, the latest edition of the cited publication applies, together with any amendments. 3 Definitions For the purposes of this Part of BS 4272, the definitions in BS EN ISO 4135 apply, together with the following. 3.1 intermittent flow analgesic machine; demand flow analgesic machine analg
24、esic machine which, in response to the patients inspiratory effort, delivers a flow of gas which ceases at the start of the expiratory phase 3.2 direct probe probe that has either the nut and nipple of a non-interchangeable screw-threaded (NIST) connector as integral parts, or is directly coupled to
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BS427221996ANAESTHETICANDANALGESICMACHINESSPECIFICATIONFORINTERMITTENTDEMANDFLOWANALGESICMACHINESFORUSEWITH5050VVNITROUSOXIDEANDOXYGEN

链接地址:http://www.mydoc123.com/p-544265.html