ASTM F1949-1999(2013) Standard Specification for Medical Oxygen Delivery Systems for EMS Ground Vehicles《EMS陆上车辆用医用氧输送系统的标准规范》.pdf
《ASTM F1949-1999(2013) Standard Specification for Medical Oxygen Delivery Systems for EMS Ground Vehicles《EMS陆上车辆用医用氧输送系统的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1949-1999(2013) Standard Specification for Medical Oxygen Delivery Systems for EMS Ground Vehicles《EMS陆上车辆用医用氧输送系统的标准规范》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F1949 99 (Reapproved 2013)Standard Specification forMedical Oxygen Delivery Systems for EMS GroundVehicles1This standard is issued under the fixed designation F1949; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the
2、year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONWithin the United States, there are several widely recognized national and international standardsorganization
3、s that have established standards and guidelines for oxygen delivery systems. Thesestandards and guidelines were largely developed for intra-facility use. This standard, developed byASTM Subcommittee F30.01, addresses the requirements for oxygen systems, both liquid andgaseous, for emergency medical
4、 services (EMS) ground vehicles.1. Scope1.1 This standard covers minimum requirements for pri-mary medical oxygen delivery systems for EMS groundvehicles used in the following applications:1.1.1 The transportation of the sick and injured to or from anappropriate medical facility while basic, advance
5、d, or special-ized life support services are being provided,1.1.2 The delivery of interhospital critical transport care,1.1.3 The delivery of nonemergency, medically required,transport services, and1.1.4 The transportation and delivery of personnel andsupplies essential for proper care of an emergen
6、t patient.1.2 This standard establishes criteria to be considered in theperformance, specification, purchase, and acceptance testing ofground vehicles for EMS use.1.3 This entire standard should be read before ordering anambulance in order to be knowledgeable of the types ofequipment that are availa
7、ble and their performance require-ments. Due to the variety of ambulance equipment or features,some options may be incompatible with all chassis manufac-turers models. Detailed technical information is availablefrom the chassis manufacturers.1.4 The sections in this standard appear in the followings
8、equence:SectionScope 1Referenced Documents 2Terminology 3Significance and Use 4Design Requirements 5Performance Requirements 6Installation Requirements 7GOX System Design Requirements 8GOX System Installation Requirements 9GOX System Test Requirements 10LOX System Design Requirements 11LOX System In
9、stallation Requirements 12LOX System Test Requirements 13Keywords 142. Referenced Documents2.1 The following documents, of the issue currently ineffect, form a part of this standard to the extent specifiedherein.2.2 ASTM Standards:2F1177 Terminology Relating to Emergency Medical Ser-vices2.3 Militar
10、y Standards:3MIL-STD-461 Requirements for the Control of Electromag-netic Interference Emissions and SusceptibilityMS 33584 Standard Dimensions for Flared Tubing EndMS 33611 End Bend Radii2.4 Federal Specifications:3RR-C-901 Cylinders, Compressed Gas: High Pressure, SteelDOT3AA, and Aluminum Applica
11、tions, General Specifi-cation for1This specification is under the jurisdiction of ASTM Committee F30 onEmergency Medical Services and is the direct responsibility of SubcommitteeF30.01 on EMS Equipment.Current edition approved June 1, 2013. Published August 2013. Originallyapproved in 1999. Last pre
12、vious edition approved in 2007 as F1949 99 (2007).DOI: 10.1520/F1949-99R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe
13、 ASTM website.3Available from Standardization Document Order Desk, 700 Robbins Ave.,Building #4, Section D, Philadelphia, PA 19111-5094.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States12.5 ASME Standard:4B31.3 Chemical Plant and Petrol
14、eum Refinery Piping2.6 CGA Standards:5E-7 Standard for Medical Gas Regulators and FlowmetersV-1 Compressed Gas Cylinder Valve Outlet and Inlet Con-nectionsS-1.1 Pressure Relief Device Standards Part 1 - Cylinders forCompressed Gases3. Terminology3.1 DefinitionsSpecific terms used throughout this spe
15、ci-fication are defined in 3.2. Other applicable terms are containedin Terminology F1177.3.2 Definitions of Terms Specific to This Standard:3.2.1 design operating pressure, nthe nominal pressure atwhich the oxygen equipment or container is designed tooperate during normal use.3.2.2 LOX container, na
16、 vessel used to store or transportliquid oxygen.3.2.3 maximum allowable working pressure, n the maxi-mum gage pressure to which the equipment or container can besubjected without exceeding the allowable design stress.3.2.4 maximum filling volume, nthe maximum filling vol-ume of liquid at its maximum
17、 permissible level.3.2.5 pressure relief device, na device designed to open inorder to prevent a rise of internal fluid pressure in excess of aspecified value.3.3 Symbol:g = the normal or standard constant of gravity at sea level;approximately 32.2 ft/s/s (9.81 m/s/s)3.4 Acronyms:3.4.1 GOX, ngaseous
18、 oxygen.3.4.2 LOX, nliquid oxygen.4. Significance and Use4.1 The intent of this standard is to establish minimumrequirements, test parameters, and other criteria essential foroxygen system design, performance, and appearance, and toprovide for a practical degree of standardization. The object isto p
19、rovide oxygen systems that are properly constructed, andwhich, when properly serviced and maintained, will reliablyfunction on an EMS ground vehicle.5. Design Requirements5.1 The medical oxygen delivery system may be either agaseous oxygen (GOX) system, or a liquid oxygen (LOX)system.5.2 The oxygen
20、delivery system shall be a piped oxygensystem designed and installed as follows:5.2.1 CapacityThe oxygen system shall be capable ofstoring and supplying a minimum of 3000 L of gaseousmedical oxygen.5.2.2 ComponentsAll oxygen delivery system compo-nents shall be approved by the manufacturer of the co
21、mponentfor the intended service. The system shall include the follow-ing:5.2.2.1 Oxygen Piping System, designed and sized to deliverthe required flow rates at the utilization pressures. Piping andtubing shall be of non-ferrous or corrosion-resistant steelmaterial and shall comply with the design req
22、uirements ofASME B31.3. Hose shall be electrically conductive andapproved by the manufacturer for oxygen service at thepressure and temperatures the hose will be subjected to inservice. Fittings shall be of non-ferrous or corrosion-resistantsteel material and shall comply with the design requirement
23、s ofASME B31.3. Cast fittings shall not be used.5.2.2.2 Flow Control Device, of a pressure-compensatedtype that includes a means to display and monitor deliveredflow rate. It shall be continuously adjustable over a minimumrange of 0 to 15 L/min, with a calibrated display resolution ofat least 0.5 L/
24、min. The flow control device shall be calibratedfor 50 psig inlet pressure and be able to withstand a minimuminlet pressure of 200 psig without damage or failure. It shallincorporate an inlet filter and be electrically conductive frominlet to outlet. Flow control device accuracy shall be within610 %
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF194919992013STANDARDSPECIFICATIONFORMEDICALOXYGENDELIVERYSYSTEMSFOREMSGROUNDVEHICLESEMS 陆上 车辆 医用

链接地址:http://www.mydoc123.com/p-536116.html