ASTM F1831-1997(2006) Standard Specification for Cranial Traction Tongs and Halo External Spinal Immobilization Devices《颅骨牵引钳和晕轮外部脊椎固定装置的标准规范》.pdf
《ASTM F1831-1997(2006) Standard Specification for Cranial Traction Tongs and Halo External Spinal Immobilization Devices《颅骨牵引钳和晕轮外部脊椎固定装置的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1831-1997(2006) Standard Specification for Cranial Traction Tongs and Halo External Spinal Immobilization Devices《颅骨牵引钳和晕轮外部脊椎固定装置的标准规范》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 1831 97 (Reapproved 2006)Standard Specification forCranial Traction Tongs and Halo External SpinalImmobilization Devices1This standard is issued under the fixed designation F 1831; the number immediately following the designation indicates the year oforiginal adoption or, in the case
2、of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers standards a manufacturer shallmeet in the designing, manufacturing
3、, testing, labeling, anddocumenting of halo and tong external spinal immobilizationdevices, but it is not to be construed as production methods,quality control techniques, manufacturers lot criteria, or clini-cal recommendations.1.2 This specification represents the best currently availabletest proc
4、edures at this time and is a minimum safety andperformance standard.1.3 This specification covers only those halo and tongdevices intended for use on humans for therapeutic purposes.This specification assumes the user is well-trained in theprocedures and maintenance of halo and tong application andh
5、as the ability to determine if an abnormality is treatable bythese procedures.1.4 This specification describes those devices commonlyknown as halo external fixation devices and what is known ascranial traction tongs.1.5 Cranial traction tongs and halo devices are used toachieve and maintain optimal
6、spinal alignment, in order toenhance fusion and decrease neurological deficit.1.6 Monitoring the progress of treatment after application ofthese devices is important, this should be done in accordancewith the manufacturers recommendation and guidelines per-taining to the specific device.1.7 The valu
7、es stated in both inch-pound and SI units are tobe regarded separately as the standard. The values given inparentheses are for information only.1.8 The following precautionary statement pertains only tothe test method portions Sections 10-15 of this specification:This standard does not purport to ad
8、dress all of the safetyconcerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety andhealth practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 IEC Standard:IEC 601-1 Medical E
9、lectrical Equipment23. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 cranial traction tonga device providing weightedcervical traction to a patient through invasive attachment to theskull. This traction instrument is indicated for closed reductionof a cervical spine injury (tha
10、t is, fracture or dislocation).3.1.1.1 adjustable tonga cranial traction tong that adjustsfor size, pin positioning, or pin pressure.3.1.1.2 one-piece tonga rigid, single-piece, semicircularcranial traction tong designed to accommodate a minimum oftwo skull pins for mounting the device to the patien
11、ts headbelow the equator.3.1.2 halo devicean external fixator for cervical stabiliza-tion that fastens by invasive means to a patients skull, andmaintains the position of the skull in relation to the thoracicarea of the patient.3.1.3 halo ringthe portion of the halo device that fastensby invasive me
12、ans to a patients skull below the head equator.3.1.3.1 closed loop halo ringa halo ring incorporating aclosed loop anywhere in the design for purposes of structuralintegrity when the ring is in use. This type of ring has multiplepositioning options for the selection of pin sites and is mountedto the
13、 head with multiple skull pins.3.1.3.2 head equatorthe greatest circumference of thehead in the coronal aspect3.1.3.3 open loop halo ringa halo ring with a posterioropening, such that the part does not incorporate a closed loopanywhere in the design for structural integrity. This ring hasmultiple po
14、sition options for the selection of pin sites and ismounted to the head with multiple skull pins.3.1.4 halo superstructurea rigid external framework usedto maintain positioning of the skull and cervical spine inrelation to the thoracic and lumbar spine. Connects the haloring to halo vest.1This speci
15、fication is under the jurisdiction of ASTM Committee F04 onMedical and Surgical Material and Devices and is the direct responsibility ofSubcommittee F04.31 on Neurosurgical Standards.Current edition approved March 1, 2006. Published April 2006. Originallyapproved in 1997. Last previous edition appro
16、ved in 1997 as F 1831 97.2Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.4.1 halo superstructure adjustment mechanisms com
17、-ponents that allow adjustment of angles and distances betweenring and uprights or vest and uprights.3.1.4.2 transverse bara rigid horizontal component of thehalo superstructure.3.1.4.3 upright bara rigid vertical component of the halosuperstructure.3.1.4.4 vest attachment mechanismattaches inferior
18、ly tothe halo superstructure and connects to vest shell, maintainspositioning of the halo superstructure in relation to the vestshell.3.1.4.5 vest platepart of the superstructure attached to thevest shell to provide a stable mounting point for the vestattachment mechanisms.3.1.5 halo vesta body-orth
19、osis that serves as a mountingpoint for the halo and superstructure.3.1.5.1 C.P.R. accessmechanism in vest or superstructureto allow quick access to patients chest for cardiopulmonaryresuscitation (C.P.R.).3.1.5.2 vest linerpadding worn inside of halo vest shelland against the skin which distributes
20、 the pressure of the vestshell against the skin.3.1.5.3 vest shellrigid portion of body orthosis.3.1.6 skull pina rigid device used to invasively anchor thehalo ring or cranial traction tongs to the skull at selectedmounting points.3.1.6.1 adjustable skull pin(1) a pin that is force con-trolled by a
21、 mechanical mechanism, that is, spring-loadedpressure pin. (2) a solid threaded pin that maintains pressureand fixation against the skull through application of a cali-brated torque.3.1.6.2 fixed skull pina pin that is mounted directly to atong structure requiring a drilled skull hole for positionin
22、g andfixation. Pressure is not adjusted directly through the pin.3.1.7 traction bail (traction hoop)a device that may beattached to the halo ring to facilitate the application ofweighted longitudinal traction.4. Conformance4.1 Presently, this specification is voluntary and not by law.A manufacturer
23、may label a product as conforming to thisspecification only if the product indeed meets all the require-ments of this specification.5. Classification5.1 Halo External FixatorTypically a complete systemconsisting of the halo ring, skull pins, vest and superstructure.The uniqueness of this system is i
24、ts ability to provide self-contained cervical stabilization.5.2 Cranial Traction TongsEither a rigid single-piece,semicircular device or an adjustable device. Both designs haveaccommodations for at least two skull pins to be mounted tothe skull. Typically designed to be fitted over the top of thehea
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF183119972006STANDARDSPECIFICATIONFORCRANIALTRACTIONTONGSANDHALOEXTERNALSPINALIMMOBILIZATIONDEVICES

链接地址:http://www.mydoc123.com/p-535786.html