ASTM F1611-2000(2009) Standard Specification for Intramedullary Reamers《髓内铰具的标准规范》.pdf
《ASTM F1611-2000(2009) Standard Specification for Intramedullary Reamers《髓内铰具的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1611-2000(2009) Standard Specification for Intramedullary Reamers《髓内铰具的标准规范》.pdf(2页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 1611 00 (Reapproved 2009)Standard Specification forIntramedullary Reamers1This standard is issued under the fixed designation F 1611; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A numbe
2、r in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification provides requirements for material,dimensions and tolerances, finish and marking, and care andhandling for reamers intend
3、ed to cut a cylindrical path alongthe medullary canal of diaphyseal bone.1.2 Intramedullary reamers are commonly used to preparethe medullary canal for the insertion of intramedullary fixationdevices (IMFD). As such, the relationship between the in-tramedullary reamer diameter and the IMFDs diameter
4、 areconsidered.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.2. Referenced Documents2.1 ASTM Standards:2A 564/A 564M Specification for Hot-Rolled and Cold-Finished Age-Hardening Stainless Steel Bars and ShapesA 693 Specifi
5、cation for Precipitation-Hardening Stainlessand Heat-Resisting Steel Plate, Sheet, and StripA 705/A 705M Specification for Age-Hardening StainlessSteel ForgingsF86 Practice for Surface Preparation and Marking of Me-tallic Surgical ImplantsF 565 Practice for Care and Handling of Orthopedic Im-plants
6、and InstrumentsF 899 Specification for Wrought Stainless Steels for Surgi-cal InstrumentsF 983 Practice for Permanent Marking of Orthopaedic Im-plant ComponentsF 1264 Specification and Test Methods for IntramedullaryFixation Devices3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1
7、.1 cutting head, nthe portion of the reamer, whichconsists of flutes, or edges, which cut the bone.3.1.2 reamer diameter, nthe diameter of the circum-scribed circle of the cutting heads cross-section (shown in Fig.1).3.1.3 reamer shaft diameter, nthe diameter of the circum-scribed circle of the long
8、 portion of the reamer, which connectsthe cutting portion of the reamer to the drill.4. Classification4.1 In general, intramedullary reamers consist of two types:4.1.1 One-piece reamerA design where the reamer shaftand cutting head are permanently attached to each other.4.1.2 Modular ReamerA design
9、where the reamer shaftand cutting head are two separate components, fixed to eachother temporarily at the time of use via a geometric connection,for example, dovetail joint.5. Dimensions and Tolerances5.1 The reamer diameter shall be measured at the largestportion of the cutting heads cross section
10、and reported to thenearest 0.2 mm. The reamer diameter shall be measured usinga micrometer or an appropriate ring gage. When using amicrometer to measure reamers with an odd number of flutes,a V-anvil micrometer (with the appropriate angle, based on thenumber of flutes) will be used to accurately de
11、termine thereamer diameter.5.2 The tolerance of a reamer diameter shall be no morethan 60.075 mm.6. Material Requirements6.1 The reamers shaft and cutting head shall be fabricatedfrom materials with suitable strength, hardness, and corrosionresistance. The materials described in Specifications A 564
12、/A 564M A 693, A 705/A 705M and F 899 have been found tobe suitable for this use.7. Finish and Marking7.1 The shaft and cutting head shall be free from burrs,nicks, dents, and scratches when examined in accordance withPractice F86.7.2 The flutes of the cutting head will be of the appropriategeometry
13、 to perform the intended use of reaming bone.1This specification is under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibliltiy ofSubcommittee F04.21 on Osteosynthesis.Current edition approved April 1, 2009. Published April 2009. Origin
14、allyapproved in 1995. Last previous edition approved in 2004 as F 1611 00(2004).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF161120002009STANDARDSPECIFICATIONFORINTRAMEDULLARYREAMERS 髓内铰具 标准规范 PDF

链接地址:http://www.mydoc123.com/p-535262.html