《ASTM E2075 E2075M-2015 Standard Practice for Verifying the Consistency of AE-Sensor Response Using an Acrylic Rod《采用丙烯酸棒测定AE传感器响应一致性的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2075 E2075M-2015 Standard Practice for Verifying the Consistency of AE-Sensor Response Using an Acrylic Rod《采用丙烯酸棒测定AE传感器响应一致性的标准实施规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2075/E2075M 15Standard Practice forVerifying the Consistency of AE-Sensor Response Using anAcrylic Rod1This standard is issued under the fixed designation E2075/E2075M; the number immediately following the designation indicates the yearof original adoption or, in the case of revision,
2、the year of last revision. A number in parentheses indicates the year of last reapproval.A superscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This practice is used for routinely checking the sensi-tivity of acoustic emission (AE) sensors. It is in
3、tended toprovide a reliable, precisely specified way of comparing a setof sensors, or telling whether an individual sensors sensitivityhas degraded during its service life, or both.1.2 This practice is not a “calibration” nor does it givefrequency response information.1.3 UnitsThe values stated in S
4、I units or inch-pound unitsare to be regarded separately as standard. The values stated ineach system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-conformancewith the standard.1.4 This standard do
5、es not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Sta
6、ndards:2E650 Guide for Mounting Piezoelectric Acoustic EmissionSensorsE750 Practice for Characterizing Acoustic Emission Instru-mentationE976 Guide for Determining the Reproducibility ofAcousticEmission Sensor ResponseE2374 Guide for Acoustic Emission System PerformanceVerification3. Significance an
7、d Use3.1 Degradation in sensor performance can occur due todropping, mechanical shock while mounted on the teststructure, temperature cycles, and so forth. It is necessary anddesirable to have a simple measurement procedure that willcheck the consistency of sensor response, while holding allother va
8、riables constant.3.2 While test blocks of many different kinds have beenused for this purpose for many years, an acrylic polymer rodoffers the best all-around combination of suitable acousticproperties, practical convenience, ease of procurement and lowcost.3.3 Because the acoustic properties of the
9、 acrylic rod areknown to depend on temperature, this practice requires that therod, sensors, and couplant be stabilized at the same workingtemperature, prior to verifying the sensors.3.4 Attention should be paid to storage conditions for theacrylic polymer rod. For example, it should not be left in
10、afreezing or hot environment overnight, unless it is given timefor temperature stabilization before use.3.5 Properly applied and with proper record keeping, thispractice can be used in many ways. The user organization mustdetermine the context for its use, the acceptance standards andthe actions to
11、be taken based on the lead break results. Thefollowing uses are suggested:3.5.1 To determine when a sensor is no longer suitable foruse.3.5.2 To check sensors that have been exposed to high-riskconditions, such as dropping, overheating, and so forth.3.5.3 To get an early warning of sensor degradatio
12、n overtime. This can lead to identifying conditions of use, which aredamaging sensors, and thus, to better equipment care and lowerreplacement costs.3.5.4 To obtain matched sets of sensors, preamplifiers,instrumentation channels, or a combination thereof, for moreuniform performance of the total sys
13、tem.3.5.5 To save time and money, by eliminating the installa-tion of bad sensors.3.5.6 To verify sensors quickly but consistently in the fieldand to assist trouble-shooting when a channel does not pass aperformance check.1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-struct
14、ive Testing and is the direct responsibility of Subcommittee E07.04 onAcoustic Emission Method.Current edition approved June 1, 2015. Published June 2015. Originallyapproved in 2000. Last previous edition approved in 2010 as E2075 - 10. DOI:10.1520/E2075_E2075M-15.2For referenced ASTM standards, vis
15、it the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 1
16、00 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.6 All the above uses are recommended for consideration.The purpose of this practice is not to call out how these uses areto be implemented, but only to state how the test itself is to beperformed so that the results
17、 obtained will be accurate andreliable.4. Apparatus4.1 Acrylic Polymer Cylindrical Rod (Fig. 1), should beused. The actual material of the acrylic polymer rod ispolymethylmethacrylate.34.1.1 The rod shall be cast, not extruded.4.1.2 Dimensions of the rod should be 78.7 cm 31 in. longby 3.8 cm 1.5 in
18、. in diameter, sensors end cut true and smoothwith a surface finish of 0.4 m RMS 0.16 in.4.1.3 Other lengths of rod are acceptable, provided thatthere is sufficient distance to attenuate and prevent reflectedsignals from the nonsensor end of the rod reaching the sensor.4.1.4 A permanent reference ma
19、rk, for example an “X”, isplaced on the rod at a distance of 10.2 cm 4 in., or 30.5 cm12 in., or both, from one end; marking the spots where thepencil lead is to be broken. It is convenient to provide a verysmall spotface, for example, 0.8 mm 0.03 in. diameter and0.1 mm 0.004 in. deep at these refer
20、ence mark points, to restthe tip of the pencil lead to avoid slippage during the lead breakprocess.NOTE 1The surface finish of the cylindrical rod section could producereflections that affect AE response. The surface finish should be main-tained in a clean, undamaged condition.4.2 Hsu-Nielsen Pencil
21、 Lead Break Source, with 2H pencillead, as described in 4.3.3 of Guide E976.4.3 Sensor(s), to be tested.4.4 Acoustic Emission Equipment, with amplitude measure-ment capability, for recording sensor response. (Operatingfamiliarity with the apparatus is assumed.)4.5 Couplant, to be standardized and do
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME2075E2075M2015STANDARDPRACTICEFORVERIFYINGTHECONSISTENCYOFAESENSORRESPONSEUSINGANACRYLICROD 采用 丙烯酸

链接地址:http://www.mydoc123.com/p-530217.html