ASTM E1852-2013 Standard Test Method for Determination of Low Levels of Antimony in Carbon and Low-Alloy Steel by Graphite Furnace Atomic Absorption Spectrometry《采用石墨炉原子吸收光谱法测定碳和低合.pdf
《ASTM E1852-2013 Standard Test Method for Determination of Low Levels of Antimony in Carbon and Low-Alloy Steel by Graphite Furnace Atomic Absorption Spectrometry《采用石墨炉原子吸收光谱法测定碳和低合.pdf》由会员分享,可在线阅读,更多相关《ASTM E1852-2013 Standard Test Method for Determination of Low Levels of Antimony in Carbon and Low-Alloy Steel by Graphite Furnace Atomic Absorption Spectrometry《采用石墨炉原子吸收光谱法测定碳和低合.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1852 08E1852 13Standard Test Method forDetermination of Low Levels of Antimony in Carbon andLow-Alloy Steel by Graphite Furnace Atomic AbsorptionSpectrometry1This standard is issued under the fixed designation E1852; the number immediately following the designation indicates the year o
2、foriginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of antimony in c
3、arbon and low-alloy steel in the 0.0005 % 0.0005 % through0.010 % range.1.2 If this test method is used to test materials having contents less than 0.001 % antimony, users of different laboratories willexperience more than the usual 5 % risk that their results will differ by more than 50 % relative
4、error.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenc
5、ed Documents2.1 ASTM Standards:2E50 Practices forApparatus, Reagents, and Safety Considerations for ChemicalAnalysis of Metals, Ores, and Related MaterialsE135 Terminology Relating to Analytical Chemistry for Metals, Ores, and Related MaterialsE1184 Practice for Determination of Elements by Graphite
6、 Furnace Atomic Absorption SpectrometryE1601 Practice for Conducting an Interlaboratory Study to Evaluate the Performance of an Analytical MethodE1770 Practice for Optimization of Electrothermal Atomic Absorption Spectrometric EquipmentE1806 Practice for Sampling Steel and Iron for Determination of
7、Chemical Composition2.2 ISO Standards:3ISO 5725 Precision of Test MethodsDetermination of Repeatability and Reproducibility for a Standard Test Method byInter-Laboratory TestsISO 10698 SteelDetermination of Antimony ContentElectrothermal Atomic Absorption Spectrometric Method3. Terminology3.1 For de
8、finitions of terms used in this test method, refer to Terminology E135.4. Summary of Test Method4.1 The sample is dissolved in hydrochloric and nitric acids and diluted to volume. An appropriate aliquot is injected into thegraphite furnace atomizer of an atomic absorption spectrometer, which is equi
9、pped with a background correction. The sample isdried, pyrolized, and atomized. The absorbance of the radiation from the external light source is measured and compared to theabsorbance of samples of known position.NOTE 1In general, the deuterium correction system should be able to correct for the br
10、oad-band background absorbance up to 0.5 to 0.6 absorbanceunits. Zeeman systems should compensate for background levels as high as 1.0 to 1.5 absorbance units.1 This test method is under the jurisdiction of ASTM Committee E01 on Analytical Chemistry for Metals, Ores, and Related Materials and is the
11、 direct responsibility ofSubcommittee E01.01 on Iron, Steel, and Ferroalloys.Current edition approved March 1, 2008Nov. 15, 2013. Published March 2008January 2014. Originally approved in 1996. Last previous edition approved in 20012008as E1852 96 (2001).E1852 08. DOI: 10.1520/E1852-08.10.1520/E1852-
12、13.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 Available from American National Standards Institute (AN
13、SI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http:/www.ansi.org.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depi
14、ct all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 194
15、28-2959. United States15. Significance and Use5.1 This test method is utilized for the determination of trace levels of antimony in carbon and low-alloy steel. It is assumedthat the procedure will be performed by trained analysts capable of performing common laboratory practices skillfully and safel
16、y.It is expected that the work will be performed in a properly equipped laboratory and proper waste disposal procedures will befollowed.6. Hazards6.1 For precautions to be observed in the use of certain reagents and equipment in this test method, refer to Practices E50.7. Apparatus7.1 Atomic Absorpt
17、ion Spectrometer with Graphite Furnace Atomizer, equipped with background corrector and appropriatesignal output device, such as video display screen, digital computer, printer or strip chart recorder, and autosampler. It isrecommended that the instrument meet the following performance requirements
18、after adjusting the instrument and optimizing thefurnace heating program as described in Practice E1770.7.1.1 The characteristic mass determined in accordance with Practice E1770 for antimony shall be less than 25 pg or within themanufacturers tolerance.7.1.2 The precision of the most concentrated b
19、lank addition solution shall not exceed 10 % of the mean absorbance of the samesolution. The precision of the least concentrated blank addition solution (excluding Solution B0) shall not exceed 4 % of the meanabsorbance of the most concentrated blank addition solution when determined in accordance w
20、ith Practice E1770.7.1.3 The limit of detection of antimony as described in Practice E1770 shall be less than 20 pg.7.1.4 Unless the instrument is provided with automatic curve correction circuitry, the graph linearity shall not be less than 0.95when determined in accordance with Practice E1770.7.2
21、Graphite Tubes, with pyrolytic coating and grooves for graphite platform, suitable for use with the graphite furnace unit.7.3 Graphite Platform, pyrolytic graphite, Lvov design, to fit graphite tubes specified in 6.27.2.7.4 LabwareTo prevent contamination of the sample(s), all beakers, lids, volumet
22、ric flasks, and funnels must be cleaned withhot HNO3 (1 + 1) before use.8. Reagents8.1 Purity and Concentration of ReagentsThe purity and concentration of common chemical reagents shall conform toPractices E50. It is important that antimony shall not exceed 0.01 g/mL in each of the reagents and 0.00
23、1 g/mL in the water.8.2 Antimony Stock Solution (1 mL = 1 mg Sb)Dissolve 0.1000 g 6 0.0001 g high-purity antimony (minimum 99.9 % Sb) in30 mLHCl + 5 mLHNO3 in a 100-mLbeaker. Boil gently to expel oxides of nitrogen. Cool and transfer the solution into a 100-mLvolumetric flask. Dilute to mark with HN
24、O3 (1 + 1) and mix. Store in polypropylene or high density polyethylene bottle.8.3 Antimony Standard Solution (1 mL = 10 g Sb)Transfer 1.0 mL of the antimony stock solution to a 100-mL volumetricflask, dilute to the mark with HNO3 (1 + 1), and mix. Prepare this solution immediately before use.9. Sam
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME18522013STANDARDTESTMETHODFORDETERMINATIONOFLOWLEVELSOFANTIMONYINCARBONANDLOWALLOYSTEELBYGRAPHITEFURNACEATOMICABSORPTIONSPECTROMETRY

链接地址:http://www.mydoc123.com/p-529644.html