《ASTM E128-1999(2005) Standard Test Method for Maximum Pore Diameter and Permeability of Rigid Porous Filters for Laboratory Use《实验室用刚性多孔过滤器的最大孔隙直经和渗透性的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E128-1999(2005) Standard Test Method for Maximum Pore Diameter and Permeability of Rigid Porous Filters for Laboratory Use《实验室用刚性多孔过滤器的最大孔隙直经和渗透性的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 128 99 (Reapproved 2005)Standard Test Method forMaximum Pore Diameter and Permeability of Rigid PorousFilters for Laboratory Use1This standard is issued under the fixed designation E 128; the number immediately following the designation indicates the year oforiginal adoption or, in th
2、e case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of maximumpore diameter and permeability of
3、 rigid porous filters used inthe laboratory for filtration or diffusion. They are applicable tofilters made of sintered glass, ceramic, metal, or plastic. Thistest method establishes a uniform designation for maximumpore diameter and also provides a means of detecting andmeasuring changes which occu
4、r through continued use.1.2 This standard does not purport to address all of thesafety problems, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prio
5、r to use.2. Referenced Documents2.1 ASTM Standards:2D 845 Specifications for Five-Degree Xylene3. Terminology3.1 Definitions:3.1.1 maximum pore diameterthe diameter in microme-tres of a capillary of circular cross section that is equivalent(with respect to characteristics related to surface-tension
6、ef-fects) to the largest pore in the filter under consideration.NOTE 1It is recognized that the maximum pore diameter as definedherein does not necessarily indicate the physical dimensions of the largestpore in the filter, and furthermore, that the pores are highly irregular inshape. Because of this
7、 irregularity in shape and other phenomenacharacteristic of filtration, a filter may be expected to retain all particleslarger than the maximum pore diameter as defined and determined by thistest method, and will generally retain particles which are much smallerthan the determined diameter.3.1.2 per
8、meabilitythe flow of air, in millimetres perminute per square centimetre of filter area per 1 cm of waterpressure differential.4. Summary of Test Method4.1 Maximum Pore Diameter, is determined by immersingthe filter in a suitable test liquid and applying air pressure untilthe first bubble of air pas
9、ses through the filter. The maximumpore diameter is calculated from the surface tension of the testliquid and the applied pressure.4.2 Permeability is determined by measuring the flow ofair through the filter when subjected to a pressure differential.5. Apparatus5.1 Because of the variety of shapes
10、of apparatus in whichporous filters are incorporated, the apparatus for this testmethod is not specified in detail.Apparatus that has been foundsatisfactory is illustrated in Fig. 1 and Fig. 2.6. Procedure6.1 Maximum Pore DiameterThoroughly wet the cleanfilter to be tested by soaking it in the presc
11、ribed test liquid (seeTable 1). Connect the filter to a controllable source of clean,dry compressed air, and a manometer (Fig. 1). Immerse thefilter just below the surface of the test liquid (Note 2) andgradually increase the air pressure at a rate of about 5 mm/minin the area of the test until the
12、first dynamic bubble passesthrough the filter and rises through the liquid. The appearanceof the first true dynamic bubble is readily recognized since it isfollowed by a succession of additional bubbles. Read thepressure from the manometer. If the test is to be repeated,thoroughly re-wet the filter
13、by soaking it in the test liquid,before proceeding with the retest. Calculate the maximum porediameter from the following equation (see Appendix X1 forderivation):D 5 30g/p (1)where:D = maximum pore diameter as defined in 3.1.1,g = surface tension of test liquid in dynes/cm at thetemperature of the
14、test, andp = pressure, mm Hg (Note 3).The uniformity of distribution of pores approaching themaximum pore size may be observed by gradually increasingthe air pressure and noting the uniformity with which streams1This test method is under the jurisdiction of ASTM Committee E41 onLaboratory Apparatus
15、and is the direct responsibility of Subcommittee E41 E41.05on Filtration Materials.Current edition approved May 1, 2005. Published October 2005. Originallyapproved in 1957. Last previous edition approved in 1999 as E 128 99.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcont
16、act ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.of bubbles are distrib
17、uted over the surface of the filter. Cracks,fissures, and clogged areas are easily discerned by this opera-tion (Note 4).NOTE 2If there is significant head of liquid above the surface of thefilter, the back pressure so produced shall be deducted from the observedpressure, as follows:NOTE 1For tests
18、on filter-type crucibles, use a standard rubber crucible adapter. For test sticks and candles a rubber stopper with a rubber tubingattached is suitable. Clean dry compressed air shall be used.FIG. 1 Apparatus for Determining Maximum Pore Diameter of Rigid Porous FiltersNOTE 1For tests on filter-type
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME12819992005STANDARDTESTMETHODFORMAXIMUMPOREDIAMETERANDPERMEABILITYOFRIGIDPOROUSFILTERSFORLABORATORYUSE

链接地址:http://www.mydoc123.com/p-526920.html