ASTM D8028-2017 7500 Standard Test Method for Measurement of Dissolved Gases Methane Ethane Ethylene and Propane by Static Headspace Sampling and Flame Ionization Detection (GC FID.pdf
《ASTM D8028-2017 7500 Standard Test Method for Measurement of Dissolved Gases Methane Ethane Ethylene and Propane by Static Headspace Sampling and Flame Ionization Detection (GC FID.pdf》由会员分享,可在线阅读,更多相关《ASTM D8028-2017 7500 Standard Test Method for Measurement of Dissolved Gases Methane Ethane Ethylene and Propane by Static Headspace Sampling and Flame Ionization Detection (GC FID.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D8028 17Standard Test Method forMeasurement of Dissolved Gases Methane, Ethane,Ethylene, and Propane by Static Headspace Sampling andFlame Ionization Detection (GC/FID)1This standard is issued under the fixed designation D8028; the number immediately following the designation indicates
2、the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of the
3、 dis-solved gases methane, ethane, ethylene, and propane indrinking, waste, and ground waters, including domestic wellswhether naturally occurring or resulting from commercialactivities, or both. This test method has an analytical rangefrom 10 g/L to 11 000 g/L for methane, and 10 g/L to 5000g/L for
4、 ethane, ethylene, and propane. If a sample has adissolved gas concentration above the calibration range, thesample would require dilution. Other analytical ranges may beused as long as the instrument calibration encompasses thedesired range and demonstrates adequate linearity.1.2 This test method i
5、s applicable to samples collected at anominal ambient pressure. Samples collected under highpressures, that is, above ambient pressure, may not be accu-rately analyzed using this method.1.3 This test method is applicable to a variety of dissolvedgases; however, for the present standard, only methane
6、, ethane,ethylene, and propane were validated. In order to validate fora gas not listed in this method, the analyst would need toestablish the saturation concentration of the new gas, andestablish a valid curve, method detection limits, and demon-stration of capability following the procedures outli
7、ned in thismethod.1.4 It is the analysts responsibility to ensure the validity ofthis test method for dissolved gases other than methane, ethane,ethylene, and propane and other sample matrices.1.5 UnitsThe values stated in SI units are to be regardedas the standard. No other units of measurement are
8、 included inthis standard.1.6 A single laboratory study was done in order to establishprecision and bias for the surrogate (see Section 16).1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard t
9、o establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations prior to use.1.8 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Prin
10、ciples for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD8006 Guide for Sampl
11、ing and Analysis of Residential andCommercial Water Supply Wells in Areas of Explorationand Production (Ewhether naturally occurring or resulting from commercialactivities, dissolved gases can be hazardous both to theenvironment and humans. Due to increased scrutiny of dis-solved gases, developing a
12、 valid method for the testing ofdissolved gases in water has become more imperative. Staticheadspace sampling is generally applicable for sampling thesecompounds. However, as these compounds are extremelyvolatile, a closed sampling system is recommended. This testmethod can be used for quantitativel
13、y determining dissolvedgases in ground, waste, and drinking waters.5.2 This test method is suitable for analysis of domesticdrinking water samples from residential and commercial watersupply wells prior to oil and gas exploration activities to forma baseline for monitoring potential impacts during e
14、xploration;development, including hydraulic fracturing; operation; clo-sure; and reclamation of nearby oil and gas production andinjection wells.6. Interferences6.1 Impurities in the air, the field sample vials, vial septa,the preservatives, or improper sealing during transport cancause contaminatio
15、n. Field reagent blanks, recommended, andmethod blanks, required, shall be used to verify propersampling and handling procedures.6.2 Contamination or carryover can occur whenever ahighly concentrated sample is run. A blank shall be run after ahighly concentrated sample to ensure system cleanliness.6
16、.2.1 A carryover study should be done in order to deter-mine the nominal level of analytes that will be prone to carryover into subsequent samples. From this study, the laboratoryshall be required to run blanks after samples containinganalytes above the established levels. Otherwise, if a sample isr
17、un directly after a high level sample, reference the laboratorycarryover study, and displays an analyte of interest, the samplewould need to be run again in order to ensure the accuracy ofthe results.6.3 The FID is a nonselective detector. If matrix interferencecaused by sample contamination or comp
18、ound co-elutionoccurs, an experienced analyst is required for proper interpre-tation.6.4 In order to inhibit analyte loss, samples should be storedin 40-ml vials with polytetrafluoroethylene (PTFE)-lined chlo-robutyl septa with no headspace, and placed on ice within 15minutes of collection. Samples
19、should be transported and keptcooled to approximately 0 to 6C but not frozen. If any visibleheadspace exists in a sample vial, it is to be noted in the resultsreport as analyte loss can exist with sample headspace.Samples may be preserved with sulfuric acid in order to extendthe expiration date to t
20、wo weeks from the sampling date.7. Apparatus7.1 Sample containers, 40-ml screw top vials sealed withPTFE-lined chlorobutyl septa.7.2 Water bath, capable of maintaining water temperature at20C (60.5C) to prepare calibration and laboratory controlstandards. Temperatures other than 20C may be used, as
21、longas the saturated solution concentration at the temperature isknown.7.3 Digital thermometer.7.4 Gastight syringes, 10 l to 2.5 ml.7.5 Two gastight syringes fitted with removable needles andLuer lock valves, 25 ml.7.6 Glass-stoppered class A volumetric flasks, 5 to 100 ml.7.7 Stir bars, (required
22、only for sample agitation throughstirring) magnetic stir bars sized to fit in the 40-ml screw-topvials.7.8 Ultra high purity (UHP) inert gas, used to purge theinstrument headspace vial to displace any contaminants in thevial. Acceptable gases are helium and nitrogen.7.9 Sample Transfer DeviceA trans
23、fer device capable oftransferring sample or standard from the sample 40-ml vial toa separate 40-ml vial purged with an inert gas such as nitrogenor helium (recommended) and incorporated with a stir bar andsurrogate standard. (A stir bar is not needed if sample agitationis done by means other than st
24、irring such as vial shaking.) It isrecommended that sample transfer is done with a sealeddevice/system. See Fig. 1.NOTE 1The removable needle needs to be fitted with a Luer lockvalve to ensure the sample is not exposed to atmosphere during the sampletransfer. See Fig. 1.7.10 Sample AgitationSamples
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD802820177500STANDARDTESTMETHODFORMEASUREMENTOFDISSOLVEDGASESMETHANEETHANEETHYLENEANDPROPANEBYSTATICHEADSPACESAMPLINGANDFLAMEIONIZATIONDETECTIONGCFIDPDF

链接地址:http://www.mydoc123.com/p-526651.html