1、Designation: D5928 96 (Reapproved 2010)1D5928 18Standard Test Method Practice forScreening of Waste for Radioactivity1This standard is issued under the fixed designation D5928; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1 NOTEEditorial changes were made throughout in January 2010.1. Scope1.1 This test method covers the screening for , , and r
3、adiation above ambient background levels or user-defined criteria,or both, in liquid, sludge, or solid waste materials.1.2 This test method is intended to be a gross screening method for determining the presence or absence of radioactive materialsin liquid, sludge, or solid waste materials. It is no
4、t intended to replace more sophisticated quantitative analytical techniques, butto provide a method for rapidly screening samples for radioactivity above ambient background levels or user-defined criteria, orboth, for facilities prohibited from handling radioactive waste.1.3 This test method may not
5、 be suitable for applications such as site assessments and remediation activities.1.4 The values stated in SI units are to be regarded as the standard.1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of thi
6、s standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C859 Terminology Relating to Nuclear Materials3. Terminology3.1 For terminology related to radioactive materials, see Termi
7、nology C859.4. Summary of Test Method4.1 A sample is held within 6 mm of the detector window of a radiation survey meter, and the visible or audible reaction of themeter, or both, is noted. The user defines an application/project specific “negative” and “positive” result criteria. A “negative” testr
8、esult indicates radiation levels are below the user-established criteria; a “positive” test result indicates the radiation levels areabove the user-established criteria.5. Significance and Use5.1 Most facilities disposing or utilizing waste materials are prohibited from handling wastes that contain
9、radioactive materials.This test method provides the user a rapid method for screening waste material samples in the field or laboratory for the presenceor absence of radioactivity at user-established criteria. It is important to these facilities to be able to verify generator-suppliedinformation tha
10、t radioactive or mixed wastes have not been included in shipments of waste materials.6. Interferences6.1 Needle deflections or audible clicks of the survey meter, or both, occur due to naturally occurring omni-directionalbackground radiation. This level of ambient background radiation should be peri
11、odically assessed. See Section 10.1 This test method practice is under the jurisdiction of ASTM Committee D34 on Waste Management and is the direct responsibility of Subcommittee D34.01.05 onScreening Methods.Current edition approved Jan. 1, 2010Jan. 15, 2018. Published January 2010January 2018. Ori
12、ginally approved in 1996. Last previous edition approved in 20032010 asD5928 96D5928(2003). 96 (2010)1. DOI: 10.1520/D5928-96R10E01.10.1520/D5928-18.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standa
13、rdsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible t
14、o adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Con
15、shohocken, PA 19428-2959. United States16.2 Possible sources of interference include pacemakers, X-ray generating equipment, radium-based luminescent dials,polonium-based static eliminators, and smoke detectors containing a radioactive isotope sensing mechanism. Such interferencescan usually be trac
16、ed to their source using the portable instrument specified in this test method.6.3 Alarge amount of potassium in the waste sample may produce a positive result due to the natural presence of the radioactiveisotope, Potassium-40.6.4 The sensitivity of this test method to beta and gamma radiation may
17、be dependent on sample volume. A small samplevolume with readings near background levels may give a false negative result.6.5 Some radioactive isotopes, such as 3H and 14C, may not emit radiation of sufficient energy to be detected. If suspected tobe present in the waste, another procedure should be
18、 used that is appropriate to their determination.6.6 Liquid samples, as well as moisture in solid samples, are good attenuators of radiation and will hinder detection of manyradionuclides unless they emit high-energy gamma radiation. The possible inability to detect alpha particles and low-level bet
19、aemissions that may be attenuated, in many cases, should not be a serious shortcoming in this test method because these emissionsare often accompanied by higher energy gamma emissions.6.6.1 Moisture-laden Americium-241 bearing waste would be a case where there is a high probability of non-detection
20、due toattenuation.6.7 Survey meter contamination can cause artificially low or high results when reading standard sources or ambient background.If this occurs, the survey meter performance shall be considered unreliable and should be appropriately decontaminated byqualified personnel, or disposed of
21、 in accordance with applicable regulations.6.8 The radiation monitor should be operated in accordance with the manufacturers instructions.7. Apparatus7.1 Radiation Survey Meter, with a halogen-quenched uncompensated Geiger-Mueller tube with thin mica end-window. It isadvisable to choose an instrumen
22、t model that has an internal counter (cpm) or is capable of operating in a scaler mode.NOTE 1The meter used in the development of this test method was the Monitor 4.3Anumber of other survey meters4 are suitable for this test method.Through the end-window, this unit is capable of detecting and indica
23、ting alpha radiation (down to 2.5 MeV with a typical detection efficiency of 80 %at 3.6 MeV), beta radiation (down to 150 KeV with a typical 75 % detection efficiency), gamma radiation and X-rays (down to 10 KeV), with the surveymeter set on its lowest energy range of 0.1 Sv/h.8. Reagents and Materi
24、als8.1 The check sources used in the development of this test method were manufactured by Oxford Instruments, Inc.5 Othersuitable check sources are available from a number of suppliers.6 The check sources used in the development of this test methodare federally exempt and do not require licensing. S
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD59282018RED8125STANDARDPRACTICEFORSCREENINGOFWASTEFORRADIOACTIVITY 放射性 废弃物 筛选 标准 实施 规程 PDF

链接地址:http://www.mydoc123.com/p-521075.html