《ASTM D3792-2005 Standard Test Method for Water Content of Coatings by Direct Injection Into a Gas Chromatograph《直接注入式气相色谱法测定涂层中水含量的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D3792-2005 Standard Test Method for Water Content of Coatings by Direct Injection Into a Gas Chromatograph《直接注入式气相色谱法测定涂层中水含量的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 3792 05Standard Test Method forWater Content of Coatings by Direct Injection Into a GasChromatograph1This standard is issued under the fixed designation D 3792; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 This test method is for the det
3、ermination of the totalwater content of waterborne paints. It has been evaluated forlatex systems (styrene-butadiene, poly(vinylacetate)-acrylic,acrylic), epoxy acrylic resin systems and acrylic systems. Theestablished working range of this test method is from 15 to90 %. There is no reason to believ
4、e that it will not work outsideof this range.1.2 The values stated in SI units are to be regarded as thestandard. The values given in parentheses are for informationonly.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility
5、of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1193 Specification for Reagent WaterE 180 Practice for Determining Precision Data of ASTMMethods for
6、 Analysis and Testing of Industrial Chemicals3. Summary of Test Method3.1 A suitable aliquot of whole paint is internally standard-ized, diluted with dimethylformamide, and then injected into agas chromatographic column containing a porous polymerpacking that separates water from other volatile comp
7、onents.4. Significance and Use4.1 In order to calculate volatile organic content (VOC) inwaterborne paints, it is necessary to know the water content.This gas chromatographic test method provides a relativelysimple and direct way to determine water content.5. Apparatus5.1 Gas ChromatographyAny gas-l
8、iquid chromato-graphic instrument equipped with a thermoconductivity detec-tor may be used. Temperature programming capability ispreferable, but isothermal operations may be adequate. SeeTable 1.5.2 ColumnThe column should be at least 1.22 m (4 ft) of3.2-mm (18-in.) outside diameter tubing of stainl
9、ess steel, orother suitable material, lined with a TFE-fluorocarbon coatingpacked with 60/80 mesh (180 to 250 m) porous polymerpacking material.3A longer 1.83 m (6-ft) column can be usedto improve resolution.5.3 IntegratorAny electronic integrator that can accu-rately quantify a gas chromatographic
10、peak is acceptable.Alternatively, a recording potentiometer with a full-scaledeflection of 1-10mV, full scale response time of2sorless andsufficient sensitivity and stability to meet the requirements of5.1.5.4 Liquid Charging DevicesMicro syringes of 510-Lcapacity with a precision of 60.01 L. Automa
11、tic injection ofsamples improves the precision of this test method.6. Column Conditioning6.1 ProcedureInstall the packed column in the gas chro-matographic unit leaving the exit end disconnected from thedetector. This will prevent any contamination of the detectorwith the column bleed.6.1.1 Set the
12、carrier gas flow rate at 20 to 30 mL/min if a3.2-mm (18-in.) outside diameter column is used. Purge thecolumn 5 or 10 min before heating.6.1.2 Heat the column from room temperature to 200C at5C/min and hold this temperature for at least 12 h (overnight).At the end of this time, heat the column at 5C
13、/min to 250C1This test method is under the jurisdiction of ASTM Committee D01 on Paintand Related Coatings, Materials, and Applications and is the direct responsibility ofSubcommittee D01.21 on Chemical Analysis of Paints and Paint Materials.Current edition approved Jan. 1, 2005. Published February
14、2005. Originallyapproved in 1979. Last previous edition approved in 1999 as D 3792 99.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary
15、 page onthe ASTM website.3HayeSep R (silanized), was used in the round robin. Any other porous polymerpacking or other column giving equivalent or superior performance may be used.These products are available from most gas chromatography suppliers and distribu-tors.1Copyright ASTM International, 100
16、 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.(the maximum temperature for this packing) and hold forseveral hours. Cool the column to room temperature andconnect to the detector.6.1.3 Reheat the column to 250C at 5C/min to observe ifthere is column bleed. Optimum
17、conditioning of this columnmay take several cycles of the heating program before a flatrecorder baseline is achieved.6.2 Before each calibration and series of determinations (ordaily) condition the column at 200C for 1 h with carrier gasflow and to eliminate any residual volatile compounds retainedo
18、n the column.7. Reagents and Materials7.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specific
19、ations are available.4Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.7.2 Purity of WaterUnless otherwise indicated, referenceto water shall be understood to mean reagent wat
20、er conformingto Type II of Specification D 1193.7.3 Carrier GasHelium of 99.995 % or higher purity.High-purity nitrogen may also be used.NOTE 1Care should be taken that any moisture that may be present inthe carrier gas is eliminated through the use of a suitable carrier gaspurifier. Trace levels of
21、 water will accumulate on the column at low oventemperatures and may affect the reproducibility as well as the accuracy ofthe determination.7.4 Dimethylformamide (DMF) (Anhydrous) gas chroma-tography, spectrophotometric quality (See Note 2).7.5 2-Propanol (Anhydrous) (Isopropanol)See Note 2.7.6 Meth
22、anol (Anhydrous)See Note 2.7.7 Septum Sample Vials, 10-mL capacity withfluorocarbon-faced septa are preferred.7.8 Molecular Sieve, 2A-3A, 8-12, mesh.NOTE 2Dry the DMF, 2-propanol and methanol with the molecularsieve. Verify the absence of water by analysis of the solvents by this GCmethod.8. Hazards
23、8.1 Dimethylformamide is hazardous. Check the suppliersMaterial Safety Data Sheet (MSDS) before use.9. Preparation of Apparatus9.1 Install the column in the chromatograph and establishthe operating conditions required to give the desired separation(see Table 1).9.2 Allow sufficient time for the inst
24、rument to reach equi-librium as indicated by a stable base line.9.3 Control the detector temperature so that it is constant towithin 1C without thermostat cycling, which causes anuneven baseline.9.4 Adjust the carrier-gas flow to a constant value.10. Calibration10.1 Using the information in Table 1
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD37922005STANDARDTESTMETHODFORWATERCONTENTOFCOATINGSBYDIRECTINJECTIONINTOAGASCHROMATOGRAPH 直接 注入

链接地址:http://www.mydoc123.com/p-515514.html