1、Designation: D3104 14D3104 14aStandard Test Method forSoftening Point of Pitches (Mettler Softening Point Method)1This standard is issued under the fixed designation D3104; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year o
2、f last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of the softening point of pitches having softening points in the range fr
3、om 50 Cto 180C 180 C by this test method, and gives results comparable to those obtained by Test Method D2319 above 176F (80C).80 C (176 F).1.2 The values stated in SI units are to be regarded as the standard.1.3 This standard does not purport to address all of the safety concerns, if any, associate
4、d with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2A314 Specification for Stainless Steel Billets and Bars for Forgi
5、ngD2319 Test Method for Softening Point of Pitch (Cube-in-Air Method)D4296 Practice for Sampling Pitch3. Summary of Test Method3.1 In this test method, the softening point is defined as the temperature at which pitch, suspended in a cylindrical cup with a6.35-mm 6.35 mm hole in the bottom, flows dow
6、nward a distance of 19 mm 19 mm to interrupt a light beam, as the sample isheated at a linear rate in air.4. Significance and Use4.1 Pitch does not go through a solid-liquid phase change when heated, and therefore does not have a true melting point. Asthe temperature is raised pitch gradually soften
7、s or becomes less viscous. For this reason, the determination of the softening pointmust be made by an arbitrary, but closely defined, method if the test values are to be reproducible.4.2 This test method is useful in determining the consistency of pitches as one element in establishing the uniformi
8、ty ofshipments or sources of supply.5. Apparatus5.1 A Mettler METTLER TOLEDO dropping point cell3 shall be used to determine pitch softening points by this test method.These commercially available instruments consist of a control unit with a digital temperature indicator, matched furnace, withfurnac
9、e built in or attached, sample cartridges, and accessories. The control unit automatically regulates the heating rate of thefurnace. The softening point is indicated on the readout, and the heating program stopped, when the sample flow triggers aphotocell detector. the softening point detection. A g
10、eneral view of the contents of the Mettler a METTLER TOLEDO droppingpoint instrument is shown in Fig. 1. (old instrument) and Fig. 2 (new instrument).1 This test method is under the jurisdiction ofASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricantsand is the direct responsibility
11、of SubcommitteeD02.05 on Properties of Fuels, Petroleum Coke and Carbon Material.Current edition approved June 1, 2014Oct. 1, 2014. Published July 2014November 2014. Originally approved in 1972. Last previous edition approved in 20102014 asD3104 99 (2010).D3104 14. DOI: 10.1520/D3104-14.10.1520/D310
12、4-14A.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 The sole source of supply of the apparatus known to t
13、he committee at this time is available from the Mettler Toledo, Inc., Balances and Instruments, 69Princeton-Hightstown Rd., Hightstown, NJ 08520-0071. Mettler-Toledo, LLC., 1900 Polaris Pkwy, Columbus, OH 43240, . If you are aware of alternativesuppliers, please provide this information to ASTM Inte
14、rnational Headquarters. Your comments will receive careful consideration at a meeting of the responsible technicalcommittee,1 which you may attend.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the pr
15、evious version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of C
16、hanges section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.1.1 Integrated or Separate Control UnitThe control unit shall provide a continuous, linear temperature increase from 25 Cto 250 C at a
17、 rate of 2 C/min. A digital readout shall indicate the temperature to 0.1 C throughout.5.1.2 Control and Integrated or Separate Furnace UnitThis shall provide a continuous, linear temperature increase from 25to 250C at a rate of 2C/min.Adigital readout shall indicate the temperature to 0.1C througho
18、ut. This shall The furnace unit shallbe capable of heating one or two sample cup assemblies, as described in 5.1.25.1.3, at a linear rate of 2 C 60.3C/min.0.3 Cmin. It shall include a sensing system capable of detecting the softening point with a precision of 0.1C.0.1 C.5.1.3 Sample Cup AssemblyA cu
19、p of chromium-plated brass, or of stainless steel, aluminum, or of aluminum, stainless steelconforming to the requirements for Type 303 (UNS S30300) stainless steel as prescribed in Specification A314, with thedimensions shown in Fig. 23. It shall be placed in the assembly so that the pitch sample s
20、oftening point will be detected when ithas flowed down a distance of 19 mm.19 mm.6. Reagents6.1 Xylene, Xylene,industrial grade.6.2 Benzoic Acid.Benzoic Acid.FIG. 1 General View of the Mettler FP-5/53METTLER TOLEDO FP90 Control Unit with Heater FP83HTD3104 14a27. Calibration of the Mettler METTLER T
21、OLEDO Apparatus7.1 This step, required only occasionally, is designed to establish that the temperature indicated by the instrument is inagreement with a known standard.Aspecial cup with a bottom orifice of 2.8 mm is used instead of the one prescribed for the testingof pitch.7.2 ReagentUse either an
22、alytical reagent or primary standard grade benzoic acid for the calibration. As this material ishygroscopic it must be stored in a tightly sealed container, and replaced with fresh material from a newly opened supply ifhydration or other contamination is suspected.7.3 Procedure: ProcedureOld Instrum
23、ents:7.3.1 Filling the Sample CupPlace the cup on a clean, flat surface. Add a small amount of benzoic acid crystals and pressdown with a rod (approximately 4.5 mm in diameter). Check that the bottom orifice is completely filled. Refill and repeat thepressing step until the cup is filled with benzoi
24、c acid. Remove any crystals from the exterior of the cup.7.3.2 HeatingPreheat the Mettler furnace to 121C, 121 C, and maintain it at that temperature. Place the cartridge assemblycontaining the benzoic acid in position in the furnace, taking care that the slits for the light beam are properly positi
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD31042014ASTANDARDTESTMETHODFORSOFTENINGPOINTOFPITCHESMETTLERSOFTENINGPOINTMETHOD 沥青 软化 标准 试验 方法

链接地址:http://www.mydoc123.com/p-514082.html