ASTM D2268-1993(2003)e1 Standard Test Method for Analysis of High-Purity n-Heptane and Isooctane by Capillary Gas Chromatography《用毛细管气相色谱法分析高纯度n-庚烷和异辛烷的标准试验方法》.pdf
《ASTM D2268-1993(2003)e1 Standard Test Method for Analysis of High-Purity n-Heptane and Isooctane by Capillary Gas Chromatography《用毛细管气相色谱法分析高纯度n-庚烷和异辛烷的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D2268-1993(2003)e1 Standard Test Method for Analysis of High-Purity n-Heptane and Isooctane by Capillary Gas Chromatography《用毛细管气相色谱法分析高纯度n-庚烷和异辛烷的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 2268 93 (Reapproved 2003)e1An American National StandardStandard Test Method forAnalysis of High-Purity n-Heptane and Isooctane by CapillaryGas Chromatography1This standard is issued under the fixed designation D 2268; the number immediately following the designation indicates the yea
2、r oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Departm
3、ent of Defense.e1NOTEWarning notes were editorially moved into the standard text in July 2003.1. Scope1.1 This test method provides for the analysis of high-purity(greater than 99.5 % by volume) n-heptane and isooctane(2,2,4-trimethylpentane), which are used as primary referencestandards in determin
4、ing the octane number of a fuel. Indi-vidual compounds present in concentrations of less than0.01 % can be detected. Columns specified by this test methodmay not allow separation of all impurities in reference fuels.1.2 The values stated in SI units are to be regarded as thestandard. The values give
5、n in parentheses are for informationonly.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory
6、 limitations prior to use.2. Summary of Test Method2.1 The sample is injected into a capillary gas chromato-graphic column consisting of at least 61 m (200 ft) of stainlesssteel tubing (0.25-mm (0.010-in.) inside diameter), the innerwalls of which are coated with a thin film of stationary liquid.An
7、inert gas transports the sample through the column, inwhich it is partitioned into its individual components. As eachcomponent is eluted from the column, it is detected with ahydrogen flame ionization detector and recorded on a conven-tional strip-chart recording potentiometer. The detector re-spons
8、e from each impurity is then compared with that of aknown quantity of an internal standard. After determining thetotal impurity concentration, the n-heptane, or isooctane purityis obtained by difference.3. Significance and Use3.1 This test method is used for specification analysis ofhigh-purity n-he
9、ptane and isooctane, which are used as ASTMKnock Test Reference Fuels. Hydrocarbon impurities or con-taminants, which can adversely affect the octane number ofthese fuels, are precisely determined by this method.4. Apparatus4.1 ChromatographGas chromatograph should beequipped with a split-stream inl
10、et device for introducingminute quantities of sample without fractionation, a capillarycolumn, and a hydrogen flame ionization detector. An elec-trometer to amplify the low output signal of the hydrogen flameionization detector, and a strip-chart recorder for recording thedetector signal are needed.
11、 The time constant of neither theelectrometer nor the recorder should exceed 1 s.Aball and diskintegrator or electronic integrator for peak area measurementsshould be used. The detection system must have sufficientsensitivity to produce a recorder deflection for cyclohexane ofat least 8 divisions on
12、 a standard 0100 scale chart using 0.10volume percent of cyclohexane in n-heptane as defined in 7.1.4.2 MicrosyringeA microsyringe is needed for injectingthe sample into the split-stream inlet device.4.3 Volumetric Pipet, 0.1-mL capacity.4.4 Analytical Balance, 200-g capacity.5. Reagents and Materia
13、ls5.1 Carrier GasArgon, Nitrogen, or Helium; 99.99% orgreater purity. (WarningCompressed gases under high pres-sure.)5.2 Fuel GasHydrogen; 99.99% or greater purity.(WarningCompressed gas under high pressure. Extremelyflammable gas.)5.3 Oxidant GasAir; 99.99% or greater purity.(WarningCompressed gase
14、s under high pressure.)1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of SubcommitteeD02.04 on Hydrocarbon Analysis.Current edition approved May 10, 2003. Published July 2003. Originallyapproved in 1964. Last pre
15、vious edition approved in 1998 as D 226893 (1998).1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.4 CyclohexaneAt least 99 mol % pure, to be used asinternal standard. (WarningFlammable liquid and harmful ifingested or inhaled.)5.5
16、n-PentaneCommercial grade. (WarningVolatileand flammable liquid, and harmful if ingested or inhaled.)5.6 Isooctane (2,2,4-trimethylpentane)(WarningFlammable liquid and harmful if ingested or inhaled.)5.7 SqualaneLiquid phase for gas chromatographic col-umns.5.8 TubingType 316, 321, or 347 stainless
17、steel; 0.25 mm(0.010 in.) inside diameter.6. Preparation of Resolving ColumnNOTE 1There are many different procedures for coating capillarycolumns. A suitable procedure is given in 6.1 through 6.3. Other columnsmay be used provided they meet resolution and repeatability requirementsof the method.6.1
18、 Connect a 229 mm (9-in.) section of stainless steeltubing 6.4 mm (14-in.) outside diameter, total volume ofapproximately 5 mL) to a high-pressure cylinder of argon,helium, or nitrogen through a pressure regulator. Connect atleast 61 m (200 ft) of Type 316, 321, or 347 stainless steeltubing (0.25-mm
19、 (0.010-in.) inside diameter) to the 229-mmsection of 64 mm tubing which is to be used as a reservoir forthe coating solution. The capillary column is generally coiledon a suitable mandrel before coating. To the other end of thecapillary column, connect an additional 30 to 9 to 12 m (40 ft)of capill
20、ary tubing through a 1.6 mm (116-in.) Swagelok union.6.2 Clean the tubing by passing 25 to 30 mL (5 to 6reservoir volumes) of n-pentane through the tubing with about1.7 to 2.1 MPa ( 250 to 300 psig gage) of inert gas. After thecolumn has been cleaned, disconnect the upstream end of thereservoir tube
21、 and allow the pressure in the tubing to return toatmospheric.6.3 Prepare a solution containing 6 volume percent ofsqualane in n-pentane. Fill the reservoir tube with the coatingsolution and promptly connect to the gas cylinder. Pass thecoating solution through the column at 500 psig (3.5 MPagage) u
22、ntil the solution begins issuing from the end of thecapillary tubing; gradually reduce the inlet pressure in order tokeep the flow of the solution at a relatively even rate of 40 to60 drops/min. When the coating solution has been expelledfrom the column, reduce the inlet pressure to 345 kPa (50 psig
23、gage) and allow gas to pass through the column for 1 to 2 h.Disconnect the 9 to 12-m (30 to 40-ft) tail section and thenmount the column in the chromatograph.6.4 To test column resolution use Fig. 1 and calculate R,from the distance between the cyclohexane and n-heptanepeaks at the peak maxima, d, a
24、nd the widths of the peaks at thebaseline, Y1and Y2.R 5 2d12 d2!/Y11 Y2! (1)Resolution (R), using the above equation, must exceed avalue of 10.7. Sample Preparation7.1 Place 20 to 30 mL of the reference fuel (n-heptane orisooctane) into a 100-mL volumetric flask which has beenpreviously weighed.7.2
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD226819932003E1STANDARDTESTMETHODFORANALYSISOFHIGHPURITYNHEPTANEANDISOOCTANEBYCAPILLARYGASCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-511346.html