ASTM C1590-2004(2009) Standard Practice for Alternate Actinide Calibration for Inductively Coupled Plasma-Mass Spectrometry《电感耦合等离子体质谱法的交替锕系校正的标准实施规程》.pdf
《ASTM C1590-2004(2009) Standard Practice for Alternate Actinide Calibration for Inductively Coupled Plasma-Mass Spectrometry《电感耦合等离子体质谱法的交替锕系校正的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM C1590-2004(2009) Standard Practice for Alternate Actinide Calibration for Inductively Coupled Plasma-Mass Spectrometry《电感耦合等离子体质谱法的交替锕系校正的标准实施规程》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1590 04 (Reapproved 2009)Standard Practice forAlternate Actinide Calibration for Inductively CoupledPlasma-Mass Spectrometry1This standard is issued under the fixed designation C 1590; the number immediately following the designation indicates the year oforiginal adoption or, in the c
2、ase of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides guidance for an alternate linearcalibration for the determination
3、of selected actinide isotopesin appropriately prepared aqueous solutions by InductivelyCoupled Plasma-Mass Spectrometry (ICP-MS). This alternatecalibration is mass bias adjusted using thorium-232 (232Th) anduranium-238 (238U) standards. One of the benefits of thisstandard practice is the ability to
4、calibrate for the analysis ofhighly radioactive actinides using calibration standards atmuch lower specific activities. Environmental laboratories mayfind this standard practice useful if facilities are not available tohandle the highly radioactive standards of the individualactinides of interest.1.
5、2 The instrument response for a series of determinationsof known concentration of232Th and238U defines the massversus response relationship. For each standard concentration,the slope of the line defined by232Th and238U is used to derivelinear calibration curves for each mass of interest usinginterfe
6、rence equations. The mass bias corrected calibrationcurves, although generated from interference equations, arespecific to the instrument operating parameters and tuning ineffect at the time of data acquisition. Because interferenceequations are part of the normal ICP-MS manufacturerssoftware packag
7、e, this calibration methodology is widelyapplicable.1.3 For this standard practice, the actinide atomic massrange that has been studied is from amu 232244. Guidance foran extended range of amu 228248 is given in this practice.1.4 Using this practice, analyte concentrations are reportedat each amu an
8、d not by element total (that is,239Pu versusplutonium).1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is therespon
9、sibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C 859 Terminology Relating to Nuclear MaterialsC 1168 Practice for Preparation and Dissolutio
10、n of Pluto-nium Materials for AnalysisC 1347 Practice for Preparation and Dissolution of UraniumMaterials for AnalysisC 1411 Practice for The Ion Exchange Separation of Ura-nium and Plutonium Prior to Isotopic AnalysisC 1414 Practice for The Separation of Americium fromPlutonium by Ion ExchangeC 146
11、3 Practices for Dissolving Glass Containing Radioac-tive and Mixed Waste for Chemical and RadiochemicalAnalysisD 1193 Specification for Reagent Water3. Terminology3.1 Definitions:3.1.1 For definitions of terms relating to nuclear materials,refer to Terminology C 859.3.1.2 AMUatomic mass unit.4. Summ
12、ary of Practice4.1 Calibration for the actinides by ICP-MS can be per-formed in a variety of ways with varying degrees of dataquality. An alternative calibration method has been developedto compensate for instrument mass bias using a generated massresponse curve defined by the232Th and238U data poin
13、ts. Themass response curve defined by232Th and238U approximatesthe mass response curve from amu 232244 as verifiedexperimentally and graphically depicted in Fig. 1. The massresponse curve shown reflects the operating parameters andtune of the particular instrument in use at the time of datacollectio
14、n. Different tuning parameters or instrumentationcould result in varying degrees of negative, neutral, or positive1This practice is under the jurisdiction of ASTM Committee C26 on NuclearFuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved Jun
15、e 1, 2009. Published July 2009. Originally approvedin 2004. Last previous edition approved in 2004 as C 1590 04.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the
16、standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.mass bias. Because the mass response curve defined by232Thand238U used in this standard practice is determined duringeach calibration,
17、 all potential linear variations in mass bias arecompensated for.4.2 The alternative calibration in this standard practicecombines the features of an external linear calibration at eachmass of interest with the mass bias correction of a mass/response curve. The correction for mass bias is integrated
18、 intothe acquisition of the standard data through the use ofinterference equations which, are part of the normal softwarepackage for correction of isobaric interferences in ICP-MSanalyses. Multipoint calibration curves are generated at eachmass of interest, resulting in more accurate quantification
19、thanthat of the typical “semi-quantitative” single point calibrationbased on the mass/response curve alone.4.3 Sample analyses for blanks and samples are performedusing a data acquisition method file without the interferenceequations that were used to derive the calibration curves. Aftercalibration
20、curves are generated, the calibration files are copiedor linked to the analytical acquisition method. The sampleresponses are acquired at each mass, and concentrationscalculated from the mass bias corrected calibration curves.Some ICP-MS vendor supplied control software will permit thelinking of sep
21、arate calibration and acquisition files (that is, youcan choose which calibration files to use to quantitate anyparticular data set regardless of the acquisition file that wasused to acquire the data).4.4 Mixed calibration standard solutions are preparedthrough dilution of single element stock stand
22、ards of thoriumand known abundance uranium (normally depleted in235U)with dilute nitric acid to develop a calibration series coveringthe desired concentration range. Standard concentrations arecalculated for232Th and238U for each calibration solution.4.5 Bismuth-209 (209Bi) is used as an internal st
23、andard andis added in a fixed quantity in all standards and samples tocorrect for both instrument drift and physical sample transportfluctuations.5. Significance and Use5.1 One of the benefits of this standard practice is the abilityto calibrate for the analysis of highly radioactive actinidesusing
24、calibration standards at much lower specific activities(that is,232Th and238U). Environmental laboratories may findthis standard practice useful if facilities are not available tohandle the highly radioactive standards of the individualactinides of interest.5.2 The degree of actual mass bias is vari
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC159020042009STANDARDPRACTICEFORALTERNATEACTINIDECALIBRATIONFORINDUCTIVELYCOUPLEDPLASMAMASSSPECTROMETRY

链接地址:http://www.mydoc123.com/p-465237.html