ASTM ISO ASTM51538-2017 Standard Practice for Use of the Ethanol-Chlorobenzene Dosimetry System《使用乙醇-氯苯放射量系统的标准实施规程》.pdf
《ASTM ISO ASTM51538-2017 Standard Practice for Use of the Ethanol-Chlorobenzene Dosimetry System《使用乙醇-氯苯放射量系统的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM ISO ASTM51538-2017 Standard Practice for Use of the Ethanol-Chlorobenzene Dosimetry System《使用乙醇-氯苯放射量系统的标准实施规程》.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、ISO/ASTM 51538:2009(E)ISO/ASTM 51538 2017(E)Standard Practice forUse of the Ethanol-Chlorobenzene Dosimetry System1This standard is issued under the fixed designation ISO/ASTM 51538; the number immediately following the designation indicates theyear of original adoption or, in the case of revision,
2、the year of last revision.1. Scope1.1 This practice covers the procedure for preparation, handling, testing, and use of procedure for using the ethanol-chlorobenzene (ECB) dosimetry system to determinemeasure absorbed dose (in terms of absorbed dose to water) in materialsirradiated by photons (gamma
3、 radiation or X-radiation/bremsstrahlung) or high energy electrons. to water when exposed toionizing radiation. The system consists of a dosimeter and appropriate analytical instrumentation. For simplicity, the system willbe referred to as the ECB system. It The ECB dosimeter is classified as a refe
4、rence-standard dosimetry system and is also typeI dosimeter on the basis of the effect of influence quantities. The ECB dosimetry system may be used as a routine referencestandard dosimetry system (see ISO/ASTM Guideor as a 51261). routine dosimetry system.1.2 This document is one of a set of standa
5、rds that provides recommendations for properly implementing dosimetry in radiationprocessing, and describes a means of achieving compliance with the requirements of ISO/ASTM Practice 52628 for the ECBsystem. It is intended to be read in conjunction with ISO/ASTM Practice 52628.1.3 This practice desc
6、ribes the mercurimetric titration analysis as a standard readout procedure for the ECB dosimeter whenused as a reference standard dosimetry system. Other readout methods (spectrophotometric, oscillometric) that are applicable whenthe ECB system is used as a routine dosimetry system are described in
7、Annex A1 and Annex A2Annex A1 and Annex A2.1.4 This practice applies only to gamma radiation, X-radiation/bremsstrahlung, and high energy electrons.1.5 This practice applies provided the following conditions are satisfied:1.5.1 The absorbed dose range is between 10 Gy and 2 MGy for gamma radiation a
8、nd between 10 Gy and 200 kGy for highcurrent electron accelerators (1, 2).2 (Warningthe boiling point of ethanol chlorobenzene solutions is approximately80C.80 C. Ampoules may explode if the temperature during irradiation exceeds the boiling point. This boiling point may beexceeded if an absorbed do
9、se greater than 200 kGy is given in a short period of time.)1.5.2 The absorbed-dose rate is less than 106 Gy s1(2).1.5.3 For radionuclide gamma-ray sources, the initial photon energy is greater than 0.6 MeV. For bremsstrahlung photons, theenergy of the electrons used to produce the bremsstrahlung ph
10、otons is equal to or greater than 2 MeV. For electron beams, the initialelectron energy is equal to or greater than 48 MeV (3). (see ICRU Reports 34 and 35).NOTE 1The same response relative to 60Co gamma radiation was obtained in high-power bremsstrahlung irradiation produced by a 5 MeV electronacce
11、lerator (4).NOTE 2The same response relative to 60Co gamma radiation was obtained in high-power bremsstrahlung irradiation produced by a 5 MeV electronaccelerator (4). The lower limits of energy givenlimits are appropriate for a cylindrical dosimeter ampoule of 12-mm diameter. Corrections for dosegr
12、adients across an ampoule of that diameter or less are not required. the ampoule may be required for electron beams. The ECB system may be usedat energies of incident electrons lower than 4 MeV lower energies by employing thinner (in the beam direction) dosimeters. dosimeters (see ICRU Report35). Th
13、e ECB system may also be used at X-ray energies as low as 120 kVp (5). However, in this range of photon energies the effect caused by theampoule wall is considerable.NOTE 3The effects of size and shape of the dosimeter on the response of the dosimeter can adequately be taken into account by performi
14、ng theappropriate calculations using cavity theory (6).1.5.4 The irradiation temperature of the dosimeter is within the range from 40C to 80C.30 C to 80 C.NOTE 4The temperature dependence of dosimeter response is known only in this range (see 4.35.2). For use outside this range, the dosimetry system
15、should be calibrated for the required range of irradiation temperatures.1 This practice is under the jurisdiction ofASTM Committee E61 on Radiation Processing and is the direct responsibility of Subcommittee E61.02 on Dosimetry Systems,and is also under the jurisdiction of ISO/TC 85/WG 3.Current edi
16、tion approved June 18, 2008.April 25, 2017. Published June 2009.2017. Originally published as ASTM E1538-93. Last previous ASTM edition E153899.ASTM E153893 was adopted by ISO in 1998 with the intermediate designation ISO 15563:1998(E). The present International Standard ISO/ASTM 51538:200951538:201
17、7(E) is a major revision of ISO/ASTM 51538:2002(E), which replaced ISO 15563.51538:2009(E). DOI:10.1520/ISOASTM51538-17.2 The boldface numbers in parentheses refer to the bibliography at the end of this practice.This document is not an ASTM standard and is intended only to provide the user of an AST
18、M standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published
19、by ASTM is to be considered the official document. ISO/ASTM International 2017 All rights reserved11.4 The effects of size and shape of the dosimeter on the response of the dosimeter can adequately be taken into account byperforming the appropriate calculations using cavity theory (6).1.6 This stand
20、ard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use. Specific warnings are given in
21、 1.5.1, 8.29.2 and 9.210.2.1.7 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade Organizati
22、on Technical Barriers to Trade (TBT) Committee.2. Referenced documents2.1 ASTM Standards:3C912 Practice for Designing a Process for Cleaning Technical GlassesD1193 Specification for Reagent WaterE170 Terminology Relating to Radiation Measurements and DosimetryE275 Practice for Describing and Measuri
23、ng Performance of Ultraviolet and Visible SpectrophotometersE666 Practice for Calculating Absorbed Dose From Gamma or X RadiationE668 Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining Absorbed Dose inRadiation-Hardness Testing of Electronic DevicesE925 Practice f
24、or Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidth does notExceed 2 nmE958 Practice for Estimation of the Spectral Bandwidth of Ultraviolet-Visible Spectrophotometers2.2 ISO/ASTM Standards:351261 GuidePractice for Selection and Calibration of Routine Dos
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMISOASTM515382017STANDARDPRACTICEFORUSEOFTHEETHANOLCHLOROBENZENEDOSIMETRYSYSTEM 使用 乙醇 氯苯 放射 系统 标准

链接地址:http://www.mydoc123.com/p-540773.html