ASTM ISO ASTM51205-2017 Standard Practice for Use of a Ceric-Cerous Sulfate Dosimetry System《使用硫酸铈和硫酸高铈剂量测定系统的标准实施规程》.pdf
《ASTM ISO ASTM51205-2017 Standard Practice for Use of a Ceric-Cerous Sulfate Dosimetry System《使用硫酸铈和硫酸高铈剂量测定系统的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM ISO ASTM51205-2017 Standard Practice for Use of a Ceric-Cerous Sulfate Dosimetry System《使用硫酸铈和硫酸高铈剂量测定系统的标准实施规程》.pdf(19页珍藏版)》请在麦多课文档分享上搜索。
1、ISO/ASTM 51205:2009(E)ISO/ASTM 51205 2017(E)Standard Practice forUse of a Ceric-Cerous Sulfate Dosimetry System1This standard is issued under the fixed designation ISO/ASTM 51205; the number immediately following the designation indicates theyear of original adoption or, in the case of revision, the
2、 year of last revision.1. Scope1.1 This practice covers the procedures for preparation, testing, and procedure for using the ceric-cerous sulfate dosimetrysystem to determinemeasure absorbed dose (in terms of absorbed dose to water) in materials irradiated by photons (gammaradiation or X-radiation/b
3、remsstrahlung) or high-energy electrons. to water when exposed to ionizing radiation. The systemconsists of a dosimeter and appropriate analytical instrumentation. For simplicity, the system will be referred to as the ceric-ceroussystem. It The ceric-cerous dosimeter is classified as a referencestan
4、dard dosimetry system (see ISO/ASTM Guidetype 1dosimeter on the basis 51261). Ceric-cerous dosimeters are also used as transferstandard dosimeters or routine dosimeters.of theeffect of influence quantities. The ceric-cerous system may be used as a reference standard dosimetry system or as a routined
5、osimetry system.1.2 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiationprocessing, and describes a means of achieving compliance with the requirements of ISO/ASTM Practice 52628 for theceric-cerous system. It is intended to be rea
6、d in conjunction with ISO/ASTM Practice 52628.1.3 This practice describes both the spectrophotometric and the potentiometric readout procedures for the ceric-cerous system.1.4 This practice applies only to gamma radiation, X-radiation/bremsstrahlung, and high energy electrons.1.5 This practice appli
7、es provided the following conditions are satisfied:1.5.1 The absorbed-dose range is between 0.5 and 50from 5 102 kGy to 5 104 Gy (1).21.5.2 The absorbed-dose rate is less thandoes not exceed 106 Gy s1 (1).1.5.3 For radionuclide gamma-ray sources, the initial photon energy is greater than 0.6 MeV. Fo
8、r bremsstrahlung photons, theinitial energy of the electrons used to produce the bremsstrahlung photons is equal to or greater than 2 MeV. For electron beams,the initial electron energy is greater than 8 MeV.NOTE 1The lower energy limits are appropriate for a cylindrical dosimeter ampoule of 12-mm d
9、iameter. Corrections for dose gradientsgradient acrossan ampoule of that diameter or less are not required for photons, but the ampoule may be required for electron beams (2). The ceric-cerous system maybe used at lower energies by employing thinner (in the beam direction) dosimeters.dosimeters (see
10、 ICRU Report 35).1.5.4 The irradiation temperature of the dosimeter is above 0C and below 62C (3).NOTE 2The temperature dependencecoefficient of dosimeter response is known only in this range (see 4.35.2). Use outside this range requiresdetermination of the temperature dependence.coefficient.1.6 Thi
11、s standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.1.7 This international s
12、tandard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Committee.1 Th
13、is guidepractice is under the jurisdiction of ASTM Committee E61 on Radiation Processing and is the direct responsibility of Subcommittee E61.02 on DosimetrySystems, and is also under the jurisdiction of ISO/TC 85/WG 3.Current edition approved June 18, 2008. Published June 2009.March 8, 2017. Publis
14、hed May 2017. Originally published asASTM E120588. Last previousASTM editionE120599. ASTM E120593 was adopted by ISO in 1998 with the intermediate designation ISO 15555:1998(E). The present International Standard ISO/ASTM51205:2009(E)51205:2017(E) is a major revision of ISO/ASTM 51205-2002(E) which
15、replaced ISO 15555.51205-2009(E). DOI:10.1520/ISOASTM51205-17.2 The boldface numbers in parentheses refer to the bibliography at the end of this standard.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to
16、 the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document. ISO/AST
17、M International 2017 All rights reserved12. Referenced documents2.1 ASTM Standards:3C912 Practice for Designing a Process for Cleaning Technical GlassesE170 Terminology Relating to Radiation Measurements and DosimetryE178 Practice for Dealing With Outlying ObservationsE275 Practice for Describing an
18、d Measuring Performance of Ultraviolet and Visible SpectrophotometersE666 Practice for Calculating Absorbed Dose From Gamma or X RadiationE668 Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining Absorbed Dose inRadiation-Hardness Testing of Electronic DevicesE925 P
19、ractice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidth does notExceed 2 nmE958 Practice for Estimation of the Spectral Bandwidth of Ultraviolet-Visible Spectrophotometers2.2 ISO/ASTM Standards:351261 GuidePractice for Selection and Calibration of Ro
20、utine Dosimetry Systems for Radiation Processing51707 Guide for Estimation of Measurement Uncertainty in Dosimetry for Radiation Processing5140052628 Practice for Characterization and Performance of a High-Dose Radiation Dosimetry Calibration LaboratoryDo-simetry in Radiation Processing5170752701 Gu
21、ide for Estimating Uncertainties in Dosimetry for Performance Characterization of Dosimeters and DosimetrySystems for Use in Radiation Processing2.3 ISO Standards:412749-4 Nuclear energy Vocabulary Part 4: Dosimetry for radiation processing2.4 ISO/IEC Standards:417025 General Requirements for the Co
22、mpetence of Testing and Calibration Laboratories2.5 Joint Committee for Guides in Metrology (JCGM) Reports:JCGM 100:2008, GUM 1995, with minor corrections, Evaluation of measurement data Guide to the Expression of Uncertaintyin Measurement5JCGM 200:2012 (JCGM 200:2008 with minor revisions),VIM, Inte
23、rnational Vocabulary of Metrology Basis and GeneralConcepts and Associated Terms62.6 International Commission on Radiation Units and Measurements (ICRU) Reports:7ICRU Report 1410b (NBS Handbook 85) Radiation Dosimetry: X-Rays and Gamma Rays with Maximum Photon EnergiesBetween 0.6 and 60 MeVPhysical
24、Aspects of IrradiationICRU Report 34 The Dosimetry of Pulsed RadiationICRU Report 35 Radiation Dosimetry: Electrons Electron Beams with Initial Energies Between 1 and 50 MeVICRU Report 3780 Stopping Powers for Electrons and PositronsDosimetry Systems for Use in Radiation ProcessingICRU Report 6085a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMISOASTM512052017STANDARDPRACTICEFORUSEOFACERICCEROUSSULFATEDOSIMETRYSYSTEM 使用 硫酸 剂量 测定 系统 标准 实施 规程

链接地址:http://www.mydoc123.com/p-540771.html