《ASTM E664 E664M-2010 1250 Standard Practice for the Measurement of the Apparent Attenuation of Longitudinal Ultrasonic Waves by Immersion Method《浸渍法测量纵向超声波的表面衰减的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E664 E664M-2010 1250 Standard Practice for the Measurement of the Apparent Attenuation of Longitudinal Ultrasonic Waves by Immersion Method《浸渍法测量纵向超声波的表面衰减的标准实施规程》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E664/E664M 10Standard Practice for theMeasurement of the Apparent Attenuation of LongitudinalUltrasonic Waves by Immersion Method1This standard is issued under the fixed designation E664/E664M; the number immediately following the designation indicates the yearof original adoption or, i
2、n the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval.A superscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes a procedure for measuring theapparent attenuation of ultra
3、sound in materials or componentswith flat, parallel surfaces using conventional pulse-echoultrasonic flaw detection equipment in which reflected indica-tions are displayed in an A-scan presentation.1.2 The measurement procedure is readily adaptable for thedetermination of relative attenuation betwee
4、n materials. Forabsolute (true) attenuation measurements, indicative of theintrinsic nature of the material, it is necessary to correct forspecimen geometry, sound beam divergence, instrumentation,and procedural effects. These results can be obtained with morespecialized ultrasonic equipment and tec
5、hniques.1.3 The values stated in either SI units or inch-pound unitsare to be regarded separately as standard. The values stated ineach system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-conforma
6、ncewith the standard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior t
7、o use.2. Referenced Documents2.1 ASTM Standards:2E214 Practice for Immersed Ultrasonic Testing by theReflection Method Using Pulsed Longitudinal Waves3E317 Practice for Evaluating Performance Characteristicsof Ultrasonic Pulse-Echo Testing Instruments and Systemswithout the Use of Electronic Measure
8、ment InstrumentsE1316 Terminology for Nondestructive Examinations3. Terminology3.1 DefinitionsFor definitions of terms used in this prac-tice, see Terminology E1316.4. Summary of Practice4.1 This practice describes a procedure for determiningapparent attenuation by measuring the decay of multiple ba
9、ckreflections of longitudinal ultrasonic waves introduced intospecimens with flat, parallel surfaces by the immersion tech-nique.5. Significance and Use5.1 The measurement of apparent attenuation in materials isuseful in applications such as the comparison of heat treat-ments of different lots of ma
10、terial or the assessment of thedegradation of materials due to environment.5.2 Several different modes of wave vibration can bepropagated in solids. This practice is concerned with theattenuation associated with longitudinal waves introduced intothe specimen by the immersion method.5.3 This practice
11、 allows for the comparison of the apparentattenuations of geometrically similar specimens.5.4 For the determination of apparent attenuation, the pro-cedures described herein are valid only for measurements inthe far field of the ultrasonic beam.6. Apparatus6.1 Ultrasonic Flaw Detection System A syst
12、em capableof generating, receiving, and displaying electrical pulses at thefrequency of interest. Display shall be an A-scan presentation.6.1.1 Performance CharacteristicsThe vertical linearitylimits shall be determined as specified in Practice E317. Allmeasurements shall be made only within the lin
13、ear ranges ofthe system.6.2 Search UnitThe size and frequency should be deter-mined to suit the application, and only non-focused searchunits may be used.1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommittee E07.06
14、onUltrasonic Method.Current edition approved June 1, 2010. Published July 2010. Originally approvedin 1978. Last previous edition approved in 2005 as E664 - 93(2005). DOI:10.1520/E0664_E0664M-10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at
15、serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn. The last approved version of this historical standard is referencedon www.astm.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West
16、 Conshohocken, PA 19428-2959, United States.6.3 CouplantNormally water. See Practice E214 for alter-natives.6.4 Reference BlockThe use of a reference block issuggested to evaluate the stability of the measurement systemif measurements will be made over a period of time. Thereference block should hav
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME664E664M20101250STANDARDPRACTICEFORTHEMEASUREMENTOFTHEAPPARENTATTENUATIONOFLONGITUDINALULTRASONICWAVESBYIMMERSIONMETHOD

链接地址:http://www.mydoc123.com/p-533264.html