ASTM E2069-2006 Standard Test Method for Temperature Calibration on Cooling of Differential Scanning Calorimeters《差示扫描热量计制冷却温度校准用标准试验方法》.pdf
《ASTM E2069-2006 Standard Test Method for Temperature Calibration on Cooling of Differential Scanning Calorimeters《差示扫描热量计制冷却温度校准用标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2069-2006 Standard Test Method for Temperature Calibration on Cooling of Differential Scanning Calorimeters《差示扫描热量计制冷却温度校准用标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2069 06Standard Test Method forTemperature Calibration on Cooling of Differential ScanningCalorimeters1This standard is issued under the fixed designation E 2069; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the y
2、ear of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the temperature calibration ofdifferential scanning calorimeters on cooling using the
3、differ-ence between transition temperatures upon heating and coolingin the temperature range of 50 to 185 C. An offset in theindicated temperature between heating and cooling experi-ments, within this temperature range, may be used to providetemperature calibration on cooling at other temperature ra
4、nges.1.2 SI units are the standard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limit
5、ations prior to use. Specific precau-tionary statements are given in Section 6.2. Referenced Documents2.1 ASTM Standards:2D 3418 Test Method for Transition Temperatures of Poly-mers By Differential Scanning CalorimetryE 473 Terminology Relating to Thermal Analysis and Rhe-ologyE 794 Test Method for
6、Melting And Crystallization Tem-peratures By Thermal AnalysisE 928 Test Method for Purity by Differential ScanningCalorimetryE 967 Test Method for Temperature Calibration of Differ-ential Scanning Calorimeters and Differential ThermalAnalyzersE 1970 Practice for Statistical Treatment of Thermoanalyt
7、i-cal Data3. Terminology3.1 Specific technical terms used in this test method aredefined in Terminology E 473.4. Summary of Test Method4.1 The temperature sensor of the DSC, used to determinethe temperature of a transition, is located close to but externalto the test specimen. This arrangement cause
8、s the indicatedtemperature to lead or lag the actual specimen temperature onheating/cooling causing the reported temperature to be higheron heating and lower on cooling than the actual specimentransition temperature. A DSC apparatus temperature, that hasbeen calibrated for heating experiments, needs
9、 to be re-calibrated for cooling experiments. Such a calibration oncooling is performed using a liquid crystal material, thetransition(s) for which are not subject to super-heating orsuper-cooling.4.2 The transition temperature of a rapid, non-superheatingand non-supercooling transition is determine
10、d upon heatingand upon cooling. The difference between these two indicatedtemperatures provides an offset temperature value betweenheating and cooling experiments at the indicated rate. Thisoffset temperature value, when used with a precise temperaturecalibration upon heating, may serve as an instru
11、ment calibra-tion function upon cooling.5. Significance and Use5.1 This test method is useful in calibrating the temperaturesignal of a differential scanning calorimeter for cooling experi-ments such as the determination of crystallization temperaturesin Test Method D 3418 and Test Method E 794.1Thi
12、s test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on TestMethods and Recommended Practices.Current edition approved Aug. 15, 2006. Published October 2006. Originallyapproved in 2000. Last previous edition appro
13、ved in 2000 as E 206900.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Ba
14、rr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.2 This test method may be used for research, develop-ment, analytical, specification acceptance, quality assuranceand control purposes.6. Precautions6.1 Toxic or corrosive effluents, or both, may be releasedwhen heating
15、the material of this test method and may beharmful to personnel and to the apparatus.7. Apparatus7.1 Differential Scanning Calorimeter (DSC)The essen-tial instrumentation required providing the minimum differen-tial scanning calorimeter capability for this test method in-cludes:7.1.1 A DSC Test Cham
16、ber, composed of:7.1.1.1 A Furnace(s), to provide uniform controlled heatingand cooling of a specimen and reference material to a constanttemperature or at a constant rate within the applicable tempera-ture range of this method.7.1.1.2 A Temperature Sensor, that indicates specimen orfurnace temperat
17、ure to 60.01 C.7.1.1.3 A Differential Sensor, to detect a heat flow difference(DSC) between the specimen and reference with a range of atleast 100 mW readable to 6 1 W (DSC).7.1.1.4 A means of sustaining a purge gas rate of 10 to 1006 5 mL/min in the test chamber.NOTE 1Typically inert purge gases th
18、at inhibit specimen oxidation are99 + % pure nitrogen, argon or helium. Subambient operation requires drypurge gases. Dry gases are recommended for all experiments unless theeffect of moisture is part of the study.7.1.2 A Temperature Controller, capable of executing aspecific temperature program by
19、operating the furnace orfurnaces between selected temperature limits at a rate oftemperature change of 10 C/min constant to within 6 0.1C/min or at an isothermal temperature constant to 6 0.1 C.7.1.3 A Recording Device, capable of recording and display-ing fractions of the heat flow signal (DSC curv
20、e), including thesignal noise, on the Y-axis versus fractions of temperaturesignal, including the signal noise, on the X-axis.7.1.4 Containers, (pans, crucibles, vials, lids, closures,seals, etc.) that are inert to the specimen and referencematerials and that are of suitable structural shape and int
21、egrityto contain the specimen and reference in accordance with therequirements of this test method.NOTE 2DSC containers are commonly composed of aluminum orother inert material of high thermal conductivity. Aluminum has beentested and found compatible with the materials used in this standard.7.1.5 C
22、ooling Capability, at constant cooling rates of up to10 C/min in the temperature range of 185 to 50 C, to hastencool down from elevated temperatures, or to sustain anisothermal subambient temperature, or both.7.2 A Balance, to weigh specimen and/or containers to 6 10g with a capacity of 100 mg or gr
23、eater.8. Calibration Materials8.1 For the temperature range covered by many applica-tions, the liquid crystal transitions of 99.8 to 99.9 % purematerials in Table 1 may be used for calibration. The calibrat-ing liquid crystal materials3are known as M-24, BP-53 andBCH-52.NOTE 3The purity of these liq
24、uid crystal materials may be determinedby Test Method E 928 using the first liquid crystal transition prior to use(see Table 2).8.2 The approximate heat of transitions for these samples isshown in Table 2. The enthalpy of transition for M-24 is sosmall that it is detectable only on the most sensitiv
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME20692006STANDARDTESTMETHODFORTEMPERATURECALIBRATIONONCOOLINGOFDIFFERENTIALSCANNINGCALORIMETERS 扫描

链接地址:http://www.mydoc123.com/p-530199.html