ASTM E1618-2010 Standard Test Method for Ignitable Liquid Residues in Extracts from Fire Debris Samples by Gas Chromatography-Mass Spectrometry《使用气相色谱-质谱分析法测定火场碎屑样品的提取物中的可点燃液态残渣的标准.pdf
《ASTM E1618-2010 Standard Test Method for Ignitable Liquid Residues in Extracts from Fire Debris Samples by Gas Chromatography-Mass Spectrometry《使用气相色谱-质谱分析法测定火场碎屑样品的提取物中的可点燃液态残渣的标准.pdf》由会员分享,可在线阅读,更多相关《ASTM E1618-2010 Standard Test Method for Ignitable Liquid Residues in Extracts from Fire Debris Samples by Gas Chromatography-Mass Spectrometry《使用气相色谱-质谱分析法测定火场碎屑样品的提取物中的可点燃液态残渣的标准.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1618 10Standard Test Method forIgnitable Liquid Residues in Extracts from Fire DebrisSamples by Gas Chromatography-Mass Spectrometry1This standard is issued under the fixed designation E1618; the number immediately following the designation indicates the year oforiginal adoption or, in
2、 the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the identification of residues ofignitable liquids in extra
3、cts from fire debris samples. Extrac-tion procedures are described in the referenced documents.1.2 Although this test method is suitable for all samples, itis especially appropriate for extracts that contain high back-ground levels of substrate materials or pyrolysis and combus-tion products. This t
4、est method is also suitable for the identi-fication of single compounds, simple mixtures, or non-petroleum based ignitable liquids.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address
5、 all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E1386 Practice fo
6、r Separation of Ignitable Liquid Residuesfrom Fire Debris Samples by Solvent ExtractionE1388 Practice for Sampling of Headspace Vapors fromFire Debris SamplesE1412 Practice for Separation of Ignitable Liquid Residuesfrom Fire Debris Samples by Passive Headspace Concen-tration With Activated Charcoal
7、E1413 Practice for Separation and Concentration of Ignit-able Liquid Residues from Fire Debris Samples by Dy-namic Headspace ConcentrationE2154 Practice for Separation and Concentration of Ignit-able Liquid Residues from Fire Debris Samples by PassiveHeadspace Concentration with Solid Phase Microext
8、rac-tion (SPME)E2451 Practice for Preserving Ignitable Liquids and Ignit-able Liquid Residue Extracts from Fire Debris Samples3. Summary of Test Method3.1 The sample is analyzed with a gas chromatograph (GC)which is interfaced to a mass spectrometer (MS) and a datasystem (DS) capable of storing and
9、manipulating chromato-graphic and mass spectral data.3.2 Post-run data analysis generates extracted ion profiles(mass chromatograms) characteristic of the chemical com-pound types commonly found in ignitable liquids.Additionally,specific chemical components (target compounds) may beidentified by the
10、ir mass spectra and retention times. Semi-quantitative determination of target compounds which areidentified by mass spectra and retention time may be used todevelop target compound chromatograms (TCCs).3.2.1 The total ion chromatogram (TIC), extracted ionprofiles (EIP) for the alkane, alkene, alcoh
11、ol, aromatic, cycloal-kane, ester, ketone and polynuclear aromatic compound types,or TCCs, or combination thereof, are evaluated by visualpattern matching against known reference ignitable liquids.3.2.2 Ignitable liquids may be grouped into one of sevenmajor classifications or one miscellaneous clas
12、s, as describedin this test method.4. Significance and Use4.1 The identification of an ignitable liquid residue insamples from a fire scene can support the field investigatorsopinion regarding the origin, fuel load, and incendiary natureof the fire.4.1.1 The identification of an ignitable liquid res
13、idue in afire scene does not necessarily lead to the conclusion that a firewas incendiary in nature. Further investigation may reveal alegitimate reason for the presence of ignitable liquid residues.1This test method is under the jurisdiction ofASTM Committee E30 on ForensicSciences and is the direc
14、t responsibility of Subcommittee E30.01 on Criminalistics.Current edition approved Jan. 15, 2010. Published March 2010. Originallyapproved in 1994. Last previous edition approved in 2006 as E1618 061. DOI:10.1520/E1618-10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontac
15、t ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.4.1.2 Because of the vol
16、atility of ignitable liquids andvariations in sampling techniques, the absence of detectablequantities of ignitable liquid residues does not necessarily leadto the conclusion that ignitable liquids were not present at thefire scene.4.2 Materials normally found in a building, upon exposureto the heat
17、 of a fire, will form pyrolysis and combustionproducts. Extracted ion profiling and target compound identi-fication techniques described herein may facilitate the identi-fication of an ignitable liquid in the extract by reducinginterference by components generated as products of pyrolysis.5. Apparat
18、us5.1 Gas ChromatographA chromatograph capable of us-ing capillary columns and being interfaced to a mass spectrom-eter.5.1.1 Sample Inlet SystemA sample inlet system that canbe operated in either split or splitless mode with capillarycolumns; the inlet system may use on-column technology.5.1.2 Colu
19、mnA capillary, bonded phase, methylsiliconeor phenylmethylsilicone column or equivalent. Any columnlength or temperature program conditions may be used pro-vided that each component of the test mixture (see 6.4)isadequately separated.5.1.3 GC OvenA column oven capable of reproducibletemperature prog
20、ram operation in the range from 50 to 300C.5.2 Mass SpectrometerCapable of acquiring mass spectrafrom m/z 40 to m/z 400 with unit resolution or better, withcontinuous data output. Values above m/z 40 may not besufficient to detect or identify some lower molecular weightcompounds; for example, methan
21、ol, ethanol, acetone.5.2.1 SensitivityThe system shall be capable of detectingeach component of the test mixture (see 6.4) and providingsufficient ion intensity data to identify each component, eitherby computer library search or by comparison with referencespectra.5.3 Data StationA computerized dat
22、a station, capable ofstoring time sequenced mass spectral data from sample runs.5.3.1 Data HandlingThe data system shall be capable ofperforming, either through its operating system or by userprogramming, various data handling functions, including inputand storage of sample data files, generation of
23、 extracted ionprofiles, searching data files for selected compounds, andqualitative and semi-quantitative compound analysis.5.3.2 Mass Spectral LibrariesThe system shall be capableof retrieving a specified mass spectrum from a data file andcomparing it against a library of mass spectra available to
24、thedata system. This capability is considered an aid to the analyst,who will use it in conjunction with chromatographic data andknown reference materials to identify unknown components.5.4 Syringes:5.4.1 For liquid samples, a syringe capable of introducing asample size in the range from 0.1 to 10.0
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME16182010STANDARDTESTMETHODFORIGNITABLELIQUIDRESIDUESINEXTRACTSFROMFIREDEBRISSAMPLESBYGASCHROMATOGRAPHYMASSSPECTROMETRY

链接地址:http://www.mydoc123.com/p-529062.html