1、Designation: E1615 08E1615 16Standard Test Method forIron in Trace Quantities Using the FerroZine Method1This standard is issued under the fixed designation E1615; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last re
2、vision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of iron in the range from 0.01 to 0.2 g/g using FerroZine2 reagent solution. Thera
3、nge may be extended through the use of a 5- or 10-cm cell or by suitable dilution of the sample solution.1.2 This test method is intended to be general for the final steps in the determination of iron and does not include proceduresfor sample preparation.1.3 The values stated in SI units are to be r
4、egarded as standard. No other units of measurement are included in this standard.1.4 Review the current material safety data sheets (MSDS) Safety Data Sheets (SDS) for detailed information concerningtoxicity, first-aid procedures, and safety precautions.1.5 This standard does not purport to address
5、all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use. For specific warning statements, see 7.4.2. Referenced Documen
6、ts2.1 ASTM Standards:3D1193 Specification for Reagent WaterE60 Practice for Analysis of Metals, Ores, and Related Materials by SpectrophotometryE180 Practice for Determining the Precision of ASTM Methods for Analysis and Testing of Industrial and Specialty Chemicals(Withdrawn 2009)4E200 Practice for
7、 Preparation, Standardization, and Storage of Standard and Reagent Solutions for Chemical AnalysisE275 Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers3. Summary of Test Method3.1 This test method is based upon a photometric determination of the FerroZi
8、ne2 complex with the iron (II) ion.5,6 The sampleis dissolved in a suitable solvent and the iron is reacted with FerroZine reagent solution which will convert the dissolved ironcompounds to form a magenta color iron (II) complex. The iron content of the sample solution is determined by measurement o
9、fthe magenta color at 560 nm using a suitable photometer.4. Significance and Use4.1 This test method is suitable for determining trace concentrations of iron in a wide variety of products, provided thatappropriate sample preparation has rendered the iron and sample matrix soluble in water or other s
10、uitable solvent. Each samplematrix must be investigated for suitability using this test method.4.2 This test method assumes that the amount of color developed is proportional to the amount of iron in the test solution. Thecalibration curve is linear over the specified range.1 This test method is und
11、er the jurisdiction of ASTM Committee E15 on Industrial and Specialty Chemicalsand is the direct responsibility of Subcommittee E15.01 onGeneral Standards.Current edition approved April 1, 2008April 1, 2016. Published May 2008May 2016. Originally approved in 1994. Last previous edition approved in 2
12、0052008 asE161505.08. DOI: 10.1520/E1615-08.10.1520/E1615-16.2 FerroZine is a trademark of Hach Chemical Company.3 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the
13、standards Document Summary page on the ASTM website.4 The last approved version of this historical standard is referenced on www.astm.org.5 Stookey, L. L., “FerroZineA New Spectrophotometric Reagent for Iron,” Analytical Chemistry, Vol 42, No. 7, June 1970, pp. 779 781.6 Gibbs, C. R., “Characterizat
14、ion and Application of FerroZine Iron Reagent as a Ferrous Iron Indicator,” Analytical Chemistry , Vol 48, No. 8, July 1976, pp. 11971201.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous ve
15、rsion. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes se
16、ction appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Interferences5.1 Any ion that absorbs light at 560 nm will interfere with the determination. Anionic interferences include oxalate inconcentra
17、tions over 500 g/g, cyanide, and nitrate.55.2 Of copper, cobalt, calcium, magnesium, lead, silver, molybdenum, aluminum, nickel, zinc, arsenic, manganese, hexavalentchromium, trivalent chromium, divalent cobalt and monovalent copper are the only metals other than iron that form colored specieswith F
18、erroZine under test conditions. At least 1000 mg/L of the alkali metals and the alkaline earths had no effect on thedetermination. Many heavy metals will react with FerroZine in competition with iron, but with the excess reagent used in the testthere is no effect on the results.55.3 The pH range of
19、the final solution should be from 4 to 9 to give the best test results.5,65.4 All glassware used in this test method must be iron-free and scrupulously clean by precleaning with dilute hydrochloric acidand FerroZine reagent solution followed by a water rinse.6. Apparatus6.1 Photometer, capable of me
20、asuring light absorption at 560 nm and holding a 5-cm or 10-cm cell. Check the performance ofthe photometer at regular intervals according to the guidelines given in Practice E275 and the manufacturers manual.6.2 Absorption Cells, 5-cm or 10-cm light path.NOTE 1A discussion of photometers and photom
21、etric practice is given in Practice E60.7. Reagents7.1 Unless otherwise indicated, it is intended that all reagents shall conform to the specifications of the Committee onAnalyticalReagents of the American Chemical Society, where such specifications are available.7 Other grades may be used, provided
22、 it isfirst ascertained that the reagent is of sufficiently high purity to permit its use without lessening the accuracy of the determination.7.2 Purity of WaterUnless otherwise indicated, references to water shall be understood to mean Type II or Type III reagentwater conforming to Specification D1
23、193.7.3 Iron, Standard Solution, 1 mL = 1 g Fe (see Note 2Notes 2 and 3 and Note 3)Dissolve 0.1000 g of iron wire in 10 mLof hydrochloric acid (HCl, 1 + 1) and 1 mL of saturated bromine water (400 mL water + 20 mL bromine). Boil until the excessbromine is removed. Add 200 mL of HCl, cool, and dilute
24、 to 1 L 1 L in a volumetric flask. Dilute 10 mL of this solution to 1 L.NOTE 2The preparation of this reagent is also described in Practice E200.NOTE 3As an alternative, the standard iron solution may be prepared by diluting 1.00 mL of commercially available iron standard stock solution(1000 mg iron
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME16152016STANDARDTESTMETHODFORIRONINTRACEQUANTITIESUSINGTHEFERROZINEMETHOD 采用 铁合金 测定 痕量 标准 试验 方法

链接地址:http://www.mydoc123.com/p-529056.html